18
ALARM TEMPERATURE SETTING
This unit is provided with both high and low temperature alarms. The temperature at which the alarm is
activated can be changed.The high temperature alarm can be set to a temperature of + 5 °C to + 15 °C.
The low temperature alarm can be set to a temperature between -15 °C and -5 °C.
Note: The temperature alarm is set at ±10 oC of the set temperature at the factory.
As an example, Table 3 shows the procedure to set the high temperature alarm so that the alarm can
activate when the chamber temperature is 5 oC higher than the set temperature.
Table 4 shows the procedure to set the low temperature alarm so that the alarm can activate when the
chamber temperature is 5 oC lower than the set temperature.
Table 3. Procedure for setting high temperature alarm
Description of operation Key operated Indication after operation
1 ― ― The current chamber temperature is
displayed.
2 Press numerical value shift key for
about 5 seconds.
The display changes to F00 and
the first digit blinks.
3
Set the figure to F01 by using
digit shift key and numerical value
shift key
When pressed, the figure of
settable digit changes.
4 Press set key. SET The current setting is displayed and
the first digit is flashed.
5
Set the figure to 005 by using
digit shift key and numerical value
shift key
When pressed, the settable digit is
shifted.
When pressed, the figure of
settable digit changes.
6 Press set key. SET
Alarm temperature is memorized and
the current chamber temperature is
displayed.
Table 4. Procedure for setting low temperature alarm
Description of operation Key operated Indication after operation
1 ― ― The current chamber temperature is
displayed.
2 Press numerical value shift key for
about 5 seconds.
The display changes to F00 and
the first digit blinks.
3
Set the figure to F02 by using
digit shift key and numerical value
shift key
When pressed, the figure of
settable digit changes.
4 Press set key. SET The current setting is displayed and
the first digit is flashed.
5
Set the figure to -05 by using
digit shift key and numerical value
shift key
When pressed, the settable digit is
shifted.
When pressed, the figure of
settable digit changes.
6 Press set key. SET
Alarm temperature is memorized and
the current chamber temperature is
displayed.
• The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.