Grundfos AQC-D12 Installation And Operating Instructions Manual

Tip
Installation And Operating Instructions Manual
AQC-D12
Potentiostatic Measuring Cell,
Compact Measuring Systems
Installation and operating instructions
GRUNDFOS INSTRUCTIONS
2
Table of contents
3
AQC-D12
English (GB)
Installation and operating instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Deutsch (DE)
Montage- und Betriebsanleitung . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Español (ES)
Instrucciones de instalación y funcionamiento. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Français (FR)
Notice d'installation et de fonctionnement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Magyar (HU)
Telepítési és üzemeltetési utasítás . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Italiano (IT)
Istruzioni di installazione e funzionamento. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Nederlands (NL)
Installatie- en bedieningsinstructies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Polski (PL)
Instrukcja montażu i eksploatacji . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Português (PT)
Instruções de instalação e funcionamento. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
Română (RO)
Instrucţiuni de instalare şi utilizare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
Русский (RU)
Паспорт, Руководство по монтажу и эксплуатации . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
Slovensko (SI)
Navodila za montažo in obratovanje . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224
Türkçe (TR)
Montaj ve kullanım kılavuzu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265
English (GB)
4
English (GB) Installation and operating instructions
Original installation and operating instructions
CONTENTS
Page
1. Symbols used in this document
1. Symbols used in this document
4
2. Description of preassembled systems with
AQC-D12
5
3. Identification
5
3.1 Nameplate, AquaCell
5
3.2 Type key, AquaCell
6
3.3 Type key, preassembled systems
7
4. Description of AQC-D12
8
5. General information
8
6. Applications
8
7. Safety information
9
7.1 Obligations of the owner
9
7.2 Avoidance of danger
9
8. Technical data
9
8.1 General data
9
8.2 Versions
9
8.3 Measuring ranges
10
8.4 Dimensions / drilling diagram
10
9. Installation
11
9.1 Transport and storage
11
9.2 Unpacking
11
9.3 Installation requirements
11
9.4 Mounting
11
10. Startup
12
10.1 Installation of electrodes and sensors
12
10.2 Checking the position of the cleaning wing
12
10.3 Water connections
13
10.4 Preparing the electrode cable for connection to the
measuring amplifier
13
10.5 Electrical connections
14
10.6 Checks prior to start-up
15
10.7 Switching on
15
10.8 Calibrating the parameters Cl
2
, ClO
2
, O
3
16
11. Operation
18
11.1 Function
18
11.2 Operation
19
11.3 Switching off
19
11.4 Switching on again
19
11.5 Fault finding
20
12. Maintenance
21
12.1 Intervals for cleaning and maintenance
21
12.2 Cleaning and replacing the filter
21
12.3 Cleaning the measuring cell
21
13. Starting up the measuring cell
23
14. Spare parts and accessories
23
14.1 Electrodes, sensors and cables
23
14.2 Other parts
23
15. Disposal
23
Warning
Prior to installation, read these installation and
operating instructions. Installation and operation
must comply with local regulations and accepted
codes of good practice.
Note
These complete installation and operating
instructions are also available on www.grundfos.com.
Warning
If these safety instructions are not observed, it may
result in personal injury.
Caution
If these safety instructions are not observed, it may
result in malfunction or damage to the equipment.
Note
Notes or instructions that make the job easier and
ensure safe operation.
English (GB)
5
2. Description of preassembled systems with
AQC-D12
Fig. 1 AQC-D12 measuring cell with Conex DIA / DIS
measuring amplifier and controller
Fig. 2 AQC-D12 measuring cell with sensor interface
Fig. 3 AQC-D12 measuring cell with DIP measuring amplifier
and controller
3. Identification
3.1 Nameplate, AquaCell
Fig. 4 Nameplate, AquaCell
TM04 8617 3912
TM04 8618 3912
TM04 8619 3912
1
Conex DIA-1, Conex DIA-2, Conex DIA-2Q, Conex
DIS-D measuring amplifier and controller
2
Sensor interface for Conex DIA-1, Conex DIA-2,
Conex DIA-2Q, for panel-mounting
3 DIP measuring amplifier and controller
Man
Cal
OKEsc
1
2
DIP
Man Cal
OKEsc
Select
Select
3
TM04 8668 3013
Pos. Description
1 Type designation
2 Model
3 Serial number
4 Maximum pressure [bar]
5 Product number
6 Year and week code
7 Country of origin
AQC-D11, P-AU-PCB-RCB, QS-T, H
314-306-10048
S/N: 07/44888
3,00 bar
95737669P11124563516
1
2
4
5
6
7
3
English (GB)
6
3.2 Type key, AquaCell
Example: AQC -D11, P AU-PCB-RCB, QS -T, H
Model
AQC AquaCell
AquaCell type
D11 Pressure-proof, with cleaning motor (Cl
2
, ClO
2
, O
3
)
D12 Pressure-proof, with hydromechanical cleaning (Cl
2
, ClO
2
, O
3
)
D13 Pressureless, with hydromechanical cleaning (Cl
2
, ClO
2
, O
3
)
Pressure-loading valve
P With pressure-loading valve
X Without pressure-loading valve
Electrodes for disinfection parameters
AU Gold
PT Platinum
Electrodes for pH
PCB pH, ceramic diaphragm, with buffer solution
PCX pH, ceramic diaphragm, without buffer solution
PTB pH, PTFE diaphragm, with buffer solution
PTX pH, PTFE diaphragm, without buffer solution
PKB pH, KCl filling, with buffer solution
PKX pH, KCl filling, without buffer solution
PGB pH, gel filling, with buffer solution
PGX pH, gel filling, without buffer solution
X Without electrode
Electrodes for redox potential (ORP)
RCB ORP, ceramic diaphragm, with buffer solution
RCX ORP, ceramic diaphragm, without buffer solution
RTB ORP, PTFE diaphragm, with buffer solution
RTX ORP, PTFE diaphragm, without buffer solution
RRB ORP, without reference system, with buffer solution
RRX ORP, without reference system, without buffer solution
X Without electrode
Water sensor
QS With water sensor
X Without water sensor
Temperature sensor
T With Pt100 temperature sensor
X Without temperature sensor
Power supply
G 230/240 V, 50/60 Hz
H 115/120 V, 50/60 Hz
I 24 VDC
X Without power supply
English (GB)
7
3.3 Type key, preassembled systems
Example: DIA -1 -A D1 P -AU -PCB -QS -T W -H
Measuring amplifier and controller
DIA-1 Dosing Instrumentation Advanced with 1 input
DIA-2 Dosing Instrumentation Advanced with 2 inputs
DIA-2Q Dosing Instrumentation Advanced with 1 input + flow measurement
DIP Dosing Instrumentation Pool
DIS-PR Dosing Instrumentation Standard for pH/ORP measurement
DIS-D Dosing Instrumentation Standard for Cl
2
/ ClO
2
/ O
3
Assembly
APreassembled
Cell type
D11 Pressure-proof, with cleaning motor
D12 Pressure-proof, with hydromechanical cleaning
D13 Pressureless, with hydromechanical cleaning
D4 Pressureless, with cleaning motor, for total chlorine
D5 Pressureless, with cleaning motor, for free chlorine
P/R pH or ORP
PA/HP Peracetic acid or hydrogen peroxide
Pressure-loading valve
P With pressure-loading valve
X Without pressure-loading valve
Electrodes for disinfection parameters
AU Gold
PT Platinum
X Without electrode
Other electrodes
PCB pH, ceramic diaphragm, with buffer solution
PCX pH, ceramic diaphragm, without buffer solution
PTB pH, PTFE diaphragm, with buffer solution
PTX pH, PTFE diaphragm, without buffer solution
PKB pH, KCl filling, with buffer solution
PKX pH, KCl filling, without buffer solution
PGB pH, gel filling, with buffer solution
PGX pH, gel filling, without buffer solution
RCB ORP, ceramic diaphragm, with buffer solution
RCX ORP, ceramic diaphragm, without buffer solution
RTB ORP, PTFE diaphragm, with buffer solution
RTX ORP, PTFE diaphragm, without buffer solution
RRB ORP, without reference system, with buffer solution
RRX ORP, without reference system, without buffer solution
PA Peracetic acid
HP Hydrogen peroxide
X Without electrode
Water sensor
QS With water sensor
X Without water sensor
Temperature sensor
T With Pt100 temperature sensor
X Without temperature sensor
Mounting
W Wall-mounting
P Panel-mounting
Power supply
G 230/240 V, 50/60 Hz
H 115/120 V, 50/60 Hz
I 24 VDC
English (GB)
8
4. Description of AQC-D12
Fig. 5 AQC-D12 measuring cell
Fig. 6 AQC-D12, top view
5. General information
This manual contains all important information for users of the
AQC-D12 and the preassembled systems:
technical data
instructions for start-up, use and maintenance
safety information.
Should you require further information, or should problems occur
which are not handled in sufficient depth in this manual, please
contact Grundfos.
6. Applications
The potentiostatic measuring cell AQC-D12 measures the
concentration of chlorine (Cl
2
) in the pH-range 4.5 to 8.2, chlorine
dioxide (ClO
2
) or ozone (O
3
). By means of the suitable
electrodes, it can also measure the pH value and redox potential
for treatment of swimming-pool water and drinking water.
The preassembled systems measure and control the following
parameters, depending on the measuring amplifier and controller
installed:
Preassembled system with Conex DIA-1 measuring amplifier
and controller:
– Measurement: free chlorine (Cl
2
), as an option with pH
compensation, chlorine dioxide (ClO
2
), ozone (O
3
), pH,
temperature.
– Control: chlorine (Cl
2
), chlorine dioxide (ClO
2
), ozone (O
3
).
Preassembled system with Conex DIA-2 measuring amplifier
and controller:
– Measurement: free chlorine (Cl
2
), as an option with pH
compensation, chlorine dioxide (ClO
2
), ozone (O
3
), pH,
temperature.
– Control: chlorine (Cl
2
), chlorine dioxide (ClO
2
), ozone (O
3
),
pH.
Preassembled system with Conex DIA-2Q measuring amplifier
and controller:
– Measurement: free chlorine (Cl
2
), as an option with pH
compensation, chlorine dioxide (ClO
2
), ozone (O
3
), pH,
temperature.
– Control: chlorine (Cl
2
), chlorine dioxide (ClO
2
), ozone (O
3
),
pH, redox potential.
Preassembled system with Conex DIS-D measuring amplifier
and controller:
– Measurement: chlorine (Cl
2
), chlorine dioxide (ClO
2
), ozone
(O
3
).
– Control: chlorine (Cl
2
), chlorine dioxide (ClO
2
), ozone (O
3
).
Preassembled system with DIP measuring amplifier and
controller:
– Measurement: chlorine (Cl
2
), chlorine dioxide (ClO
2
), ozone
(O
3
), pH, redox potential, temperature.
– Control: chlorine (Cl
2
), chlorine dioxide (ClO
2
), ozone (O
3
),
pH.
TM04 8620 3912TM04 8621 3912
Pos. Description
A pH electrode / location for pH electrode
B Redox electrode / location for redox electrode
C
Pt100 temperature sensor / location for temperature
sensor (C1) with screw plug (C2)
D Water sensor
E Connection for sample water inlet
F Connection for sample water outlet
X Sample water shut-off valve
G Regulation of sample water quantity
H Sample water filter
I Sampling cock
J Measuring electrode for Cl
2
, ClO
2
, O
3
K
Reference electrode for Cl
2
, ClO
2
, O
3
/ location for
reference electrode
L Cover with threaded ring
M Acrylic glass cylinder
A / B
L
K
M
X, G
E
J
H
I
G
X
D
C2
F
A / B
KDC
English (GB)
9
7. Safety information
7.1 Obligations of the owner
The owner of the plant is responsible for
compliance with country-specific safety regulations
training of operating personnel
provision of prescribed protective gear
implementation of regular maintenance.
7.2 Avoidance of danger
8. Technical data
8.1 General data
The AQC-D12 is a pressure-proof measuring cell with
hydromechanical cleaning for recirculation of sample water.
Measuring parameters:
– free chlorine, chlorine dioxide, ozone
– optional: pH value, redox potential, temperature (for
compensation of pH, Cl
2
, ClO
2
, O
3
).
Sample water:
– free of solids
– min. conductivity: 50 µS/cm
– temperature: 0-40 °C
– pH value (only measuring parameter Cl
2
): pH 4.5 - 8.2
– inlet pressure: max. 3 bar, min. 0.2 bar
– counterpressure: max. 3 bar
– pressure difference: approx. 0.2 bar.
Permissible ambient temperature: 0 to 45 °C.
Permissible storage temperature:
– -20 to +65 °C
– Electrodes: -10 to +30 °C.
Permissible humidity: max. 90 % relative humidity (without
condensation).
Sample water flow rate: min. 20; max. 45 l/h.
Materials of parts that come into contact with the media:
PMMA, PVC, EPDM.
Electrode materials: gold, platinum, glass, nickel, PVC.
Connections:
– inlet: for PVC hose 6/12 or PVC pipe 12 x 1.2
– outlet: for PVC hose 6/12 or PVC pipe 12 x 1.2.
Measuring parameters: chlorine, chlorine dioxide and ozone
Measuring ranges: Dependent on the measuring amplifier.
See section 8.3.1 Measuring ranges for Cl
2
, ClO
2
, O
3
, pH,
ORP.
Sensitivity: < 20 ppb.
Accuracy: < - 5 %/+ 5 % of full-scale value.
Repeatability: < - 5 %/+ 5 %.
Response time: t
90
< 60 s.
8.2 Versions
Power supply:
– 230/240 V, 50/60 Hz (standard)
– 115/120 V, 50/60 Hz
– 24 VDC (not with DIP).
Equipment/options:
– gold electrode for free chlorine, chlorine dioxide, ozone
(standard)
– platinum electrode
– redox single-rod measuring chain (electrode)
– redox electrode (with DIP)
– pH single-rod measuring chain (electrode).
8.2.1 AquaCell AQC-D12
Cables:
– connecting cables for the electrodes, 3 m, with free cable
end
– cable for measuring electrode, 2 m, with free cable end
– water sensor with cable, 3 m, with free cable end.
Options:
– Pt100 temperature sensor with cable, 3 m.
8.2.2 Preassembled systems
Cables:
– connecting cables for the electrodes, 1 m, preconnected
– cable for measuring electrode, 1.2 m, preconnected
– water sensor with cable, 1 m, preconnected.
Options:
– Pt100 temperature sensor with cable, 1 m.
Product numbers of the spare parts and optional accessories, see
section 14. Spare parts and accessories.
Warning
Cleaning, maintenance and repairs must only be
carried out by authorised personnel!
Warning
Other applications than those described in section
6. Applications are considered as non-approved and
are not permissible.
Derivates of chloro isocyanuric acid cannot be
measured or controlled.
Grundfos cannot be held liable for any damage
resulting from incorrect use.
Caution
Observe the installation and operating instructions of
the electrodes!
English (GB)
10
8.3 Measuring ranges
8.3.1 Measuring ranges for Cl
2
, ClO
2
, O
3
, pH, ORP
Preassembled system with Conex DIA
Preassembled system with Conex DIS-D
* Effective measuring range: 0.00 - 5.00
Preassembled system with DIP
8.3.2 Measuring ranges for temperature
(not with Conex DIS-D)
8.4 Dimensions / drilling diagram
Fig. 7 AQC-D12 measuring cell
Fig. 8 Preassembled systems
(example: system with Conex DIA for wall-mounting)
Cl
2
[mg/l]
ClO
2
[mg/l]
O
3
[mg/l]
pH
ORP
[mV]
0.00 - 0.50 0.00 - 0.50 0.00 - 0.50 0.00 - 14.00
-1500 -
+1500
0.00 - 1.00 0.00 - 1.00 0.00 - 1.00 2.00 - 12.00 0-1000
0.00 - 2.00 0.00 - 2.00 0.00 - 2.00 5.00 - 9.00
0.00 - 5.00 0.00 - 5.00 0.00 - 5.00
0.0 - 10.0 0.0 - 10.0
0.0 - 20.0
freely selectable between
0.0 - 50.0 0.0 - 50.0 0.00 - 5.00 0.00 - 14.00
-1500 -
+1500
Cl
2
[mg/l]
ClO
2
[mg/l]
O
3
[mg/l]
0.00 - 2.00 0.00 - 2.00 0.00 - 2.00
0.00 - 20.00 0.00 - 20.00 0.00 - 20.00*
Cl
2
[mg/l]
ClO
2
[mg/l]
O
3
[mg/l]
pH
ORP
[mV]
0.00 - 0.50 0.00 - 0.50 0.00 - 0.50 0.00 - 14.00
-1500 -
+1500
0.00 - 1.00 0.00 - 1.00 0.00 - 1.00 2.00 - 12.00 0-1000
0.00 - 2.00 0.00 - 2.00 0.00 - 2.00 5.00 - 9.00
0.00 - 5.00 0.00 - 5.00 0.00 - 5.00
0.0 - 10.0 0.0 - 50.0
0.0 - 30.0
freely selectable between
0.0 - 30.0 0.0 - 50.0 0.00 - 5.00 0.00 - 14.00
-1500 -
+1500
°C °F
0 to +50 32 to 122
0 to +100 32 to 212
-5 to +120 23 to 248
TM04 8622 3912TM04 8623 3912
365
105
120
125
137
475
495
475
133
117
Man
Cal
OKEsc
495
English (GB)
11
9. Installation
9.1 Transport and storage
9.2 Unpacking
Check scope of delivery.
Assemble as soon as possible after unpacking.
9.3 Installation requirements
The location must be vibration-free, dry, dust-free and free of
corrosive, pungent fumes or aggressive solvents.
Maximum permissible cable length:
– Individual AQC-D12 measuring cell: distance between
measuring cell and measuring amplifier or sensor interface
3 m.
– Preassembled systems with Conex DIS, DIA or DIP for
wall-mounting: completely prewired.
– Preassembled systems with Conex DIA for panel-mounting:
the distance between measuring amplifier and sensor
interface must be less than 100 m.
Ensure a continuous supply of sample water.
Install the measuring cell so that the sample water feed line is
as short as possible, in order to reduce the delay time of the
measuring cell.
Observe the permissible inlet pressure and counterpressure of
the sample water.
– Fit a pressure booster pump or pressure reducer, if
necessary.
If the counterpressure may fall below 0.3 bar or if there is an
open outflow, a pressure-loading valve must be fitted to the
outlet of the measuring cell (F). The pressure-loading valve
(V) is available with an adaptor for the measuring cell
(Grundfos product number 99253231).
Provide possibilities to shut-off the sample water (U). You can
use the Grundfos installation set 95727420 on the discharge
side.
Fig. 9 AQC-D12 with pressure-loading valve / installation set
9.4 Mounting
The measuring cell is fastened to the mounting plate at the
factory.
Individual AQC-D12 measuring cell
1. Drill two dowel holes with a diameter of 8 mm and min. 50 mm
deep. See section 8.4 Dimensions / drilling diagram.
2. Mount the measuring cell and the mounting plate on the wall
using the mounting material provided.
Fig. 10 Mounting of AQC-D12 measuring cell
Preassembled systems
1. Drill four dowel holes with a diameter of 10 mm and min. 60
mm deep. See section 8.4 Dimensions / drilling diagram.
2. Mount the preassembled system and the mounting plate on
the wall using the mounting material provided. Distance
between the mounting plate and the wall:
min. 20 mm.
Fig. 11 Mounting sequence for preassembled systems
Caution
Transport device carefully, do not throw.
Store in dry conditions between -20 and +65 °C.
Store electrodes between -10 and +30 °C. Keep
protective caps moist with 3-molar potassium
chloride solution.
Caution
Do not allow any foreign bodies to enter!
Warning
For safety reasons, the customer must install an
earth leakage circuit breaker for the measuring
amplifier.
Caution
Non-observance of the installation requirements may
result in damage or errors in measurement!
Note
For safety during service and maintenance, switch
off the measuring amplifier and the measuring cell at
all poles.
For easy switch-off, install an all-pole mains switch in
front of the measuring amplifier.
Note
The installation of the sample-water feed pipe must
not contain any copper. Copper interferes with the
measurement and can damage the electrodes.
TM04 8624 3912
Caution
The measuring cell or preassembled system must be
mounted vertically!
TM03 5856 1106
Caution
Do not pinch the cable!
Observe the mounting sequence below.
TM03 5857 1106
F
V
U
F
20 mm
English (GB)
12
10. Startup
10.1 Installation of electrodes and sensors
10.1.1 Fitting of electrodes
1. Unscrew the plugs of the holders (A, B) to be used.
2. Remove the protective caps from the electrodes. Keep them
for electrode removal.
3. Screw the electrodes into the holders by hand.
– Screw reference electrode in location K.
– Screw pH electrode in location A.
– Screw redox electrode in location B.
Fig. 12 Holders for electrodes
10.1.2 Water sensor
10.1.3 Temperature sensor
1. Remove lock screw (C2), store for later use.
2. Screw in Pt100 temperature sensor (C1) (optional).
Fig. 13 Temperature sensor
10.2 Checking the position of the cleaning wing
1. Visual inspection:
– The cleaning wing (S) must be seated on the measuring
electrode as shown.
Fig. 14 Position of the cleaning wing
2. If the cleaning wing is not seated correctly:
– Screw out the measuring electrode.
– Insert cleaning wing into the bore of the measuring electrode
using the entry pilot. It must be possible to freely move the
cleaning wing.
– Carefully screw in the measuring electrode again.
Caution
Observe the installation and operating instructions of
the electrodes!
TM04 8625 3912
Note
The water sensor is next to the cleaning wing.
pH (A) / Redox (B)
K
pH (A) / Redox (B)
J
F
E
E
TM04 8626 3912
Caution
The cleaning wing falls out when the measuring
electrode (J) is removed.
TM04 8627 3912
C
C2
S
J
English (GB)
13
10.3 Water connections
If the counterpressure may fall below 0.3 bar or if there is an open
outflow, a pressure-loading valve must be fitted to the outlet of
the measuring cell.
1. Connect the sample water inlet and outlet to the connections
for the sample water inlet (E) and outlet (F).
Fig. 15 Water connections
10.3.1 Mounting the pressure-loading valve
1. Unscrew the connection of the sample water outlet (F).
2. Screw the pressure-loading valve and adaptor (V) onto the
measuring cell outlet.
3. Screw the connection of the sample water outlet onto the
pressure-loading valve.
Fig. 16 Pressure-loading valve
10.4 Preparing the electrode cable for connection to
the measuring amplifier
Cables for electrodes (A, B, K)
1. Cut the electrode cables to the desired length plus
approximately 80 mm for the connections.
2. Remove 80 mm of the outer insulation.
3. Disentangle the braided screen until you reach the next part of
the insulation, and twist it to form a wire.
4. Strip the braided screen (for instance using a shrink hose),
and fit a wire end ferrule.
5. Remove the black (conductive) sheath of the insulated
electrode wire until you reach the insulation.
6. Dismantle the end of the electrode wire.
7. Fit a wire end ferrule to the electrode wire.
Fig. 17 Cable for electrodes (A, B, K)
Cable for measuring electrode (D), temperature sensor (C),
water sensor (D)
1. Cut cables to the desired length plus approximately 65 mm for
the connections.
2. Remove 65 mm of the outer insulation.
3. Dismantle the end of the wires.
4. Fit wire end ferrules to the wires.
Warning
Danger of injuries!
At a pressure of more than 3 bar and if the
measuring cell is not deaerated, the cell might burst.
Do not exceed the max. system pressure of 3 bar.
Fit pressure reducer, if necessary.
Caution
Observe the local pressure! The permissible primary
water pressure is 0.3 to 3.0 bar.
Fit a pressure booster pump, if necessary.
Check the tightness of the measuring cell.
Caution
Only tighten the union nut by hand. Do not use any
tools!
TM04 8628 3912TM04 8788 1313
Caution
Observe the correct installation of the
pressure-loading valve. The arrow on the
pressure-loading valve must point in flow direction!
F
E
E
F
V
Note
The preassembled systems are prewired.
Warning
The electrical connection should be carried out by
qualified personnel!
Observe the local safety regulations!
Protect the cable connections and plugs against
corrosion and humidity.
TM03 5868 1106
TM03 5869 1106
70 10
Insulated
braided
screen
English (GB)
14
10.5 Electrical connections
Preassembled systems
The preassembled systems are prewired.
1. Connect an earth leakage circuit breaker in front of the
measuring amplifier.
2. Connect the electricity supply to the measuring amplifier.
AQC-D12 measuring cell
1. Connect electrodes to the corresponding terminals of the
measuring amplifier.
2. Connect measuring electrodes to the corresponding terminals
of the measuring amplifier. Fit the screen according to the
installation and operating instructions of the measuring
amplifier and controller.
3. Connect the water sensor to the corresponding terminals of
the measuring amplifier.
4. Connect an earth leakage circuit breaker in front of the
measuring amplifier.
5. Connect the electricity supply to the measuring amplifier.
Fig. 18 Conex DIA-1 / DIA-2 / DIA-2Q for panel-mounting
Fig. 19 Conex DIA-1 / DIA-2 / DIA-2Q for wall-mounting
Warning
The electrical connection should be carried out by
qualified personnel!
Before connecting the supply cables, check that the
supply voltage specified on the nameplate
corresponds to the local conditions!
Before connecting the supply cables, switch off the
electricity supply!
Observe the local safety regulations!
Protect the cable connections and plugs against
corrosion and humidity.
Connect an earth leakage circuit breaker in front of
the measuring amplifier.
Caution
Observe the installation and operating instructions of
the measuring amplifier and controller!
Note
The cables are not preconnected. See section
10.4 Preparing the electrode cable for connection to
the measuring amplifier.
Caution
Observe the installation and operating instructions of
the measuring amplifier and controller!
Caution
Before connecting the supply cables, check that the
supply voltage specified on the nameplate
corresponds to the local conditions!
Observe the installation and operating instructions of
the measuring amplifier and controller!
TM04 8629 3912TM04 8630 3912
1/11
1/13
4/1
1/12 1/14
4/3
4/2
4/4
3/1 3/3
3/2 3/4
B
A
1
4
Conex
2/3
2/4
+
-
12
MB/RG/C
7
1
6
Pt 100
2
8
9
1
4
11
8
11
12
9
9
10
Sensorinterface
7
1
37
38
39 40 41 42
6
12
4
1
15 17
19
21 23 25 27 29 31 33 35
363432302826242220
18
16
9
8
2
1
4
11
8
11
12
9
9
10
English (GB)
15
Fig. 20 Conex DIS-D
Fig. 21 DIP
10.6 Checks prior to start-up
Check the tightness of the whole system.
Check the electrical connections.
10.7 Switching on
Fig. 22 Setting the quantity of sample water
1. Close the sample water shut-off valve (X).
2. Close the regulation of sample water quantity (G).
3. Close the sampling cock (I).
4. Open the inlet and outlet side of the customer's sample water
connection.
5. For deaeration, loosen the pH or redox electrode until the air
has escaped, then fasten it.
6. Open sample water shut-off valve (X) approx. two turns up to
the limit stop.
7. Open regulation of sample water quantity (G) slowly.
– Avoid turbulent flow!
8. Check that the cleaning wing (S) rotates.
– Try, if necessary, to briefly increase the quantity of sample
water with the regulation of sample water quantity (G) in
order to move the cleaning wing.
– If the cleaning wing still does not rotate, check and correct
its position, if necessary. See section 10.2 Checking the
position of the cleaning wing.
9. Switch on the measuring amplifier.
10. Only switch on the device controlled after the first calibration,
if applicable.
TM04 8631 3912TM04 8632 3912
Pos. Description
1Brown
2White
3Black
4Blue
6 Outer conductor (screen)
7 Inner conductor
8 Reference electrode
9 Measuring electrode with counter-electrode
10 Counter-electrode
11 Pt100 temperature sensor (not with Conex DIS-C)
12 Water sensor
7
1
6
9
8
2
20 2122
19
12
4
1
14 16
8
12
9
9
10
7
1
6
12
4
9
8
2
1
4
1
53
55 57
54 56 58
63 65
64
66
11
8
11
12
9
9
10
Caution
Observe the installation and operating instructions of
the measuring amplifier and the device controlled!
Warning
At a pressure of more than 3 bar and if the
measuring cell is not deaerated, the cell might burst.
Ensure draining and deaeration.
Do not exceed the max. system pressure of 3 bar.
Do not put a damaged measuring cell under
pressure!
Caution
At first start-up or after long-term stop: Let the
system run for at least two hours to avoid faulty
measurements and calibration!
TM04 8633 3912
I
GX
English (GB)
16
10.7.1 Setting the quantity of sample water
A correct measurement can only be taken, if the quantity of
sample water is within a defined range. This range is from 200 to
800 rotations/minute of the cleaning wing.
Set the quantity of sample water with the regulation (G) so that
the measuring amplifier shows a sample water flow rate of 200
to 800 rotations/minute. The bar in the display of the
measuring amplifier shows the range. The exact speed can be
taken from the service menu.
10.8 Calibrating the parameters Cl
2
, ClO
2
, O
3
Due to the electro-chemical behaviour of the measuring cell, no
zero-point adjustment is necessary. Only the rate of rise
(sensitivity) must be adjusted during calibration.
Check the calibration after 24 hours. Repeat, if necessary.
10.8.1 Photometrical determination of a reference value
1. Open the sampling cock (I), and let the water run for a few
seconds.
2. Take a water sample, and close the sampling cock.
3. Determine the concentration of the measuring parameter
photometrically, for instance using the Grundfos DIT hand
photometer.
4. Keep this value as a reference value.
10.8.2 Calibration example based on chlorine measurement
with pH compensation (not with Conex DIS-D)
Define the basic settings for the measuring amplifier.
Calibrating the pH value
DIP: Push the button underneath the right-hand display.
Conex DIA and DIP: Use the [CAL] button to select the
calibration function CAL. The LED illuminates.
1. If necessary, enter the code number for CAL (or full)
authorisation.
2. Change to the menu "calibration" with the [OK] button.
3. Select the measured variable "pH", and confirm with [OK].
– Selection possibilities: "cal. meas. value" (= measured
value), "cal. result" and "cal. cycle" (= count-down function
which triggers the alarm "calibrate sensor" after a selectable
time interval of 1-100 days).
4. Select the line "cal. meas. value" with [UP]/[DOWN].
5. Select one of the defined buffer solutions ("GRUNDFOS",
"DIN/NIST", "others"), and press [OK].
– The menu "temperature" is selected automatically.
6. Enter the temperature of the buffer solutions.
7. Press [OK] to jump to the menu "buffer value 1".
8. "GRUNDFOS" or "DIN/NIST" buffer solutions: Select one of
the available buffer values.
Caution
At first start-up or after long-term stop: Let the
system run for at least two hours to avoid faulty
measurements and calibration!
Caution
During calibration: Keep pH value, sample water flow
rate and water temperature constant.
Caution
Observe the installation and operating instructions of
the photometer!
Caution
Observe the installation and operating instructions of
the measuring amplifier and controller!
Note
When carrying out a chlorine measurement with pH
compensation, the pH value must be calibrated first,
since the pH value is used when calibrating the
chlorine value!
pH
cal. meas. value
cal. result
cal. cycle
buffer
GRUNDFOS
DIN/NIST
others
buffer value 1
4.01 pH
7.00 pH
9.18 pH
English (GB)
17
Buffer 1
1. Fill buffer solution 1 from the supply bottle into a clean vessel.
2. Remove the electrode from the measuring cell, and immerse
in the buffer solution.
3. Press [OK] to read in the temperature of the sample water and
the measured signal.
4. Discard buffer solution 1.
– Do not return the used buffer solution to the supply bottle!
5. Flush electrode with water.
6. Press [OK] to jump to the menu "buffer value 2".
Buffer 2
1. Select buffer value 2.
2. Fill buffer solution 2 into a clean vessel.
3. Immerse the electrode.
4. The measured signal is read in and compared with the buffer
values.
– The calibration result (sensor slope and asymmetry
potential) is displayed.
5. Discard buffer solution 2.
6. Flush the electrode with water, and return to the measuring
cell.
7. Go back to the menu "calibration" with [ESC].
Then calibrate the chlorine value.
Calibrating the chlorine value
DIP: Push the button underneath the left-hand display.
Conex DIA and DIP: Select the calibration function with the
[CAL] button. The LED illuminates.
1. Press [OK] to select the menu "calibration".
2. Select "chlorine", and confirm with [OK].
3. Select "cal. meas. value".
4. Enter the reference value.
– The measured signal is read in automatically and compared
with the reference value.
– The calibration result (sensor slope) is displayed.
Reading calibration data in the diagnostics menu "service"
Select "service" in the "main menu" with [UP]/[DOWN], and
press [OK] (DIP: on both displays).
– The data of the last 10 calibrations can be read.
10.8.3 Calibrating the parameters chlorine, chlorine dioxide
or ozone with Conex DIS-D
1. Press the [CAL] button to select the calibration menu.
– The CAL LED next to the [CAL] button illuminates.
– The display shows the current measured value.
– To prevent overdosing, the controllers must be switched off,
and the device controlled must be closed.
2. Use [UP]/[DOWN] to select the reference value.
3. Start the calibration with [OK].
– The sensor data are then read in.
– The slope (sensitivity) of the sensor is calculated.
– The calibration result is displayed immediately after
calibration.
– The sensor slope is displayed in µA/ppm.
Calibration result
The result of the current (last) calibration can be displayed in
the code menu at any time:
– Code 51: display of calibrated slope in µA/ppm.
Error message when reading in the current signal of the
sensor system
The alarm LED flashes.
The display shows code *13* (slope error).
The alarm is triggered, if the plausibility check results in an
upward or a downward violation of the following slope ranges,
depending on the selected measuring cell:
Press [OK] to acknowledge the error message and to return to
the display level.
– The calibration data are imported.
CALDATA pH
slope
- 60.17 mV/pH
asym.pot
19 mV
chlorine
cal. meas. value
cal. result
cal. cycle
CALDATA chlorine
slope
34.67 µA/ppm
Measuring cell Lower limit Upper limit
AQC-D12 2.5 µA/ppm 70.0 µA/ppm
Note
The controller is operating in emergency mode!
Eliminate the fault, see section 11.5 Fault finding,
and calibrate again!
English (GB)
18
11. Operation
11.1 Function
Various oxidation agents are used for the disinfection of
swimming-pool water and drinking water, for instance chlorine
(Cl
2
), chlorine dioxide (ClO
2
) and ozone (O
3
).
It is necessary to measure the concentration of the oxidation
agent and to regulate its dosing.
Concentration of oxidation agent is too low:
– Disinfection effect is too weak.
Concentration of oxidation agent is too high:
– Danger to health
– Unpleasant odour and taste
– Corrosion damage
– Increased operating costs.
The AQC-D12 measuring cell is used for measuring the
concentration of chlorine (Cl
2
), chlorine dioxide (ClO
2
) or ozone
(O
3
). It is equipped with a water sensor, an optional Pt100
temperature sensor and additional holders for pH electrodes and
redox electrodes.
Preassembled systems with Conex DIA, Conex DIS and DIP are
used when the values of decontamination agents, pH and
redox-potential have to be determined and controlled. The basic
element of the preassembled systems is the AQC-D12 measuring
cell.
11.1.1 Functional principle of measurement
Sample water is taken at a representative position and passed
to the measuring cell via an integral filter.
– The water flow rate can be adjusted on the measuring cell.
– A water sensor serves to trigger an alarm or to switch off the
control function, if there is a sample water deficiency. It is
also used for checking the cleaning wing.
– The agent to be determined (Cl
2
, ClO
2
or O
3
) is measured at
the gold or platinum electrode.
The agent to be determined generates an electric current in
the µA range, proportional to its concentration.
– The measuring cell is controlled by a potentiostat in the
measuring amplifier.
– A constant voltage is applied to the measuring electrode. An
exactly defined potential of the measuring electrode is
retained by means of the reference electrode. This results in
a linear response for the measuring cell as well as a stable
zero point for the measurement.
The Conex or DIP measuring amplifiers and controllers
– amplify the current
– calculate it using the calibration parameters
– display the concentration as a digital value
– control a gas dosing unit or a dosing pump (device
controlled).
11.1.2 Electrode cleaning
The measuring electrode and the counter-electrode are
continuously cleaned of deposits by a cleaning wing.
This ensures uniform sensitivity for the measuring cell over a
long period.
The cleaning wing is driven hydraulically.
11.1.3 Influence of temperature
The current generated on the electrodes depends on the
temperature of the sample water.
The measured value increases by approx. 4 % per 1 °C
increase.
Temperature variations can be compensated by the measuring
amplifier, if the temperature compensation function is
activated.
– The temperature can be measured with a Pt100 temperature
sensor optionally integrated into the measuring electrode.
– The temperature measuring signals are transferred to the
measuring amplifier and calculated using the electrode
signals.
11.1.4 Influence of the pH value on the chlorine measurement
Free, active chlorine is considered as the total of molecular
chlorine gas (Cl
2
), hypochlorous acid (HOCl) and hypochlorite
anions (OCl
-
). Dissolved chlorine gas in molecular form
practically does not exist at the application-dependent pH values
(pH 4.5 to 8.2), but is subject to hydrolysis in the presence of
water according to the equation
Cl
2
+ H
2
O HOCl + HCl
The resulting hypochlorous acid is the actually effective
compound for disinfection of the water. The dissociation of acid to
anions is primarily according to an equilibrium dependent on the
pH value according to the equation
HOCl + H
2
O H
3
O
+
+ OCl
-
Fig. 23 HOCl-OCl diagram
The measuring cell for chlorine primarily only determines the
portion of hypochlorous acid that is relevant for disinfection.
Since the dependence on the pH value is particularly large in the
actually relevant range between pH 7 and pH 8.2, this
dependence must be compensated for in the measuring amplifier
if the pH value varies considerably. The pH value must therefore
also be measured, and the pH compensation function must be
activated on the measuring amplifier.
TM03 5879 1106
4
100
% HOCl
80
60
40
20
0
0
20
40
60
80
100
% OCl
-
pH
0 °C
10 °C
30 °C
OCl
-
HOCl
5
6
7
89
10 11
English (GB)
19
11.2 Operation
11.2.1 Switching on
See section
10.7 Switching on.
11.2.2 Operating state
11.2.3 Adjusting / setting the quantity of sample water
A correct measurement can only be taken, if the quantity of
sample water is within a defined range. This range is from 200 to
800 rotations/minute of the cleaning wing.
Set the quantity of sample water with the regulation (G) so that
the measuring amplifier shows a sample water flow rate of 200
to 800 rotations/minute. The bar in the display of the
measuring amplifier shows the range. The exact speed can be
taken from the service menu.
11.3 Switching off
11.3.1 Short-term stop (up to one week)
1. Close and/or switch off the device controlled.
2. Wait until the display of the measuring amplifier is at zero.
3. Switch off the electricity supply to the measuring amplifier.
4. Close the connection to the sample water inlet (E) and outlet
(F).
11.3.2 Long-term stop
1. Close and/or switch off the device controlled.
2. Wait until the display of the measuring amplifier is at zero.
3. Switch off the electricity supply to the measuring amplifier.
4. Close the connection to the sample water inlet (E) and outlet
(F).
5. Open the sampling cock (I), and drain water.
6. Remove the measuring electrode (J), clean and store dry.
7. Remove the reference electrode (K), and remove the pH
electrode (A) and redox electrode (B), if applicable. Fit thread
protection caps, and clean the electrodes.
8. Fit protective caps on the electrodes. The protective caps
must be filled with 3-molar potassium chloride solution.
9. Store electrodes with protective caps in a dry place. If they are
stored for an extended period of time, it may be necessary to
add 3-molar potassium chloride solution after a while.
10. Before starting up again, and if it is very dirty, clean the whole
measuring cell. See section
12. Maintenance.
11.4 Switching on again
11.4.1 After short-term stop
1. Open the connection to the sample water inlet (E).
2. Open the connection to the sample water outlet (F).
3. Check the quantity of sample water.
4. Switch on the measuring amplifier.
5. Take a reference measurement.
6. Recalibrate, if necessary. See section
10.8 Calibrating the
parameters Cl
2
, ClO
2
, O
3
.
7. Switch on the device controlled, if applicable.
11.4.2 After long-term stop / inspection
If the measuring cell has been switched off for a long period of
time or has been emptied, for instance for inspection, it is
necessary to start up the measuring cell. See section
10.6 Checks prior to start-up and 10.7 Switching on.
Caution
Observe the installation and operating instructions of
the measuring amplifier and the device controlled!
Note
For information on operation of the measuring cell in
the operating state, please see the installation and
operating instructions of the relevant measuring
amplifier.
Caution
The measuring cell may be damaged if the reference
electrode (K) is not connected, and the electricity
supply is switched on!
Caution
Observe the installation and operating instructions of
the measuring amplifier and controller!
Note
The delay time of the measuring cell can be reduced
by increasing the sample water flow rate.
Caution
To avoid a dosing error, the device controlled must
be switched off each time the measuring system is
switched off.
Observe the installation and operating instructions of
the measuring amplifier and the device controlled!
Caution
To protect the electrodes, do not drain the water!
Make sure that the water level in the measuring cell
is sufficiently high for the reference electrode (K) to
be immersed at least 2 cm during device stop.
Caution
Observe the installation and operating instructions of
the electrodes!
Caution
Observe the installation and operating instructions of
the measuring amplifier and the device controlled!
Caution
Let the system run for at least two hours to avoid
faulty measurements and calibrations!
English (GB)
20
11.5 Fault finding
Caution
Observe the installation and operating instructions of
the measuring amplifier and controller!
Fault Cause Remedy
1. No display. No electricity
supply to the measuring
cell.
a) A disinfection or oxidation agent is missing in the
sample water.
Check the concentration by making a reference
measurement. Check the dosing units, and
activate dosing. Check the settings of the
measuring amplifier.
b) The electrode cable connection is interrupted. Remake the connection. Replace damaged
cables.
c) No or too low sample water flow rate. Clean the filter, check the flow rate, and check
the sample water inlet pressure.
d) The measuring electrode is faulty. Replace the measuring electrode.
e) The measuring electrode is contaminated or made
passive by deposits.
Dismantle, clean or replace the measuring
electrode.
2. The measured value is
lower than the reference
measurement.
a) The measuring electrode is faulty. Replace the measuring electrode.
b) The measuring electrode is contaminated or made
passive by deposits.
Remove, clean or replace the measuring
electrode.
c) The temperature has fallen since the calibration. Activate the automatic temperature
compensation, if applicable. Recalibrate.
d) The system has run too short time prior to
calibration.
Let the system run for at least two hours.
Recalibrate.
e) Unsuitable chlorination agent. Only use chlorine solutions such as chloric gas,
sodium hypochlorite and chlorinated lime. Do
not use organic products such as
trichloroisocyanuric acid.
f) Incorrect reference measurement during
calibration.
Check the reference measurement, and take
another measurement. Recalibrate.
g) The pH value for the chlorine measurement has
risen since the calibration.
Keep the pH value constant. Activate the pH
value compensation, if applicable. Recalibrate.
h) The quantity of sample water is set too low. Check and set the quantity of sample water.
Check the filter, and clean it, if necessary.
i) The cleaning wing is blocked. Fit the cleaning wing correctly.
3. The measured value is
higher than the reference
measurement.
a) The temperature has risen since the calibration. Activate the automatic temperature
compensation, if applicable. Recalibrate.
b) The pH value for the chlorine measurement has
fallen since the calibration.
Keep the pH value constant. Activate the pH
value compensation, if applicable. Recalibrate.
c) Interference by other oxidation agents in the
sample water.
Analyse the sample water. Check the chemicals
used.
4. The measured value is
unstable.
a) Interferences on the signal lines. Check the screen and the connections to the
amplifier.
b) The reference electrode diaphragm is blocked. Clean the diaphragm (on the side at the bottom
of the reference electrode) using diluted
hydrochloric acid (10 %). Replace the reference
electrode, if necessary.
c) The measuring electrode is contaminated or faulty. Clean the measuring electrode using a diluted
abrasive cleaning agent. Replace measuring
electrode, if necessary.
d) The filter is contaminated, and the quantity of
sample water is therefore too low.
Check the filter, and clean it, if necessary.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268

Grundfos AQC-D12 Installation And Operating Instructions Manual

Tip
Installation And Operating Instructions Manual

în alte limbi