Ambu aBox™ 2 Manualul proprietarului

Tip
Manualul proprietarului
Ambu® aBox 2
SPARE PART
REPLACEMENT
MANUAL
For use by qualified technicians only.
Contents Page
English (Instructions for use) ................................................................................................................................1-13
Български (Инструкция за употреба) .........................................................................................................14-26
Čeština (návod k použití).....................................................................................................................................27-37
Dansk (Brugsvejledning) .................................................................................................................................... 38-48
Deutsch (Bedienungsanleitung) ..................................................................................................................... 49-60
Ελληνικά (Οδηγίες χρήσης) ................................................................................................................................ 61-72
Español (Instrucciones de uso) ........................................................................................................................ 73-84
Eesti (Kasutusjuhend) ...........................................................................................................................................85-95
Suomi (Käyttöohjeet) .........................................................................................................................................96-106
Français (Mode d'emploi) ................................................................................................................................107-117
Hrvatski (Upute za upotrebu) ........................................................................................................................118 -128
Magyar (Használati útmutató) ..................................................................................................................... 129-139
Italiano (Istruzioni per l'uso)..........................................................................................................................140-150
日本(使用説明) ............................................................................................................................................... 151-161
Lietuvių (naudojimo instrukcija) ..................................................................................................................162-173
Latviešu val. (Lietošanas instrukcija) ..........................................................................................................174-185
Nederlands (Gebruiksaanwijzing) ..............................................................................................................186-198
Norsk (Bruksanvisning) ...................................................................................................................................199-210
Polski (Instrukcja użytkowania) ....................................................................................................................211-223
Português (Instruções de utilização) .........................................................................................................224-236
Română (Instrucțiuni de utilizare) .............................................................................................................. 237-248
Slovensky (Návod na použitie) .....................................................................................................................249-260
Slovenščina (Navodila za uporabo) ............................................................................................................ 261-272
Svenska (bruksanvisning)...............................................................................................................................273-284
Türkçe (Kullanma Talimatı) .............................................................................................................................285-296
󰱛󳜂(󰹉󵌡󸄝󳟼) ...................................................................................................................................................297-307
2
Content Page
1. Important Information ............................................................................................................ 4
1.1. Qualification Requirements for Personnel Performing Spare Part Replacement ...............4
1.2. Warnings and Cautions ..............................................................................................................................4
1.3. Data Security .................................................................................................................................................. 5
1.4. System Description .......................................................................................................................................5
1.5. Spare Parts........................................................................................................................................................6
1.6. Spare Part Replacement Process .............................................................................................................7
2. How To Clean and Disinfect the Displaying Unit .................................................................. 7
3. Preparing for Spare Part Replacement .................................................................................. 9
3.1. Tools Needed ..................................................................................................................................................9
4. How to Replace Spare Parts .................................................................................................... 9
4.1. Visualization Device Interface Board Replacement ...................................................................... 10
5. Final Inspection and Test ....................................................................................................... 13
3
1. Important Information
This is a Spare Part Replacement Manual for Ambu® aBox™ 2. It is applicable to aBox 2 with
item number 505001000 only.
The following terms are used in this Spare Part Replacement Manual:
“Displaying unit” refers to Ambu® aBox™ 2.
“Visualization device” refers to compatible Ambu endoscopes and other compatible
visualization devices that can be connected to and used with the displaying unit.
This manual may be updated without further notice. Contact your local Ambu representative
or download the most recent version from ambu.com found om the download section on the
product page.
Please observe that the warranty will be void, if parts of the displaying unit has been changed
within the warranty period without the written consent from Ambu.
1.1. Qualification Requirements for Personnel Performing Spare
Part Replacement
Do not start replacing any spare part of the displaying unit, unless you have read
through this document and are qualified in the following areas:
Knowledge, experience and acquaintance with electronic repair and safety testing
according to IEC 60601-1 (Medical electrical equipment, Part 1: General requirements for
basic safety and essential performance) and IEC 62353 (Medical electrical equipment –
Recurrent test and test after repair of medical equipment).
Knowledge of and experience with local regulations.
Having read the Instruction for Use for the displaying unit (find the Instruction for Use in
the sales package or download the current version from ambu.com).
Knowledge of the environment where the displaying unit is installed and used.
1.2. Warnings and Cautions
WARNINGS
To avoid risk of electric shock, always disconnect the power cable from the
displaying unit before performing spare part replacement.
To avoid contamination, always clean and disinfect the displaying unit before and
after spare part replacement and wear gloves during handling.
CAUTIONS
To avoid harming the electrical components, always use ESD protection when
disassembling, replacing a spare part, and reassembling the displaying unit.
To ensure the displaying unit functions optimally and to avoid compromising electrical
safety, always perform the tests specified in section 5 after it has been reassembled.
If any part of the tests fails after spare part replacement, do not use the displaying unit.
To avoid malfunction of the equipment, only use spare parts supplied by Ambu.
Do not modify the spare parts.
Cleaning and disinfection wipes shall be moist, but not dripping to ensure no
damage to internal electronics of the displaying unit.
If using wipes containing hypochlorite or citric acid for cleaning, ensure that all
residue is completely removed. Wipes containing hypochlorite or citric acid may
affect the screen's antireflective coatings over time. You should limit the use of wipes
containing hypochlorite or citric acid to required cases only.
4
en
5
1.3. Data Security
To avoid possible loss of data, it is recommended to export important files before performing
any spare part replacement (please refer to the Instruction for Use).
1.4. System Description
The displaying unit is intended to display live imaging data from Ambu visualization devices.
7
8
19181716
14 15131210 11
1
23
45
6
9
1




3
2
4
Wi-Fi

Figure 1 System illustration
No. Part Function
1Touch Screen Displays the image from the Ambu visualization
device and the graphical user interface.
2
VDI port (connector port
for specific Ambu
visualization devices)
Connector port geometry and colour ensures
correct connection with compatible
visualization devices.
3USB 3.0 port (front) Enables connection of external USB flash drives.
4USB cover Protects the front USB port.
5Power button Turns the power ON or switches to STAND-BY mode.
6Base Contains the main unit.
7Power cable Connects the displaying unit to a power outlet.
8Wi-Fi antenna Connect Wi-Fi antenna to the displaying unit for
improved Wi-Fi signal.
9Positioning arm
Enables manual positioning of the touch screen.
The screen can be adjusted horizontally and
vertically as well as rotated.
10 Video output ports
(2 x DVI-D) Enable connection to external monitor or
external medical imaging recorder.
11 Video output ports
(2 x 3G-SDI)
12 Wi-Fi antenna connector Enables connection of Wi-Fi antenna.
13 USB 3.0 ports Enable connection of external USB flash drives.
14 LAN port Enables connection to ethernet.
15 USB 2.0 ports Enable connection of external USB flash drives.
16 Trigger output ports
(2 x 3.5 mm jack) Enable connection to an external medical imaging
recorder to transfer trigger signals.
17 Trigger output ports
(2 x D-SUB9)
18 Power inlet Enables connection to power cable.
19 Connector for potential
equalization cable
Used for bonding of electrical products to eliminate
potential differences between conductive parts.
1.5. Spare Parts
The spare parts may not be available in all countries. Please contact your local Ambu representative.
CAUTIONS
To avoid malfunction of the equipment, only use spare parts supplied by Ambu.
Do not modify the spare parts.
6
7
en
Spare Part Description Item numbers
Thickness: 4 pt
Ambu® aBox™ 2 – Visualization device
interface kit – Grey-Empty-Green
Contains:
One grey and one green visualization
device interface board (VDI), a front cover
with a colour ring (grey and green), a
plectrum tool, and two screws for the VDI.
505000530
Power cables – J (CH), K (DK), I (AUS/NZ)
Power cable – B (US, JP)
Power cables – G (UK), E/F (EU, not DK, CH)
505000520
505000521
505000522
1.6. Spare Part Replacement Process
This section specifies a spare part replacement process which is in compliance with IEC 62353.
Read these Instructions for Use carefully before performing any replacements on the
Ambu® aBox™ 2.
The technician and personnel being responsible for the replacement process should fulfil
the requirements outlined in section 1.1.
Follow the process in Figure 2 to replace spare parts in the displaying unit.
Contact Ambu, if
you need further
assistance
Use the Trouble
Shooting guide in
the IFU
Order and receive
relevant parts to be
replaced.
Chapter 2:
Clean and disinfect
the displaying unit
Chapter 3:
Ensure you have
the tools needed
Chapter 5:
Inspect the
displaying unit
Chapter 2:
Clean and disinfect
the displaying unit
Follow the standard operating procedures
required by your organization
Chapter 4:
Replace the
relevant part(s) of
the displaying unit
Figure 2 Spare part replacement flow
2. How To Clean and Disinfect the Displaying Unit
CAUTIONS
Cleaning and disinfection wipes shall be moist, but not dripping to ensure no
damage to internal electronics of the displaying unit.
If using wipes containing hypochlorite or citric acid, ensure that all residue is
completely removed. Wipes containing hypochlorite or citric acid may affect the
screen's antireflective coatings over time. You should limit the use of wipes
containing hypochlorite or citric acid to required cases only.
The displaying unit is a reusable medical device. According to the Spaulding classification,
the displaying unit is a non-critical device.
The displaying unit should be cleaned and disinfected before and after repair and replacement
procedures following one of the procedures below.
Any deviation from the instructions should be properly evaluated for effectiveness and
potential adverse consequences by the person responsible for cleaning and disinfection,
to ensure that the device continues to fulfil its intended purpose. Cleaning procedures should
begin as soon as possible following use. Excessive soil should be removed on accessible areas
of the device, excluding electrical ports.
Limitations: The displaying unit is not compatible with ultrasonic or automatic cleaners and
should not be immersed.
Procedure 1 – Cleaning and disinfection with hypochlorite
Hypochlorite-based wipes approved for disinfection of medical devices, e.g., Sani-Cloth®
Bleach from PDI, should be used in accordance with the wipe manufacturer’s instructions.
Cleaning: Use a wipe to remove heavy soil. All blood and other body fluids must be
thoroughly cleaned from surfaces and objects. Inspect the displaying unit for cleanliness,
function and integrity before disinfection by germicidal wipe. If visible soil remains,
reclean the displaying unit.
Disinfecting:
1. For heavily soiled surfaces, use a wipe to pre-clean the displaying unit prior
to disinfecting.
2. Unfold a clean wipe and thoroughly wet the surface of the displaying unit.
3. The treated surfaces must remain visibly wet for a full four (4) minutes (or the time
recommended by the manufacturer of the disinfectant, at least 4 minutes).
Use additional wipes if needed to assure continuous 4 minutes wet contact time.
4. Let the displaying unit air dry.
Procedure 2 – Cleaning and disinfection with quaternary ammonium compounds
Wipes containing a mixture of quaternary ammonium compounds and isopropyl alcohol
approved for disinfection of medical devices, e.g., Super Sani-Cloth® from PDI, should be
used in accordance with the wipe manufacturer’s instructions.
Cleaning: Use a wipe to remove heavy soil. All blood and other body fluids must be
thoroughly cleaned from surfaces and objects. Inspect the displaying unit for cleanliness,
function, and integrity before disinfection by germicidal wipe. If visible soil remains,
reclean the displaying unit.
Disinfecting:
1. For heavily soiled surfaces, use a wipe to pre-clean the displaying unit prior to disinfecting.
2. Unfold a clean wipe and thoroughly wet the surface of the displaying unit.
3. The treated surfaces must remain visibly wet for a full two (2) minutes (or the time
recommended by the manufacturer of the disinfectant, at least 2 minutes).
Use additional wipes if needed to assure continuous 2 minutes wet contact time.
4. Let the displaying unit air dry.
Procedure 3 – Enzymatic detergent cleaning and alcohol disinfection
Cleaning:
1. Prepare a cleaning solution using a standard enzymatic detergent prepared
per manufacturers recommendations. Recommended detergent: enzymatic,
mild pH: 7–9, low foaming (Enzol or equivalent).
2. Soak a sterile gauze in the enzymatic solution and make sure that the gauze is moist
and not dripping.
3. Thoroughly clean the button, external side of rubber covers, screen, external casing
of the monitor with the moist gauze. Avoid getting the displaying unit wet to
prevent damaging internal electronic components.
4. Wait for 10 minutes (or the time recommended by the manufacturer of the
detergent) to allow the enzymes to activate.
8
9
en
5. Wipe the displaying unit clean using sterile gauze that has been moistened with
RO/DI water. Ensure all traces of the detergent are removed.
6. Repeat steps 1 to 5.
Disinfecting: Wipe the surfaces of the displaying unit for approximately 15 minutes using a
piece of sterile gauze moistened with the alcohol mixture indicated below (approximately
once every 2 minutes). Follow safety procedures for the handling of isopropyl. The gauze
should be moist and not dripping since liquid can affect the electronics inside the
displaying unit. Pay close attention to the button, external side of rubber covers, screen,
external casing and slots and gaps on the displaying unit. Use a sterile cotton swab for
these areas. Solution: Isopropyl (alcohol) 95%; Concentration: 70–80%; Preparation: 80cc
of 95% Isopropyl (alcohol) added to 20cc of purified water (PURW). Alternatively, use EPA-
registered hospital disinfection wipes containing at least 70% isopropyl. Safety precautions
and directions of use of the manufacturer must be followed.
NOTE: After cleaning and disinfection, the displaying unit must be submitted to the pre-check
procedure in chapter 5.
3. Preparing for Spare Part Replacement
3.1. Tools Needed
Plectrum tool
Screwdriver (Phillips #0) – Not supplied
Compatible (and functioning) Visualization Device – Not supplied
The plectrum tool is provided with the spare parts kits (see section 1.5 Spare Parts).
Thickness: 2 pt
Figure 3 Plectrum tool
4. How to Replace Spare Parts
This section covers the replacement of the Visualization Device Interface board (VDI board).
WARNINGS
To avoid risk of electric shock, always disconnect the power cable from the
displaying unit before performing spare part replacement.
CAUTIONS
To avoid harming the electrical components, always use ESD protection when
disassembling, replacing a spare part, and reassembling the displaying unit.
To avoid malfunction of the equipment, only use spare parts supplied by Ambu.
Do not modify the spare parts.
4.1. Visualization Device Interface Board Replacement
Ambu® aBox™ 2 Visualization Device
Interface Kit – Grey-Empty-Green 505000530
Grey Visualization Device Interface board
(VDI board)
Green Visualization Device Interface board
(VDI board)
Front cover, incl. colour ring (grey) and
USB cover
Thickness: 2 pt
Screws for VDI board (x2)
Thickness: 2 pt
Plectrum tool
Thickness: 2 pt
Additional Tools (not supplied)
Screwdriver (Phillips #0)
IMPORTANT:
Ensure all cables and equipment are disconnected from the displaying unit before
initiating the replacement procedure!
The pictures are for illustration purposes. The number, color and position of front connectors
may differ. Please make sure to install the exchanged parts at the original positions.
a) Detach the front cover from the displaying unit
Locate the groove on one of the sides of the front cover where a snap feature is located
on the inside. Press inwards with the plectrum tool and a finger to detach the front cover
from one side.
NOTE: Be careful not to press on the top or bottom of the displaying unit, as it will
complicate removal of the front cover.
10
11
en
b) Completely remove the front cover by pulling it carefully from the displaying unit
c) Remove the 2 screws holding each VDI board using a Phillips screwdriver
d) Pull out the VDI boards from the displaying unit and dispose them according to
local guidelines for disposal of electronic waste
e) Carefully insert the new VDI boards into their original positions
Ensure that the VDI boards are oriented correctly as shown on the picture below.
NOTE: Insert the VDI boards straight and horizontally into the original main board
connector positions to ensure correct connection of the VDI board to the displaying unit.
THIS SIDE UP
THIS SIDE UP
f) Carefully fasten the VDI boards with the supplied screws using the Phillips
screwdriver into their original positions
g) Attach the new front cover supplied with the replacement kit
Start by attaching from one end. Ensure that the snap features on each end is placed
inside the housing of the displaying unit. The front cover is pressed in place until all top
and bottom snap features click in place.
NOTE: Ensure that the colour rings are fitted correctly in the front cover and that colours
and arrows on the front cover and the displaying unit match.
NOTE: Ensure the USB cover flap is inserted underneath the USB port.
12
13
en
5. Final Inspection and Test
CAUTIONS
To ensure the displaying unit functions optimally and to avoid compromising electrical
safety, always perform the tests specified in section 5 after it has been reassembled.
If any part of the tests fails after spare part replacement, do not use the displaying unit.
Perform the following inspections and tests after assembly prior to use:
a) Examine displaying unit for visual damage
Closely examine the displaying unit and other parts for any damage I. Do not use the
displaying unit if it is damaged in any way.
b) Visual inspection with a functioning scope
Connect the displaying unit to a suitable power outlet via the power cable II .
Connect a functioning visualization device to the displaying unit by plugging it into the
visualization device connector (VDC) port III . Ensure that the connector colours are
matching, and the arrows are aligned. Turn on the displaying unit on the power button.
Verify that a live video image appears on the screen by pointing the distal end of the
Ambu visualization device towards an object, e.g. the palm of your hand, when the Live
View tab is selected in the toolbar to the left IV.
I II
III IV
For other relevant testing of the displaying unit, refer to the requirements of IEC 62353 or
other relevant safety tests according to local guidelines.
Съдържание Страница
1. Важна информация ............................................................................................................. 15
1.1. Изисквания за квалификация за персонала, извършващ подмяна на резервни части
... 15
1.2. Предупреждения и предпазни мерки .......................................................................................... 15
1.3. Защита на данните .................................................................................................................................. 16
1.4. Описание на системата ......................................................................................................................... 16
1.5. Резервни части .......................................................................................................................................... 18
1.6. Процедура по подмяна на резервни части ................................................................................ 19
2. Как се почиства и дезинфекцира дисплейният модул ................................................ 19
3. Подготовка за подмяна на резервни части .................................................................... 21
3.1. Необходими инструменти ................................................................................................................... 21
4. Как да се подменят резервните части ............................................................................. 21
4.1. Подмяна на интерфейсната платка на устройството за визуализиране ..................... 22
5. Финална инспекция и тест ..................................................................................................25
14
bg
1. Важна информация
Това е ръководство за подмяна на резервни части за Ambu® aBox™ 2. То е приложимо
само за aBox 2 с номер на артикул 505001000.
В ръководството за подмяна на резервни части са използвани следните термини:
„Дисплеен модул“ се отнася за Ambu® aBox™ 2.
„Устройство за визуализиране“ се отнася за съвместими ендоскопи Ambu и други
съвместими устройства за визуализиране, които могат да бъдат свързани и
използвани с дисплейния модул.
Това ръководство може да се актуализира без допълнително предизвестие. Свържете се
с местния представител на Ambu или изтеглете последната версия от ambu.com от
раздела за изтегляне на продуктовата страница.
Имайте предвид, че гаранцията ще се анулира, ако части на дисплейния модул бъдат
разглобени по време на гаранционния период без писмено съгласие от Ambu.
1.1. Изисквания за квалификация за персонала, извършващ
подмяна на резервни части
Не започвайте да подменяте никакви резервни части на дисплейния модул, освен ако
не сте прочели този документ докрай и не сте квалифицирани в следните области:
Знания, опит и познаване поправката на електроника и тестването за безопасност
съгласно IEC 60601-1 (Електромедицински апарати, част 1: Общи изисквания за
основна безопасност и съществени характеристики) и IEC 62353
лектромедицински апарати – Периодично изпитване и изпитване след ремонт на
електромедицински апарати).
Знания относно и опит с местните разпоредби.
Прочели сте инструкциите за употреба на дисплейния модул (намерете инструкциите
за употреба в пакета за продажба или изтеглете текущата версия от ambu.com).
Познания за средата, където дисплейният модул е инсталиран и се използва.
1.2. Предупреждения и предпазни мерки
ПРЕДУПРЕЖДЕНИЯ
За да избегнете риска от електрически удар, винаги разкачвайте захранващия
кабел от дисплейния модул преди извършване на подмяна на резервни части.
За да избегнете контаминация, винаги почиствайте и дезинфекцирайте
дисплейния модул преди и след подмяна на резервни части, а също така и
носете ръкавици по време на работата.
15
УКАЗАНИЯ ЗА ВНИМАНИЕ
За да избегнете нараняване на електрическите компоненти, винаги използвайте
защита от ЕСР, когато разглобявате, подменяте резервна част и повторно
сглобявате дисплейния модул.
За да гарантирате, че дисплейният модул функционира оптимално, и за да
избегнете нарушаване на електрическата безопасност, винаги извършвайте
тестовете, указани в раздел 5, след като е сглобен отново. Ако някоя част от
тестовете не е сполучлива след подмяна на резервна част, не използвайте
дисплейния модул.
За да се избегне неизправност на оборудването, използвайте само резервни
части, предоставени от Ambu. Не модифицирайте резервните части.
Кърпичките за почистване и дезинфекция трябва да са влажни, но да не капят, за да
се гарантира, че няма да повредят вътрешната електроника на дисплейния модул.
Ако използвате кърпички, съдържащи хипохлорит или лимонена киселина за
почистване, се уверете, че всички остатъци са напълно отстранени.
Кърпичките, съдържащи хипохлорит или лимонена киселина, може да засегнат
антирефлексното покритие на екрана с течение на времето. Трябва да
ограничавате използването на кърпички, съдържащи хипохлорит или лимонена
киселина, само за случаите, в които такива се изискват.
1.3. Защита на данните
За да избегнете евентуална загуба на данни, е препоръчително да експортирате важните
файлове, преди да извършите подмяна на някоя резервна част (направете справка с
  ).
1.4. Описание на системата
Дисплейният модул е предназначен за показване на образни данни в реално време от
устройства за визуализиране на Ambu.
16
bg
7
8
19181716
14 15131210 11
1
23
45
6
9
1




3
2
4
Wi-Fi

 1   
Част Функция
1Сензорен екран
Показва изображението от устройството
за визуализиране на Ambu и графичния
потребителски интерфейс.
2
VDI порт (конекторен порт за
конкретни устройства за
визуализиране на Ambu)
Геометрията и цветът на конекторния порт
гарантират правилно свързване със
съвместими устройства за визуализиране.
17
Част Функция
3USB 3.0 порт (преден) Позволява свързване на външни USB
флаш устройства.
4Капак за USB Защитава предния USB порт.
5Бутон на захранването Включва захранването или превключва на
режим В ГОТОВНОСТ.
6Основа Съдържа основния модул.
7Захранващ кабел Свързва дисплейния модул към
електрически контакт.
8Wi-Fi антена Свържете Wi-Fi антената с дисплейния
модул за подобрен Wi-Fi сигнал.
9Позициониращо рамо
Позволява ръчно позициониране на
сензорния екран. Екранът може да се
регулира хоризонтално и вертикално,
както и да се завърта.
10 Портове за видеоизход (2 × DVI-D) Позволяват свързване към външен
монитор или външно устройство за запис
на медицински изображения.
11 Портове за видеоизход (2 × 3G-SDI)
12 Конектор за Wi-Fi антена Позволява свързване на Wi-Fi антена.
13 USB 3.0 портове Позволяват свързване на външни USB
флаш устройства.
14 LAN порт Позволява свързване с Ethernet.
15 USB 2.0 портове Позволяват свързване на външни USB
флаш устройства.
16 Портове за активаторен изход
(2 × 3,5 mm жак) Позволяват свързване към външно
устройство за запис на медицински
изображения за прехвърляне на
активаторни сигнали.
17 Портове за активаторен изход
(2 × D-SUB9)
18 Вход за захранване Позволява свързване към захранващ кабел.
19 Конектор за кабел за
изравняване на потенциала
Използва се за свързване на електрически
продукти за елиминиране на разлики на
потенциала между проводимите части.
1.5. Резервни части
Резервните части може да не са налични във всички държави. Свържете се с вашия
местен представител на Ambu.
УКАЗАНИЯ ЗА ВНИМАНИЕ
За да се избегне неизправност на оборудването, използвайте само резервни
части, предоставени от Ambu. Не модифицирайте резервните части.
18
bg
Резервна част Описание Номера на
артикули
Thickness: 4 pt
Ambu® aBox™ 2 – комплект с интер-
фейс на устройство за визуализиране
– сиво-празно-зелено
Съдържа:
Една сива и една зелена интерфейсна
платка за устройство за визуализиране
(VDI), преден капак с цветен пръстен
(сив и зелен), перце и два винта за VDI.
505000530
Захранващи кабели – J (Швейцария),
K (Дания), I (Австралия/Нова Зеландия)
Захранващ кабел – B (САЩ, Япония)
Захранващи кабели – G (Обединено
кралство), E/F (ЕС, без Дания, Швейцария)
505000520
505000521
505000522
1.6. Процедура по подмяна на резервни части
Този раздел описва процеса по подмяна на резервни части, който е в съответствие с IEC62353.
Прочетете внимателно тези инструкции за употреба, преди да извършите подмяна на
резервни части на Ambu® aBox™ 2.
Техникът и персоналът, отговорни за процеса по подмяна, трябва да отговарят на
изискванията, описани в раздел 1.1.
Следвайте процеса на фигура 2 за подмяна на резервни части в дисплейния модул.
Contact Ambu, if
you need further
assistance
Use the Trouble
Shooting guide in
the IFU
Order and receive
relevant parts to be
replaced.
Chapter 2:
Clean and disinfect
the displaying unit
Chapter 3:
Ensure you have
the tools needed
Chapter 5:
Inspect the
displaying unit
Chapter 2:
Clean and disinfect
the displaying unit
Follow the standard operating procedures
required by your organization
Chapter 4:
Replace the
relevant part(s) of
the displaying unit
 2      
2. Как се почиства и дезинфекцира дисплейният модул
УКАЗАНИЯ ЗА ВНИМАНИЕ
Кърпичките за почистване и дезинфекция трябва да са влажни, но да не капят, за да
се гарантира, че няма да повредят вътрешната електроника на дисплейния модул.
Ако използвате кърпички, съдържащи хипохлорит или лимонена киселина,
се уверете, че всички остатъци са напълно отстранени. Кърпичките, съдържащи
хипохлорит или лимонена киселина, може да засегнат антирефлексното
покритие на екрана с течение на времето. Трябва да ограничавате използването
на кърпички, съдържащи хипохлорит или лимонена киселина, само за случаите,
в които такива се изискват.
Дисплейният модул е медицинско устройство за многократна употреба. Според
класификацията на Spaulding дисплейният модул е устройство от некритично значение.
Дисплейният модул трябва да се почиства и дезинфекцира преди и след процедури по
ремонт и подмяна, като се следва една от процедурите по-долу.
19
Всякакви отклонения от инструкциите трябва да бъдат внимателно оценени за
ефективност и потенциални неблагоприятни последици от лицето, отговорно за
почистването и дезинфекцирането, за да се гарантира, че устройството продължава да
изпълнява предназначението си. Процедурите по почистване трябва да започнат
възможно найкоро след употреба. Прекомерните замърсявания трябва да се
отстраняват на достъпните места на изделието, с изключение на електрическите портове.
Ограничения: Дисплейният модул не е съвместим с ултразвукови или автоматични
почистващи препарати и не трябва да се потапя.
Процедура 1 – Почистване и дезинфекция с хипохлорит
Кърпичките на основата на хипохлорит, одобрени за дезинфекция на медицински
изделия, напр. белина Sani-Cloth® от PDI, трябва да се използват в съответствие с
инструкциите на производителя на кърпичките.
Почистване: Използвайте кърпа, за да премахнете големите замърсявания. Цялата
кръв и други телесни течности трябва да бъдат почистени щателно от повърхностите
и предметите. Проверете дисплейния модул за чистота, функциониране и цялост
преди дезинфекция с бактерицидна кърпа. Ако остане видимо замърсяване,
почистете отново дисплейния модул.
Дезинфекциране:
1. За силно замърсени повърхности използвайте кърпа, за да почистите
предварително дисплейния модул преди дезинфекцирането.
2. Разгънете чиста кърпа и внимателно намокрете повърхността на
дисплейния модул.
3. Третираните повърхности трябва да останат видимо мокри в продължение на
пълни четири (4) минути (или времето, препоръчано от производителя на
дезинфектанта, поне 4 минути). Използвайте допълнителни кърпи, ако е
необходимо, за да осигурите непрекъснат контакт с мократа субстанция в
продължение на 4 минути.
4. Оставете дисплейния модул да изсъхне на въздух.
Процедура 2 – Почистване и дезинфекция с четвъртични амониеви съединения
Кърпичките, съдържащи смес от четвъртични амониеви съединения и изопропилов
алкохол, одобрени за дезинфекция на медицински изделия, напр. Super Sani-Clot
от PDI, трябва да се използват в съответствие с инструкциите на производителя
на кърпичките.
Почистване: Използвайте кърпа, за да премахнете големите замърсявания. Цялата
кръв и други телесни течности трябва да бъдат почистени щателно от повърхностите
и предметите. Проверете дисплейния модул за чистота, функциониране и цялост
преди дезинфекция с бактерицидна кърпа. Ако остане видимо замърсяване,
почистете отново дисплейния модул.
Дезинфекциране:
1. За силно замърсени повърхности използвайте кърпа, за да почистите
предварително дисплейния модул преди дезинфекцирането.
2. Разгънете чиста кърпа и внимателно намокрете повърхността на дисплейния модул.
3. Третираните повърхности трябва да останат видимо мокри в продължение на
пълни две (2) минути (или времето, препоръчано от производителя на
дезинфектанта, поне 2 минути). Използвайте допълнителни кърпи, ако е
необходимо, за да осигурите непрекъснат контакт с мократа субстанция в
продължение на 2 минути.
4. Оставете дисплейния модул да изсъхне на въздух.
20
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280
  • Page 281 281
  • Page 282 282
  • Page 283 283
  • Page 284 284
  • Page 285 285
  • Page 286 286
  • Page 287 287
  • Page 288 288
  • Page 289 289
  • Page 290 290
  • Page 291 291
  • Page 292 292
  • Page 293 293
  • Page 294 294
  • Page 295 295
  • Page 296 296
  • Page 297 297
  • Page 298 298
  • Page 299 299
  • Page 300 300
  • Page 301 301
  • Page 302 302
  • Page 303 303
  • Page 304 304
  • Page 305 305
  • Page 306 306
  • Page 307 307
  • Page 308 308

Ambu aBox™ 2 Manualul proprietarului

Tip
Manualul proprietarului