LG 27HJ713SW Ghid de inițiere rapidă

Categorie
Bucătărie
Tip
Ghid de inițiere rapidă
www.lg.com
Please read this manual carefully before operating
your set and retain it for future reference.
SAFETY AND REGULATORY
INFORMATION
27HJ713S
MEDICAL
MONITOR
Copyright 2018 LG Electronics Inc. All Rights Reserved.
ENGLISH
2
Be sure to see the owner's manual provided with the CD for more information about the product and its use.
- The CD manual can be read on a computer with Adobe Reader installed.
- You can download Adobe Reader free from the Adobe website.
Do not modify or copy arbitrarily the owner's manual provided on CD.
If you need a printed owner's manual, please contact the manufacturer.
- At the request, the manufacturer provide a hard copy manual without the additional cost within 7 days.
Please note that this information is for proper use and safety of the equipment. The following symbols may indicate a
hazardous situation in which, if not heeded, may result in serious injury or even death to the user or others, or damage
to the equipment.
WARNING
Indicates warning and safety instructions. If not adhered to, it could result in death or serious
injury to the user or others.
CAUTION
Indicates a hazardous situation which, if not heeded, may result in minor or moderate injury
to the user or others, or damage to the equipment.
For users in the United States
United State federal law restricts this equipment to be used by or on the order of a physician.
For users in other countries
This equipment is to be used by or on the order of a licensed person under the related laws for each country.
WARNING
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
Do not modify this equipment without authorization of the manufacturer.
Intended use
This Medical Monitor is intended to provide color video displays and images from medical equipment which include
laparoscopy and endoscopy systems for surgery and various medical imaging systems. This product does not support
3D.
EC REP
Authorized representative in the European community.
LG Electronics European Shared Service Center B.V.
Krijgsman 1, 1186 DM Amstelveen, The Netherlands
Tel: +31-20-456-3132
LG Electronics Inc.
77, Sanho-daero, Gumi-si, Gyeongsangbuk-do, 39381, Republic of Korea
Tel: +82-1544-8777
ENGLISH
3
SAFETY INFORMATION
Safety Standard
Medical Device Classification
Classification by protection type against Electric
Shock Class equipment
Classification according to the degree of protection
against ingress of water or particulate matter
Front: IP35
Except for Front: IP32
Mode of operation Continuous Operation
Environment of Use This equipment is not suitable for use in the presence of
flammable anesthetic or oxygen.
Regulations
Safety and Electromagnetic Compatibility Information
This equipment has been tested and found to comply with the limits for medical devices in IEC 60601-1-2. These limits
are designed to provide reasonable protection against harmful interference in a typical medical installation.
This equipment generates, uses and can radiate radio frequency energy. If not installed and used in accordance with the
instructions, it may cause harmful interference to other devices in the vicinity.
However, there is no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to other devices, which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the following measures.
Reorient or relocate the equipment.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from that to which the other devices are connected.
Contact manufacturer or authorized agent for help.
Radio Frequency compliance
FCC (For USA)
FCC NOTICE
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device
may not cause harmful interference, and (2) this device must accept any interference received, including interference
that may cause undesired operation.
ENGLISH
4
FCC CAUTION
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority
to operate the equipment. This transmitter must not be co-located or operated in conjunction with any other antenna
or transmitter.
FCC WARNING
This equipment may generate or use radio frequency energy changes or modifications to this equipment may cause
harmful interference unless the modifications are expressly approved in the instruction manual.
The user could lose the authority to operate this equipment if an unauthorized change or modification is made.
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part
15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance of
this equipment in a residential area is likely to cause harmful interference in which cause the user will be required to
correct the interference at his own expense.
For Europe
DECLARATION OF CONFORMITY
The complete Declaration of Conformity may be requested through the following postal address:
LG Electronics European Shared Service Center B.V., Krijgsman 1, 1186 DM Amstelveen, The Netherlands
or can be requested at our dedicated DoC website: http://www.lg.com/global/support/cedoc/cedoc#
Electro-Magnetic Compatibility Information
Electro-Magnetic Emissions
This EUT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EUT should assure that it is used in such an environment.
Immunity Test Compliance Electromagnetic Environment Guidance
RF Emissions CISPR 11 Group 1
The EUT uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF Emissions CISPR 11 Class A
The EUT is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC
61000-3-2 A
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Complies
ENGLISH
5
Electro-Magnetic Immunity
This EUT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EUT should assure that it is used in such an environment.
Immunity Test IEC 60601-1-2
Test Level
Compliance
Level Electromagnetic Environment Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative
humidity should be at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV
100 kHz repetition
frequency
± 2 kV
100 kHz repetition
frequency
Mains power quality should be that of a
typical commercial or hospital environment.
Surge Line-to-line
IEC 61000-4-5
± 0,5 kV, ± 1 kV ± 1 kV Mains power quality should be that of a
typical commercial or hospital environment.
Surge Line-to-ground
IEC 61000-4-5
± 0,5 kV, ± 1 kV,
± 2 kV
± 2 kV
Voltage dips
IEC 61000-4-11
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the EUT image intensifier
requires continued operation during power
mains interruptions, it is recommended
that the EUT image intensifier be powered
from an uninterruptible power supply or a
battery.
0 % UT; 1 cycle
and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 1 cycle
and
70 % UT; 25/30 cycles
Single phase: at 0°
Voltage interruptions
IEC 61000-4-11
0 % UT; 250/300 cycle 0 % UT; 250/300 cycle
RATED power frequency
magnetic fields
(50/60Hz)
IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
NOTE
UT is the A.C mains voltage prior to application of the test level.
ENGLISH
6
Immunity Test IEC 60601-1-2
Test Level
Compliance
Level Electromagnetic Environment Guidance
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V
0,15 MHz 80 MHz
6 V in ISM bands
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
3 V
0,15 MHz 80 MHz
6 V in ISM bands
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
Portable and mobile RF communications equipment
should be used no closer to any part of the EUT,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
80 MHz to 800MHz
800 MHz to 2.7GHz
Where P is the maximum output power rating of the
transmitter in watts(W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters(M).
Field strengths from fixed RF transmitters as
determined by an electromagnetic site survey, should
be less than the compliance level in each frequency
range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Radiated RF EM
fields
IEC 61000-4-3
3 V/m
80 MHz 2,7 GHz
80 % AM at 1 kHz
3 V/m
80 MHz 2,7 GHz
80 % AM at 1 kHz
NOTE
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
1 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the EUT is used exceeds the applicable
RF compliance level above, the EUT should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the EUT.
2 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
ENGLISH
7
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Test
frequency
(MHz)
Band
(MHz) Service Modulation
Maximum
power
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
385 380 –390 TETRA 400 Pulse modulation 18 Hz 1,8 0,3 27
450 430 – 470 GMRS 460,
FRS 460
FM ± 5 kHz deviation
1 kHz sine 2 0,3 28
710
704 – 787 LTE Band 13,
17
Pulse modulation
217 Hz 0,2 0,3 9
745
780
810
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation
18 Hz 2 0,3 28
870
930
1 720
1 700 1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse modulation
217 Hz 2 0,3 28
1 845
1 970
2 450 2 400 2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse modulation
217 Hz 2 0,3 28
5 240
5 100 5 800 WLAN 802.11
a/n
Pulse modulation
217 Hz 0,2 0,3 95 500
5 785
NOTE
If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
ENGLISH
8
Symbols
Symbols Descriptions
Refer to instruction manual/ booklet
Alternate current
Direct current
Off (power: disconnect from the main switch)
On (power: connect from the main switch)
Warning
Caution
E486403
UL classified mark of medical equipment with respect to electrical shock, fire and mechanical
hazards only in accordance with ANSI/AAMI ES60601-1 (2005) + AMD 1 (2012), CAN/
CSA-C22.2 No. 60601-1 (2014)
Manufacturer
Date of manufacture
SN
Serial number
Non-ionizing radiation
WEEE: Waste Electrical and Electronic Equipment
EC REP
Authorized representative in the European community.
CE Marking
ENGLISH
9
Temperature limit
Humidity limitation
Pressure limitation
For the customers in the U.S.A. Caution
Federal law (United States of America) restricts this device to sale by or on the order of a
licensed healthcare practitioner.
IPN1N2Ingress of protection
Warning
WARNING: To avoid the risk of electric shock, this equipment must only be connected to a supply mains with
protective earth.
Connection
Do not connect the monitor with anything other than specified.
Otherwise, it may result in fire or electric shock.
To avoid the risk of electric shock, this monitor must only be connected to supply mains with
protective earth.
Handling
Always be sure to keep checking the condition of the system and the patient to ensure they are
normal during the use of the monitor. If any problem is found, take appropriate measures, such as
stopping the operation of the monitor, as required.
Never disassemble or modify the monitor as it may result in fire or electric shock.
Also, since the monitor incorporates parts that may cause electric shocks and other hazardous
parts, touching them may cause death or serious injury.
Do not hit or drop the monitor. The monitor may be damaged if it receives a strong jolt, which may
result in fire or electric shock if the monitor is used without being repaired.
The operator must not touch connectors of the monitor and the patient simultaneously.
The product has lower breaking capacity type. So do not install at the building power system
prospective short-circuit current exceeding 35 A.
ENGLISH
10
Caution
Do not install the ME Equipment in a location without easy disconnect accessibility.
Environment of Use and Storage
Do not install the monitor in a location with the conditions listed below.
Otherwise, it may result in failure or malfunction, cause fire or injury.
Close to facilities where water is used.
Where it will be exposed to direct sunlight.
Close to air-conditioner or ventilation equipment.
Close to heat source such as a heater.
Prone to vibration
Unsecure location place.
Dusty environment.
Saline or sulfurous environment.
High temperature or humidity.
Freezing or condensation.
Do not place the storage case in a location with the conditions listed below.
Where the cable of the monitor unit will be strongly pulled when the monitor is put into the
case, otherwise, the cable may be damaged, resulting in fire or electric shock.
Where someone might get their foot caught in the cable of the monitor.
Handling
For safety reasons, be sure to turn off the power when monitor is not used.
This monitor is contraindicated for pregnant woman.
MANUFACTURER will make available on request circuit diagrams, component part lists,
descriptions, calibration instructions, or other information that will assist SERVICE PERSONNEL to
repair those parts of ME EQUIPMENT.
Location of Cables
Make sure all cables are located so that they cannot be stepped on, tripped over, or otherwise
subjected to damage or stress.
If the monitor is defective, do not disassemble the monitor randomly. Maintenance of the monitor
should be done by a manufacturer.
ENGLISH
11
SAFETY PRECAUTIONS
The Safety Precautions are intended to prevent unexpected danger or harm by helping the user to use the product
safely and for its intended purpose.
WARNING
Failure to follow the instructions may result in serious injury or even death.
CAUTION
Failure to follow the instructions may result in minor injury to people or damage to the
product.
Precautions When Installing the Product
WARNING
Do not place the product close to heat sources such as radiators, fires, stoves, etc.
Be careful when disinfecting the product, as some disinfectants may cause fire.
Do not install the product in a damp or dusty area.
If the product emits smoke, strange odors or sounds, disconnect the power cord and contact the manufacturer.
Do not leave the power or signal cable, etc. in walkways.
Do not install the product onto a wall close to sources of oil or oil mist.
If the product has been dropped or its case is damaged, turn off the power and unplug the power cord from the
power outlet.
CAUTION
Ensure that the product is installed at least 100 mm (3.9 inches) away from the wall for good ventilation to prevent
deformation of the product or a fire due to the increase in the internal temperature.
Keep the product away from direct sunlight.
Do not install the product close to hot objects, such as lights.
Make sure the air vents are not blocked by a tablecloth or curtain.
Do not install the product near magnetic field generating equipment, such as transformers or high-voltage power
lines.
Installing the product in locations that do not meet the general conditions may seriously damage the product’s picture
quality, life cycle and appearance. Please check with our service engineer before installing in such locations.
Locations that do not meet the general conditions: places where the product is not intended for use, such as those
where fine dust or oil mist is generated, places where chemicals are used, places where the temperature is too high or
low, places where the humidity is too high, or places where the product is likely to be used for a long period of time (e.g.
airports or train stations).
ENGLISH
12
Precautions for the AC/DC Adaptor and Power
WARNING
If water or a foreign substance enters the product (power cord or AC/DC adaptor), disconnect the power cord
immediately and contact the manufacturer.
Always connect the power cord to a grounded outlet.
Do not touch the power plug or AC/DC adaptor with wet hands. If the pins of the plug are wet or dusty, clean and dry
them before use.
Insert the power plug securely so that it does not come loose. Do not use the product if the power outlet is loose.
Make sure the power cord is fully inserted into the AC/DC adaptor.
Do not insert any conductive objects into the device connection end of the power cord, while the power cord is
inserted into the wall outlet. Avoid touching the power plug immediately after pulling from the wall outlet.
Be sure to use the power cords and AC/DC adaptors provided or ones approved by LG Electronics, Inc.
Use the product at the rated voltage only.
Disconnect the power cable when not using the product for long periods.
Always pull by the plug when unplugging the power cord. Do not bend the power cord with excessive force.
Take care not to step on or place heavy objects (electronic appliances, clothing, etc.) on the power cord or AC/DC
adaptor. Be careful not to damage the power cord or AC/DC adaptor with sharp objects.
Never disassemble, repair or modify the power cord or AC/DC adaptor.
To turn off the main power, remove the power cord, which should be positioned so it is easy to access for operation.
If the product is connected to an AC wall outlet, it will not be disconnected from the AC power source even if you
turn off the switch.
CAUTION
Do not remove the power cord when the product is in use.
Be sure to keep the outlet, AC/DC adaptor and pins of the power plug clean from dust, etc.
Do not turn the monitor on or off by plugging in or unplugging the power plug from the power outlet. (Do not use the
power plug as a switch.)
Keep the power cord away from heating devices.
Precautions When Transporting the Product
WARNING
Before moving the product, disconnect the power cord and all other connected cables.
When carrying the product, make sure the screen faces forward and hold it firmly in both hands.
ENGLISH
13
CAUTION
Do not discard the delivery box with the original packaging. Put the product in the box when carrying.
When moving or unpacking the product, make sure that two or more people safely lift it because it is heavy. Using the
product after it has fallen may result in electric shock or fire. Contact the manufacturer.
Precautions When Using the Product
WARNING
Do not disassemble, repair or modify the product by yourself. If the product needs to be checked, reset or repaired,
contact the manufacturer.
Do not allow water to enter the product and keep the product dry at all times.
If there is a gas leak, do not touch the power outlet. Open the windows for ventilation.
Do not impact the front or sides of the screen with a hard object, such as a metal object. Do not scratch the screen.
If liquid or a foreign object falls into the product, switch the product off and unplug it from the wall outlet. Contact
the manufacturer.
Stop using the product if no image appears on the screen or no sound is heard. Switch the product off immediately.
Unplug the product from the power outlet and contact the manufacturer.
Do not use the product in any environment with excessively high temperatures or humidity.
Do not use high-voltage electrical products near the product. This may result in product malfunction due to electric
shock. (e.g. electrical mosquito swatter)
CAUTION
Do not use or store the product near inflammables.
Take a break to protect your health and vision when using the product for long periods of time.
Do not push hard against the screen surface or scratch it with your hands or sharp objects, such as a nail, pencil or
pen.
Keep the product clean.
Refer to the Owner's Manual and set the product to the correct resolution and frequency. Otherwise, you may
experience blurred vision.
Do not leave the product where it can get wet. Do not place anything containing liquid on top of the product, such as
a flower vase.
Because the LCD requires high technology consisting of hundreds of pixels, pixelated spots (red, green, white or
black) may appear on the screen during use. This is normal for an LCD screen and not an error, nor is it related to the
monitor's performance.
Given the nature of the viewing angle, the brightness and color of the LCD may vary between the left and right or top
and bottom, depending on the viewing position. This is normal for an LCD screen and not an error, nor is it related to
the monitor's performance.
ENGLISH
14
Precautions for Connecting to Other Medical Devices
Take the following precautions before using this product or connecting it to other medical devices.
Stop using this device for medical purposes if you experience interruptions or discomfort in executing medical
activities.
Quick movements, shaking and the focus of the video displayed on the monitor, its distance from the user, the user's
point of view, the user's physical fitness and other factors may cause discomfort (eye fatigue, dizziness, vomiting,
nausea, etc.).
Before using the product, check that the image of the connected device displays properly on the screen of this
product.
Precautions for Use with Electric Scalpels, etc.
When this product is used with electric scalpels, etc., the image may appear unstable, distorted or abnormal due to
the strong electromagnetic waves or voltage emitted by the device. This is not a malfunction.
If you are using this device with equipment that emits strong radio waves or voltage, install this device in a direction
that minimizes radio wave interference.
Recommendation for Using Multiple Units
Electronic products may fail unexpectedly. If using the monitor for medical, emergency or any other critical use, we
strongly recommend that you use multiple monitors or have a spare monitor ready.
Precautions When Cleaning the Product
The medical LCD monitor's front protective panel has been specially processed to reflect less light.
Using solvents such as thinner or benzene, acidic, alkaline or corrosive detergents, chemical cleaning fibers, etc. may
damage the finish material on the surface.
Unplug the power cord before cleaning.
Never spray water directly onto the monitor or use thinner or benzene solvents, acidic, alkaline or corrosive
detergents, or cleaning or sterilizing chemical cleaning fibers, etc. as they may damage the product.
To clean the product, unplug the power cord and wipe it gently with a soft cloth. Do not spray with water or wipe
with a wet cloth. When cleaning the product or the screen, do not use cleansers, automobile or industrial thinners,
abrasives or wax, benzene, alcohol, etc., as this may damage the product.
To clean the front frame, spray water onto a soft cloth two to four times and wipe in one direction only.
Do not rub the product with a stained cloth using more force than necessary. This may scratch the surface of the
product.
Do not allow the product to remain in contact with rubber or vinyl resin for extended periods. The surface finish may
deteriorate or the coating may peel off.
If you do not clean the product, dust will accumulate inside and may create a fire hazard or malfunction.
ENGLISH
15
Image Sticking Precautions
Displaying a still image for a prolonged period may damage the screen, resulting in image sticking. Most third-party
products have the same issue. Damage resulting from long-term static images is not covered by product warranty.
1
FRANÇAIS
Pour en savoir plus sur le produit et son utilisation, assurez-vous de consulter le manuel d’utilisation inclus sur le
CD.
- Le manuel inclus sur le CD peut être lu sur un ordinateur au moyen du logiciel AdobeReader.
- Vous pouvez télécharger AdobeReader gratuitement sur le site Web d’Adobe.
Prière de ne pas modifier ou copier arbitrairement le manuel d’utilisation inclus sur le CD.
Si vous avez besoin d’une copie papier du manuel d’utilisation, veuillez communiquer avec le fabricant.
- Si le client en fait la demande à l’intérieur de septjours, le fabricant peut fournir une copie papier du manuel, et
ce, sans frais supplémentaires.
Veuillez noter que les présents renseignements concernent l’utilisation appropriée et sécuritaire de l’équipement. Les
symboles suivants peuvent indiquer une situation dangereuse potentielle qui, si elle est ignorée, peut entraîner des
blessures graves ou même la mort de l’utilisateur ou d’autres personnes, ou encore endommager l’équipement.
AVERTISSEMENT
Indique un avertissement et des consignes de sécurité. À défaut de les observer,
l’utilisateur ou d’autres personnes pourraient subir des blessures graves pouvant
entraîner la mort.
MISE EN GARDE
Indique une situation dangereuse potentielle qui, si elle est ignorée, peut entraîner des
blessures graves ou même la mort de l’utilisateur ou d’autres personnes, ou encore
endommager l’équipement.
Pour les utilisateurs aux États-Unis
La loi fédérale des États-Unis stipule que cet équipement peut seulement être utilisé par un médecin ou une personne
mandatée par un médecin.
Pour les utilisateurs des autres pays
Cet équipement peut seulement être utilisé par un détenteur de licence ou une personne mandatée par un détenteur
de licence, conformément aux lois connexes en vigueur dans chaque pays.
AVERTISSEMENTS
Pour éviter les risques de décharges électriques, cet équipement doit seulement être branché dans une prise mise à la
terre reliée à l’alimentation électrique principale.
N’apporter aucune modification à cet équipement sans l’autorisation du fabricant.
Utilisation prévue
Le moniteur médical vise à fournir un affichage couleur pour les vidéos et images générées par de l’équipement
médical, y compris les systèmes de laparoscopie et d’endoscopie utilisés en chirurgie et une variété d’autres systèmes
d’imagerie médicale. Ce produit ne prend pas en charge la 3D.
EC REP
Représentant autorisé pour la communauté européenne.
LG Electronics European Shared Service Center B.V.
Krijgsman 1, 1186 DM Amstelveen, Pays-Bas
Tél: +31-20-456-3132
LG Electronics inc.
77, Sanho-daero, Gumi-si, Gyeongsangbuk-do, 39381, République de Corée
Tél: +82-1544-8777
2
FRANÇAIS
CONSIGNES DE SÉCURITÉ
Normes de sécurité
Classification d’appareil médical
Classification par type de protection contre les
décharges électriques Équipement de classe
Classification selon le degré de protection contre la
pénétration d’eau ou de matière particulaire
Avant : IP35
Sauf pour l’avant: IP32
Mode de fonctionnement Fonctionnement continu
Environnement d’utilisation
Cet équipement ne convient pas à un environnement où
des anesthésiants inflammables ou de l’oxygène sont
utilisés.
Réglementation
Renseignements sur la sécurité et la compatibilité électromagnétique
Cet équipement a été testé et jugé conforme aux limites relatives aux appareils médicaux imposées par la
partie60601-1-2 de la réglementation de laCEI. Ces limites sont établies pour fournir une protection raisonnable
contre les interférences nuisibles susceptibles d’être observées en installation médicale.
Cet équipement génère, utilise et peut rayonner de l’énergie de radiofréquence. S’il n’est pas installé selon les présentes
consignes, il pourrait causer des interférences nuisibles susceptibles d’avoir une incidence sur d’autres appareils à
proximité.
Cependant, il nexiste aucune garantie qu’il n’y aura pas d’interférences dans une installation particulière. Si cet
équipement cause des interférences nuisibles à d’autres appareils (pour s’en assurer, il suffit d’éteindre et de rallumer
l’appareil), l’utilisateur est invité à essayer de corriger les interférences par l’une ou plusieurs des mesures suivantes:
Réorienter ou relocaliser l’équipement.
Augmenter la distance entre l’équipement et les appareils qui subissent des interférences.
Brancher l’équipement sur un circuit électrique distinct de celui qui alimente les autres appareils branchés.
Communiquer avec le fabricant ou un représentant autorisé pour obtenir de l’aide.
Conformité en matière de radiofréquences
FCC (aux États-Unis)
AVIS DE LA FCC
Cet appareil est conforme aux limites imposées par le paragraphe15 de la réglementation de la FCC. Son
fonctionnement est soumis aux deux conditions suivantes: (1) Cet appareil ne doit émettre aucune interférence
nuisible, et (2) cet appareil doit accepter toute interférence reçue, y compris les interférences qui pourraient entraîner
un fonctionnement indésirable.
3
FRANÇAIS
MISE EN GARDE DE LA FCC
Les changements ou modifications non approuvés expressément par l’autorité responsable de la conformité peuvent
annuler l’autorisation accordée à l’utilisateur de faire fonctionner l’équipement. Cet émetteur ne peut être colocalisé ni
utilisé conjointement avec tout autre émetteur ou antenne.
AVERTISSEMENT DE LA FCC
Cet équipement peut générer ou utiliser de l’énergie de radiofréquence. Toute modification apportée à cet équipement
peut causer des interférences nuisibles, sauf si elle est expressément approuvée dans le manuel d’utilisation.
Tout utilisateur qui effectue une modification non autorisée s’expose à la révocation de son droit d’utiliser l’équipement.
Remarque: Cet équipement a été testé et jugé conforme aux limites relatives aux appareils numériques de classeA
imposées par le paragraphe15 de la réglementation de laFCC. Ces limites sont établies pour fournir une protection
raisonnable contre les interférences nuisibles lorsque l’équipement est utilisé dans un environnement commercial.
Cet équipement génère, utilise et peut rayonner de l’énergie de radiofréquence. Par conséquent, s’il n’est pas installé
et utilisé conformément aux consignes qui s’appliquent en installation résidentielle, il causera vraisemblablement
des interférences nuisibles, auquel cas l’utilisateur sera tenu de prendre les mesures nécessaires à l’élimination des
interférences en question, et ce, à ses propres frais.
Pour l’Europe
DÉCLARATION DE CONFORMITÉ
Il est possible d’obtenir une copie intégrale de la déclaration de conformité en présentant une demande à cet effet,
laquelle doit être envoyée à l’adresse postale suivante:
LG Electronics European Shared Service Center B.V., Krijgsman 1, 1186 DM Amstelveen, Pays-Bas
Cette demande peut aussi être soumise par l’entremise de notre site Web réservé à cet effet: http://www.lg.com/
global/support/cedoc/cedoc#
Renseignements sur la compatibilité électromagnétique
Émissions électromagnétiques
Le matériel à l’essai est prévu pour une utilisation dans les environnements électromagnétiques indiqués ci-après.
Le client, ou l’utilisateur du matériel à l’essai, doit s’assurer que celui-ci sera utilisé dans un environnement répondant à
ces critères.
Essai d’immunité Conformité Environnement électromagnétique Conseils
Émissions de
radiofréquence CISPR11 Groupe1
Le matériel à l’essai n’utilise que de l’énergie de radiofréquence pour
son fonctionnement interne.
Par conséquent, ses émissions de radiofréquence sont très faibles
et ne risquent pas de causer d’interférences aux équipements
électroniques à proximité.
Émissions de
radiofréquence CISPR11 ClasseA
Le matériel à l’essai convient à tout type d’établissement, y compris
les établissements domestiques et ceux directement reliés au réseau
d’alimentation électrique public à basse tension qui alimente les
bâtiments à visée domestique.
Émissions harmoniques
IEC61000-3-2 A
Variations de tension/
Émissions papillotantes
IEC61000-3-3
Conforme
4
FRANÇAIS
Immunité électromagnétique
Le matériel à l’essai est prévu pour une utilisation dans les environnements électromagnétiques indiqués ci-après. Le
client, ou l’utilisateur du matériel à l’essai, doit s’assurer que celui-ci sera utilisé dans un environnement répondant à ces
critères.
Essai d’immunité IEC60601-1-2
Niveau d’essai
Niveau
de conformité
Environnement électromagnétique –
Conseils
Décharge
électrostatique
(DES)
IEC61000-4-2
± 8kV (contact)
± 2kV, ± 4kV,
± 8kV, ± 15kV (air)
± 8kV (contact)
± 15kV (air)
Les planchers doivent être en bois
ou béton, ou recouverts de tuiles
de céramique. Si les planchers sont
recouverts d’un matériau synthétique,
l’humidité relative doit être maintenue
au-dessus de 30%.
Transitoire électrique
rapide en salves
IEC61000-4-4
± 2kV
100kHz (fréquence de
répétition)
± 2kV
100kHz (fréquence de
répétition)
La qualité énergétique de l’alimentation
électrique principale doit correspondre
à ce que l’on retrouve habituellement en
environnement commercial ou en milieu
hospitalier.
Surtension
ligne à ligne
IEC61000-4-5
± 0,5kV, ± 1kV ± 1kV La qualité énergétique de l’alimentation
électrique principale doit correspondre
à ce que l’on retrouve habituellement en
environnement commercial ou en milieu
hospitalier.
Surtension
ligne à terre
IEC61000-4-5
± 0,5kV, ± 1kV, ± 2kV ± 2 kV
Baisses de tension
IEC61000-4-11
0% en UT pour un cycle
de 0,5
À 0°, 45°, 90°, 135°,
180°, 225°, 270° et 315°
0% en UT pour un cycle
de 0,5
À 0°, 45°, 90°, 135°,
180°, 225°, 270° et 315°
La qualité énergétique de l’alimentation
électrique principale doit correspondre
à ce que l’on retrouve habituellement
en environnement commercial ou
en milieu hospitalier. Si l’utilisateur
de l’intensificateur d’image à l’essai
nécessite que celui-ci fonctionne de
manière interrompue lors d’interruptions
dans l’alimentation électrique principale,
il est recommandé de le brancher à une
source d’alimentation sans coupure ou
une batterie.
0% en UT
pour un cycle de 1
et
70% en UT pour des
cycles de 25/30
Monophasé à 0°
0% en UT
pour un cycle de 1
et
70% en UT pour des
cycles de 25/30
Monophasé à 0°
Coupures de courant
IEC61000-4-11
0% en UT pour un cycle
de 250/300
0% en UT pour un cycle
de 250/300
Fréquence
d’alimentation
NOMINALE
champs magnétiques
(50/60Hz)
IEC61000-4-8
30 A/m 30 A/m Les champs magnétiques de la
fréquence d’alimentation doivent
présenter des niveaux que l’on retrouve
habituellement en environnement
commercial ou en milieu hospitalier.
REMARQUE
Le UT correspond à la tension c.a. de l’alimentation électrique principale avant l’application du niveau d’essai.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78

LG 27HJ713SW Ghid de inițiere rapidă

Categorie
Bucătărie
Tip
Ghid de inițiere rapidă