Please read the operating instructions carefully before using this product, and keep the operating
instructions for future use.
See page 41 for all model numbers.
Operating Instructions
Pharmaceutical Refrigerator
MPR-1412
2
CONTENTS
INTRODUCTION ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ3
PRECAUTIONS FOR SAFE OPERATION ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ4
ENVIRONMENTAL CONDITIONS ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ9
SYMBOLS ON UNIT ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ9
REFRIGERATOR COMPONENTS
Main body ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 10
Control panel ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 12
Inside the front over ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 13
INSTALLATION SITE ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 14
INSTALLATION ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 15
START-UP PROCEDURE ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 17
PLACEMENT OF STORED ITEMS ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 18
SETTING THE CHAMBER TEMPERATURE ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 19
LOCK OF CHAMBER TEMPERATURESETTING ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 20
SETTING THE TEMPERATURE DIPLAY MODE ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 21
DEFROSTING THE EVAPORATOR ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 2 2
SETTING THE ALARM TEMPERATURE
Setting the high-temperature alarm ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 23
Setting the low-temperature alarm ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 24
SETTING THE DELAY OF DOOR ALARM ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 25
SETTING THE BUZZER SUSPENDED PERIOD ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 26
OPERATION CHECK AFTER RECOVERY ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 26
REMOTE ALARM TERMINAL ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 27
ALARM FUNCTIONS ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 28
Air circulation alarm ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 28
SAFETY FUNCTIONS ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 28
SELF-DIAGNOSTIC FUNCTIONS ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 29
ROUTINE MAINTENANCE
Cleaning of exterior, interior, and accessories ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 30
Cleaning of evaporating tray ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 30
Replacement of fluorescent lamp ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 31
Replacement of glow starter ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 32
Cleaning of condenser filter ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 32
TROUBLESHOOTING ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 33
DISPOSAL OF UNIT ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 34
Decontamination of unit ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 34
TEMPERATURE RECORDER (OPTION) ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 39
BATTERY MOUNTING BOX (OPTION) ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 3 9
INTERFACE BOARD (OPTION) ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 39
SPECIFICATIONS ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 40
PERFORMANCE ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 41
SAFETY CHECK SHEETꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ 42
3
INTRODUCTION
Read the operating instructions carefully before using the appliance and follow the instructions for safe
operation.
Handling of Operating Instructions
PHC Corporation never guarantees any safety if the appliance is used for any objects other than
intended use or used by any procedures other than those mentioned in the operating instructions.
Keep the operating instructions in an adequate place to refer to it as necessary.
Contact our sales representative or agent if any page of the operating instructions is lost or page order
is incorrect.
Contact our sales representative or agent if any point in the operating instructions is unclear or if there
are any inaccuracies.
The operating instructions is only for the pharmaceutical refrigerator MPR-1412 and MPR-1412R made by PHC
Corporation.
The contents are subject to change without notice.
No part of the operating instructions may be reproduced in any form without the expressed written permission of
PHC Corporation.
Proposal for safe storage of refrigerator contents
This pharmaceutical refrigerator can store the medical or biological material in the low temperature
environment. We can offer some precautions below to protect the valuable material against accidental
temperature rise. Contact our sales representative or agent for the details or installation.
▪ Temperature recorder (Option)
▪ Battery mounting box (Option)
▪ Remote alarm system (Marketed product)
IMPORTANT NOTICE
PHC Corporation guarantees this product under certain warranty conditions. However, please note that
PHC Corporation shall not be responsible for any loss or damage to the contents of the product.
<Intended Use>
This equipment is designed for storage of pharmaceuticals, samples and reagents.
4
PRECAUTIONS FOR SAFE OPERATION
It is imperative that the user complies with the operating instructions
as they contain important safety advice.
Items and procedures are described so that you can use this unit correctly and safely.
Following these precautions will prevent possible injury to the user and any other
person.
Precautions are illustrated in the following way:
WARNING
Waring indicates a potentially hazardous situation which, if not avoided,
could result in serious injury or death.
CAUTION
Failure to observe CAUTION signs could result in injury to personnel and
damage to the unit and associated property.
Be sure to keep the operating instructions in a place that is accessible to users of this
unit.
USA Only: This product has a fluorescent lamp that contains mercury. Disposal may be regulated in your
community due to environmental considerations. For disposal or information, please visit PHC website:
https://www.phchd.com.
Contains mercur
y
/ Contenu avec mercure
For more information on safe handling
procedures, the measures to be taken in
case of accidental breakage and safe
disposal options visit:
ec.gc.ca/mercure-mercury/.
Dispose of or recycle in accordance with
applicable laws.
Pour plus de renseignements sur les
procédures de manutention sécuritaire, les
mesures à prendre en cas de bris accidentel et
les options d’élimination sécuritaire visitez:
ec.gc.ca/mercure-mercury/.
Mettez au rebut ou recyclez conformément
aux lois applicables.
5
PRECAUTIONS FOR SAFE OPERATION
Never damage the power supply cord or power supply plug (by breaking, adapting, placing
near a source of heat, bending with force, twisting, pulling, adding weight, or binding). A
damaged power supply cord or power supply plug may cause electric shock, short circuit, or fire.
Remove dust from the power supply plug periodically. Dust on the power supply plug may cause
insulation failure due to moisture and thus cause a fire. Disconnect the power supply plug and wipe it
with a dry cloth.
Make sure the power supply plug is pushed fully in. Faulty insertion of the power supply plug may
cause electric shock or fire due to generation of heat. Never use a damaged power supply plug or
loose power outlet.
Do not touch any electrical parts (such as power supply plug) or operate switches with a wet
hand. This may cause electric shock.
Use a dedicated power source as indicated on the rating label attached to the unit. A
multiple-tap may cause fire resulting from abnormal heating.
Do not use the unit outdoors. Exposure to rain may cause leakage and/or electric shock.
Only qualified engineers or service personnel should install the unit. Installation by unqualified
personnel may cause water leakage, electric shock, or fire.
Install the unit in a location capable of bearing the total combined weight (product + optional
accessories + stored items). After installing the unit, be absolutely sure to take precautions to
prevent the unit from falling over. If the unit is installed in a location which is not strong enough or if
the proper precautions are not taken, the unit may fall over and cause injuries.
Do not install the unit in a location where flammable or volatile substances are present.
Installing the unit in a location where flammable or volatile substances are present may cause
explosions and/or a fire.
Do not install the unit where there are high levels of moisture or where it may be splashed with
water. Installing the unit where there are high levels of moisture or where it may be splashed with
water may cause the insulation to deteriorate and give rise to leakage and/or electric shock.
Be absolutely sure to earth (gound) the unit in order to prevent electric shock. Failure to earth
the product may give rise to electric shock. If necessary, ask a qualified contractor to do this work.
Do not connect the earth wire to a gas pipe, water pipe, or lightning rod when earthing
the unit. Earthing the unit improperly may give rise to electric shock.
WARNING
6
PRECAUTIONS FOR SAFE OPERATION
Never store volatile or flammable substances in this unit except in a sealed container. Such
substances may cause explosion or fire if they leak.
Never insert metal objects such as pins and wires into any vent, gap, or outlet on the unit.
This may cause electric shock or injury by accidental contact with moving parts.
When handling harmful samples (for example, those which consist of toxic, pathogenic, or
radioactive substances), install the unit inside a designated isolation facility. If the unit is
installed in a location which is not an isolation facility, there may be detrimental effects on both people
and the natural environment.
Before proceeding with maintenance or checking of the unit, set the power switch to OFF and
disconnect the power supply plug. Performing the work while power is still flowing to the product or
while the power supply plug is still connected may give rise to electric shock and/or injury.
Wear protective gloves and mask during maintenance. Touching or inhaling chemicals or aerosols
from around the unit may be detrimental to health.
Never splash water directly onto the unit as this may cause electric shock or short circuit.
Never put containers with liquid on top of the unit as this may cause electric shock or short circuit
if the liquid is spilled.
Never disassemble, repair, or modify the unit yourself. A high-voltage area is located inside the
unit. Any work carried out by unauthorized personnel may result in electric shock. Contact our sales
representative or agent for maintenance or repair.
Disconnect the power supply plug if there is anything wrong with the unit. Continued abnormal
operation may cause electric shock or fire.
Grip the power supply plug when disconnecting the power supply cord from the outlet.
Pulling the power supply cord may cause electric shock or short circuit.
Disconnect the power supply cord when the unit is not in use for long periods. Keeping the unit
connected may cause electric shock, leakage, or fire due to the deterioration of insulation.
Do not place this unit in a location where it is difficult to disconnect the power supply plug.
Failure to disconnect the power supply plug may cause fire in the event of a problem or malfunction.
If the unit is to be stored unused in an unsupervised area for a long period, ensure that children do
not have access and that doors cannot be closed completely.
WARNING
7
PRECAUTIONS FOR SAFE OPERATION
Ask a qualified contractor to carry out disassembly and disposal of the unit. Leaving the unit in
a location that can be accessed by third parties may result in unexpected accidents (e.g. the unit may
be used for unintended purposes).
Do not leave the plastic bags used for packing in a place where they can be reached by small
children as this may result in unexpected accidents such as suffocation.
When moving the unit, be sure to take precautions to prevent it from falling over. Moving the
unit with too much force may cause it to fall over, possibly resulting in injury. A qualified individual
must be assigned to supervise the safe movement and relocation of the unit.
Do not install the unit in a location where corrosive gases such as acids are present. Installing
the unit in a location where corrosive substances are present may cause electric components to
corrode, leading to leakage and/or electric shock due to the deterioration of insulation resulting from
corroded electrical components.
Connect the unit to a power source as indicated on the rating label attached to the unit. Use of
any other voltage or frequency other than that on the rating label may cause fire or electric shock.
Never replace the battery for the power failure alarm yourself. Only qualified engineers or service
personnel should replace the battery.
WARNING
8
PRECAUTIONS FOR SAFE OPERATION
This unit must be plugged into a dedicated circuit protected by branch circuit breaker.
Use a dedicated power source as indicated on the rating label attached to the unit. A
multiple-tap may cause fire resulting from abnormal heating.
Do not climb on top of the unit or put any objects on the unit. Falling from the unit may cause
injury; falling objects may cause damage to the unit.
Never install the unit in a location where corrosive materials such as sulphur compounds are
likely to be generated (e.g. near a drainage facility). Corrosion of the copper pipes may result in the
deterioration and consequently the failure of the cooling unit.
Take care when installing the shelves or drawers. They may cause injury if they fall down.
Before resuming operation after the unit has been turned off, be absolutely sure to check the
settings. The settings may have changed as a result of stopping the unit. Items stored inside the unit
may be adversely affected if operation of the unit is resumed with changed settings.
Never store corrosive substances such as acids or alkalis in the unit except in a sealed
container. These may be detrimental to health and may cause corrosion of internal components,
cooling circuit or electrical components.
When moving the unit, contact our sales representative or agent. The moving by a user may
cause damage or injury resulting from tipping over of the unit.
Disconnect the power supply plug before moving the unit. Take care not to damage the power
supply cord. A damaged cord may cause electric shock or fire.
To ensure the safety of the service engineer, submit a safety check sheet with the required
items filled out. This is provided as the photocopiable “Safety Check Sheet” at the end of these
operating instructions.
CAUTION
9
ENVIRONMENTAL CONDITIONS
This equipment is designed to be safe at least under the following conditions (based on the IEC 61010-1):
Indoor use;
Altitude up to 2000 m;
Temperature 5 oC to 40 oC
Maximum relative humidity 80 % for temperature up to 31 oC decreasing linearly to 50 % relative
humidity at 40 oC;
Mains supply voltage fluctuations up to ±10 % of the nominal voltage;
Transient overvoltages up to the levels of OVERVOLTAGE CATEGORY II;
Temporary OVERVOLTAGES occurring on the mains supply;
Applicable pollution degree of the intended environment (POLLUTION DEGREE 2 in most cases);
SYMBOLS ON UNIT
The following symbols are attached to the unit. The table describes the meaning of the symbols.
This symbol is attached to covers that access high-voltage electrical components to
prevent electric shock. Only a qualified engineer or service personnel should be
allowed to open these covers.
This symbol indicates that caution is required. Refer to product documentation for
details.
This symbol indicates a hot surface.
This symbol indicates an earth.
This symbol means “ON” for a power switch.
This symbol means “OFF” for a power switch.
10
REFRIGERATOR COMPONENTS
Main body
MPR-1412
Back side
14
15
3 1 2 4
5
8 9
10
11
12 7
13
6
11
REFRIGERATOR COMPONENTS
1. Front cover: Open this cover when using the remote alarm terminal or replacing the glow starter [page
13]. To open the cover, pull the lower portion forward.
2. Control panel: The chamber temperature and other alarms/functions can be set through the keys on
the control panel. The running status can be checked by the temperature display and indicators [page
12].
3. Light switch: This is a switch to turn on or off the fluorescent lamp.
4. Space for temperature recorder: A temperature recorder (optional) can be mounted here so that the
chamber temperature can be recorded automatically [page 39].
Contact our sales representative or agent to arrange the installation.
5. Access port: This port is used to pass the sensor or cable of measuring equipment, or the sensor of a
temperature recorder (optional) to the chamber.
Replace the insulation and the rubber caps when the access port is not in use. Improper replacement
may cause an increase in chamber temperature or condensation around the access port.
6. Keyhole: Turn the key counterclockwise through 180 degree to securely lock the door.
7. Glass window: The condensation may be found on the glass under high temperature and humidity
condition. Wipe off the condensation with a soft dry cloth.
8. Dew collector: The condensation on the glass window is accumulated in this dew collector. Wipe off
the water from the dew collector with a cloth.
9. Leveling foot: These are screw bolts used to install and fix the unit. Adjust the height of the levelling
feet by turning the screw bolts until the two front casters are away from the floor [page 15].
10. Shelf (MPR-1412): The location is selectable. The allowable load is 50 kg per shelf.
Never store items on the bottom of the chamber. Always store items in the drawers.
Drawer (MPR-1412R): The allowable load is 40 kg per drawer.
<Important>
Do not pull out plural drawers at the same time when putting in and out of the stored items. If you store
heavy objects, it may cause injury due to the fall of the unit.
11. Air intake vent: This is an air intake vent to circulating the air in the chamber. Do not block this vent.
Blocking this vent may cause unstable chamber temperature.
Do not insert fingers or similar articles into this vent.
12. Air exhaust vent: This is an air exhaust vent to circulating the air in the refrigerator chamber. Do not
block this vent.
Blocking this vent may cause unstable chamber temperature.
Arrange the stored items not to be subjected to the cold air from this vent to avoid freezing.
13. Door switch: This switch detects the door open/close status. The door check indicator is lit when the
door is opened.
14. Fixture: Use the fixtures and secure the unit to a wall with a strong rope or chain [page 15].
15. Evaporating tray: Defrost water from the evaporator accumulates on the tray and evaporates into the
atmosphere.
12
REFRIGERATOR COMPONENTS
Control panel
1. Door check indicator (DOOR): The red LED
lamp is lit when the door is opened.
2 minutes after the door check indicator is lit,
the buzzer is activated to notice the door opening
[page 28].
2. Alarm indicator (ALARM): The red LED lamp
blinks during an alarm condition [page 28].
3. Temperature display: Normally shows the
present chamber temperature and during an
alarm condition, shows a blinking chamber
temperature, and an error code and chamber
temperature are displayed alternately when the
self-diagnostic function activates [page 29].
4. Up arrow key ( ):
At “temperature display mode”; pressing this key
for more than 5 seconds leads setting mode
[page 21, page 23 to page 26].
At “setting mode”; changes number values
between choices [page 21, page 23 to page 26].
At setting of lock function”; selects ON-OFF of
key lock for chamber temperature setting [page
20].
5. Scroll key ( ):
At “temperature display mode”; pressing this key
for more than 5 seconds leads “setting of lock
function” [page 20].
At “setting mode”; moves the operator through
digit on the display.
6. Set key (SET):
At “temperature display mode”; pressing this key
leads “chamber temperature setting mode”.
At “setting mode”; stores the value into computer
memory.
7. Buzzer stop key (BUZZER): Mutes the audible
alarm during alarm status (The remote alarm is
also canceled).
The audible alarm recovers after certain period
(suspended time) if the same alarm status
continues [page 26].
“temperature display mode”: the status which
the temperature display shows the current
chamber temperature
“setting mode”; the status which the
temperature display is ready to be entered after
pressing the set key or up arrow key for 5
seconds.
7 6 5 4
3 2 1
13
REFRIGERATOR COMPONENTS
Inside the front cover
1. Condenser filter
This is a filter for preventing for clogging of the condenser. The clogged filter may cause failure of
refrigerating circuit. Clean the filter every 3 months [page 32].
2. Remote alarm terminal
The alarm is relayed to a remote location when a remote alarm device (commercially available) is
connected to the remote alarm terminal. Installation of a remote alarm device is recommended when the
unit is installed in an unattended location so that the operator of the unit is notified of the alarm. Refer to
page 27 for details.
3. Power switch (also functions as a circuit breaker)
This is the power switch of the unit. (ON=”l”, OFF=””)
The round button under the power switch is a test button for the circuit breaker. After pressing this
button, the power to the unit is disconnected.
4 Glow starter (covered with water-proof cover)
This is for the fluorescent lamp. It is recommended to replace the glow starter when the fluorescent lamp
is replaced. Refer to page 32 for the replacement.
1 2
4
3
14
INSTALLATION SITE
This unit must be installed in a location which meets all the conditions described below.
If the unit is installed in a location which does not meet the conditions, its specified performance may
not be achieved or malfunctions and accidents may occur.
A location not exposed to direct sunlight
Avoid any location which is exposed to direct sunlight. Installing the unit in a location exposed to direct
sunlight may reduce its cooling performance.
A well-ventilated (airy) location
In order to ensure ventilation, leave clearances of at least 10 cm around the unit (at the left, right, top and
back). Blocking the ventilation may reduce the unit’s cooling performance or cause malfunctions.
A location away from sources of heat
Avoid any location which is close to a major source of heat (such as a heater or boiler). Installing the unit
near a major source of heat may reduce the unit’s cooling performance.
A location with minimal changes in temperature
Avoid any location where the ambient temperature is subject to sudden changes. If the unit is installed in
a location where the ambient temperature is subject to sudden changes, it will not be possible to achieve
a stable cooling performance.
A flat surface where the floor is also capable of bearing the total combined weight (product +
optional accessories + stored items)
Install the unit on a flat surface which is even and which is capable of bearing the total combined weight
(product + optional accessories + stored items). If the unit is installed where the surface is uneven or
where the unit will be inclined at an angle, the unit will be unstable, and accidents or injuries may occur
and/or vibration or unnecessary noise may be generated.
A location with minimal humidity
Install the unit in a location where the relative humidity is less than 80 %R.H. Installing the unit in a very
humid location may cause earth faults and/or electric shock.
A location free of flammable or corrosive gases
Avoid any location exposed to flammable or corrosive gases. Flammable or corrosive gases can cause
explosions and/or a fire. Furthermore, corrosion of electrical parts may cause insulation to be reduced
and result in earth faults and/or electric shock.
A location where nothing can fall onto the unit
Avoid locations where objects may fall onto the unit. Objects falling and hitting the unit may cause it to
break down or fail.
15
INSTALLATION
1. Preparations after unpacking
Remove all the tape used to secure the doors and interior parts, and leave the doors open for a short
while for ventilation.
If any surfaces of the outer cabinet are dirty, wipe the surface using a cloth moistened with a diluted
neutral dish-washing detergent.
Using an undiluted solution of detergent may cause the unit’s plastic areas to crack. Follow the
directions on the detergent for details of dilution.
After wiping the unit using the diluted detergent, be absolutely sure to wipe the surfaces with a cloth
dipped in clean water to remove traces of the detergent. After this, be absolutely sure to wipe the
surfaces with a dry cloth, allowing the surfaces of the outer cabinet to dry out completely, and then
proceed with the installation.
Note:
Remove the cable tie that binds the power supply cord. Prolonged contact with the tie may cause
corrosion of the cord coating.
2. Securing and levelling the unit using the levelling
feet
Rotate the front levelling feet clockwise until the casters
are raised 5 mm to 10 mm above the floor surface [Fig. 1].
In addition, rotate the levelling feet slightly clockwise or
anticlockwise, and adjust them so that the unit is
completely level.
When the casters are raised from the floor surface, the
unit will be secured. If they are left touching the floor, the
unit may accidently move when its door is opened or
closed.
3. Securing the unit by using the fixtures
Use the fixtures on the rear panel of the unit, and secure
the unit to a wall with a strong rope or chain [Fig. 2].
4. Preventing electric shock by earthing the unit
When installing the unit, be absolutely sure to earth
(ground) it. Earthing is necessary to prevent electric
shocks resulting from deterioration of electrical insulations.
This unit comes with a 3-pin plug having one earth pin.
Earthing is not required in the case of a 3-pin power outlet
equipped with an earth contact.
If the power outlet is not a 3-pin outlet equipped with an
earth contact, ask a qualified contractor to do the earthing
work.
Fig. 1
Leveling foot
Fixture
Fig. 2
16
INSTALLATION
5. Setting up the shelves (MPR-1412)
All shelves are packaged at the bottom of the chamber. Set the shelves firmly in place on the shelf
stoppers at the standard locations (4 locations) below.
1) The sixth hole under the second screw from the top (top screw) [Fig. 3]
2) The first hole under the fifth screw from the top.
3) The sixth hole under the seventh screw from the top
4) The first hole under the tenth screw from the top.
<Important>
Never put the stored item on the bottom of the chamber. Always put them on the shelf or in the drawer (in
the case of MPR1412R).
Fig. 3
Shelf support
Shelf stopper
Second screw
from the top
17
START-UP PROCEDURE
Follow this procedure for the initial operation of the unit and for consequent operations (after temporary
stoppage for cleaning, maintenance or moving).
After a power failure, the unit will restart operation automatically with the same settings as before the
power failure [page 26].
1. Connect the power supply cord to the dedicated outlet with appropriate rating with the chamber empty.
►The alarm indicator and the chamber temperature may blink. And the buzzer may sound. In this case,
stop the buzzer by pressing the buzzer stop key (BUZZER). (This is not a malfunction.)
<Important>
If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes.
This protects the compressor.
Close the door securely. The door check indicator is lit when the door is open. The buzzer activates when
the door is opened for more than two minutes. The buzzer is canceled when the door is closed.
2. Allow the chamber temperature to fall to 5 oC (The chamber temperature is set to 5 oC at the factory).
Check the chamber temperature on the temperature display.
Allow the chamber temperature to fall to the set temperature when the start-up after cleaning,
maintenance or moving.
3. Switch on the light switch to check the fluorescent lamp is on. After checking, turn off the fluorescent
lamp if the light is not necessary.
4. Set the desired chamber temperature [page 19].
5. When the chamber temperature gets to the set temperature, begin slowly placing items into the
chamber to minimize the temperature rise.
<Important>
Always put the stored items on the shelf (MPR-1412) or in the drawer (MPR-1412R), not on the bottom of
the chamber. Do not contact the stored items to the chamber walls.
Do not block the air intake vent nor air exhaust vent. Place the stored items with adequate space between
them to keep the cool air circulation.
<Attention>
The chamber temperature cannot be maintained due to the heat emission from the electrical equipment if
any electrical equipment is used in the chamber. Refer to the table below when using any equipment.
(ambient temperature; 35 oC)
Heat load in chamber Lowest chamber temperature
50 W 2 oC
100 W 3 oC
200 W 9 oC
18
PLACEMENT OF STORED ITEMS
Do not place stored items so as to disrupt the air circulation in the chamber. Disruption of air circulation
can cause the freezing of stored items or can impair the temperature distribution in the chamber.
Air intake vent
Do not block the air intake vent with the stored items. Care should be taken not to put any paper or plastic
chips near the air intake vent. These may be sucked in and block the air intake vent.
Air exhaust vent
Do not block the air exhaust by the stored items near the air exhaust vent. Care should be taken not to
place stored items near to the air exhaust vent. The stored items may become frozen when the
refrigerator is operating at temperature settings lower than 5 oC.
Load line [see figure below]
The storage area is limited to ensure adequate air flow in the chamber. The label “load line” is attached to
the right and left sides of the chamber. Always place stored items within the area limited by these labels.
Load line
Load line
Stock space
Side view
Cooling air
Fan Air exhaust vent
Air intake vent
19
SETTING THE CHAMBER TEMPERATURE
Set the chamber temperature according to the condition of use. This refrigerator can keep the stored
items for long period under appropriate temperature.
Setting range of chamber temperature: between 2 oC and 23 oC
Initial setting (factory setting): 5 oC
<Important>
The chamber temperature lower than 3 oC may cause partial freeze of stored items.
Example: Change the chamber temperature to 4 oC from 5 oC
Following shows a sample setting. Set the desired chamber temperature according to the condition of
use.
Operation Ke
y
operated Displa
y
after operation
1
Connect the power supply plug to
the dedicated outlet.
(
onl
y
when start-up of unit
)
----- The current chamber temperature
blinks.
2 Press the “set” key. SET The current setting (005) is displayed
and the first di
g
it blinks.
3 Press the up arrow key and scroll
the fi
g
ure to 4. The display is changed to 004 from
005.
4 Press the “set” key. SET
Set temperature is memorized and
the current chamber temperature
blinks.
The unit will return from setting mode to temperature display mode automatically after 90 seconds if no
key is operated (auto-return function). In this case, the setting is not changed.
<Attention>
It is recommended to install a temperature recorder (optional component) to check the highest/lowest
chamber temperature when the item which needs severe temperature control is stored.
20
LOCK OF CHAMBER TEMPERATURE SETTING
The chamber temperature setting can be locked to avoid accidental changes. When the lock is ON, a
change of chamber temperature setting is not accepted even if the key on the control panel is operated.
Initial setting (factory setting): lock OFF
Displa
y
Lock s
atus Settin
g
of chamber temperature
L 0 Lock OFF Enabled to chan
g
e
L 1 Lock ON Disabled from chan
g
e
Example: Change the lock to ON from OFF (factory setting)
Operation Ke
y
operated Displa
y
after operation
----- ----- The current chamber temperature is
displa
y
ed.
1 Press the scroll key for 5 seconds.
The current setting (L 0) is displayed
and the first di
g
it blinks.
2 Press the up arrow key once.
The display is changed to L 1 from
L 0.
3 Press the “set” key. SET
Lock status is memorized and the
current chamber temperature is
displa
y
ed.
The unit will return from setting mode to temperature display mode automatically after 90 seconds if no
key is operated (auto-return function). In this case, the setting is not changed.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32