Brooks 0254 Instrucțiuni de utilizare

Categorie
Măsurare, testare
Tip
Instrucțiuni de utilizare
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Model 0254
5
Gas and Liquid Mass Flow
Secondary Electronics
Model 0254
Table Top Four Channel
Secondary Electronics
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Model 0254
ESD (Electrostatic Discharge)
CAUTION: This instrument contains electronic components that are susceptible to damage by static electricity. Proper handling procedures
must be observed during the removal, installation or other handling of internal circuit boards or devices.
Handling Procedure:
1. Power to unit must be removed.
2. Personnel must be grounded, via a wrist strap or other safe, suitable means before any printed circuit card or other internal device is installed,
removed or adjusted.
3. Printed circuit cards must be transported in a conductive container. Boards must not be removed from protective enclosure until immediately before
installation. Removed boards must immediately be placed in protective container for transport, storage or return to factory.
Comments
This instrument is not unique in its content of ESD (electrostatic discharge) sensitive components. Most modern electronic designs contain components
that utilize metal oxide technology (NMOS, SMOS, etc.). Experience has proven that even small amounts of static electricity can damage or destroy these
devices. Damaged components, even though they appear to function properly, exhibit early failure.
Brooks Instrument designs, manufactures and tests its products to meet many national and international standards. These products must be properly
installed, operated and maintained to ensure they continue to operate within their normal specifications. The following instructions must be adhered to
and integrated into your safety program when installing, operating and maintaining Brooks Instrument products.
To ensure proper performance, use qualified personnel to install, operate, update, program and maintain the product.
Read all instructions prior to installing, operating and servicing the product. If this instruction manual is not the correct manual, please see back cover
for local sales office contact information. Save this instruction manual for future reference.
WARNING: Do not operate this instrument in excess of the specifications listed in the Instruction and Operation Manual. Failure to heed
this warning can result in serious personal injury and / or damage to the equipment.
If you do not understand any of the instructions, contact your Brooks Instrument representative for clarification.
Follow all warnings, cautions and instructions marked on and supplied with the product.
Install your equipment as specified in the installation instructions of the appropriate instruction manual and per applicable local and national codes.
Connect all products to the proper electrical and pressure sources.
Operation: (1) Slowly initiate flow into the system. Open process valves slowly to avoid flow surges. (2) Check for leaks around the flow meter inlet
and outlet connections. If no leaks are present, bring the system up to the operating pressure.
Please make sure that the process line pressure is removed prior to service. When replacement parts are required, ensure that qualified people use
replacement parts specified by Brooks Instrument. Unauthorized parts and procedures can affect the product's performance and place the safe
operation of your process at risk. Look-alike substitutions may result in fire, electrical hazards or improper operation.
Ensure that all equipment doors are closed and protective covers are in place to prevent electrical shock and personal injury, except when
maintenance is being performed by qualified persons.
WARNING: For liquid flow devices, if the inlet and outlet valves adjacent to the devices are to be closed for any reason, the devices must
be completely drained. Failure to do so may result in thermal expansion of the liquid that can rupture the device and may cause personal
injury.
All pressure equipment with an internal pressure greater than 0.5 bar (g) and a size larger than 25mm or 1" (inch) falls under the Pressure Equipment Directive
(PED).
The Specifications Section of this manual contains instructions related to the PED directive.
Meters described in this manual are in compliance with EN directive 97/23/EC.
All Brooks Instrument Flowmeters fall under fluid group 1.
Meters larger than 25mm or 1" (inch) are in compliance with PED category I, II or III.
Meters of 25mm or 1" (inch) or smaller are Sound Engineering Practice (SEP).
The Brooks Instrument (electric/electronic) equipment bearing the CE mark has been successfully tested to the regulations of the Electro Magnetic
Compatibility (2004/108/EC (EMC directive 89/336/EEC)).
Special attention however is required when selecting the signal cable to be used with CE marked equipment.
Quality of the signal cable, cable glands and connectors:
Brooks Instrument supplies high quality cable(s) which meets the specifications for CE certification.
If you provide your own signal cable you should use a cable which is overall completely screened with a 100% shield.
“D” or “Circular” type connectors used should be shielded with a metal shield. If applicable, metal cable glands must be used providing cable screen
clamping.
The cable screen should be connected to the metal shell or gland and shielded at both ends over 360 Degrees.
The shield should be terminated to an earth ground.
Card Edge Connectors are standard non-metallic. The cables used must be screened with 100% shield to comply with CE certification.
The shield should be terminated to an earth ground.
For pin configuration : Please refer to the enclosed Instruction Manual.
European Pressure Equipment Directive (PED)
European Electromagnetic Compatibility (EMC)
Essential Instructions
Read before proceeding!
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Model 0254
Dear Customer,
We appreciate this opportunity to service your flow measurement and control requirements with a Brooks
Instrument device. Every day, flow customers all over the world turn to Brooks Instrument for solutions to their
gas and liquid low-flow applications. Brooks provides an array of flow measurement and control products for
various industries from biopharmaceuticals, oil and gas, fuel cell research and chemicals, to medical devices,
analytical instrumentation, semiconductor manufacturing, and more.
The Brooks product you have just received is of the highest quality available, offering superior performance,
reliability and value to the user. It is designed with the ever changing process conditions, accuracy requirements
and hostile process environments in mind to provide you with a lifetime of dependable service.
We recommend that you read this manual in its entirety. Should you require any additional information concerning
Brooks products and services, please contact your local Brooks Sales and Service Office listed on the back cover
of this manual or visit www.BrooksInstrument.com
Yours sincerely,
Brooks Instrument
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Model 0254
THIS PAGE WAS
INTENTIONALLY
LEFT BLANK
i
Contents
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Model 0254
Paragraph Page
Number Number
Section 1 Introduction
1-1 Description .................................................................................................................................1-1
1-1-1 Architecture................................................................................................................................ 1-1
1-1-2 Communication .......................................................................................................................... 1-1
1-1-3 Process Controls ....................................................................................................................... 1-1
1-1-4 Operator Controls and Indicators............................................................................................... 1-2
1-1-5 Diagnostic Capabilities...............................................................................................................1-2
1-2 Specifications ............................................................................................................................. 1-2
1-3 Signal Wiring.............................................................................................................................. 1-6
1-4 Optional Equipment ...................................................................................................................1-7
1-4-1 Mounting Kit Options..................................................................................................................1-7
1-4-2 Power Supply Output Voltage Options .......................................................................................1-7
1-4-3 Input/Output Signal Adapters ..................................................................................................... 1-7
Section 2 Installation
2-1 General ...................................................................................................................................... 2-1
2-2 Receipt of Equipment................................................................................................................. 2-1
2-3 Recommended Storage Practice ............................................................................................... 2-1
2-4 Return Shipment ........................................................................................................................ 2-2
2-5 Transit Precautions .................................................................................................................... 2-2
2-6 Removal from Storage ...............................................................................................................2-2
2-7 Ventilation and Mounting Requirements ....................................................................................2-2
2-8 Cleaning Instructions ................................................................................................................. 2-2
2-9 Cable Requirements ..................................................................................................................2-3
2-10 Installation Instructions .............................................................................................................. 2-3
2-10-1 Panel Mount Installation Instructions ......................................................................................... 2-3
2-10-2 Table Top Stand Assembly Instructions ...................................................................................... 2-5
2-10-3 Retrofit to Model 0152/0154 Table Top Housing ........................................................................2-6
2-10-4 19-in. Rack Assembly and Installation Instructions .................................................................... 2-7
Section 3 Operation
3-1 Home Screen ............................................................................................................................. 3-1
3-2 Operating Controls
.....................................................................................................................3-2
3-2-1 Primary Functions ......................................................................................................................3-2
3-2-2 Key Functions ............................................................................................................................3-3
3-3 Navigation ..................................................................................................................................3-4
3-3-1 Display or Instrument Configuration Screen Selection...............................................................3-5
3-3-2 User Interface Screen Map........................................................................................................ 3-6
3-4 Process Values (PV) and Setpoints (SP) ................................................................................... 3-7
3-4-1 PV-SP Measures and Status......................................................................................................3-7
3-4-2 PV-SP Configuration ..................................................................................................................3-8
3-4-3 Value Programming ................................................................................................................... 3-9
3-5 Global Settings.......................................................................................................................... 3-12
3-5-1 Global Information.....................................................................................................................3-12
3-5-2 Global System Power................................................................................................................ 3-13
3-5-3 Global Control Services ............................................................................................................ 3-14
3-5-4 Global Communication..............................................................................................................3-15
3-6 Instrument Configuration and Control ....................................................................................... 3-15
3-6-1 Rate (Setpoint) Control ............................................................................................................. 3-15
Model 0254
ii
Contents
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Paragraph Page
Number Number
3-6-2 PV Configuration ....................................................................................................................... 3-17
3-6-3 Batch Control ............................................................................................................................3-18
Appendix A Engineering Units
Available Engineering Units .......................................................................................................A-1
Appendix B Blending Examples
Examples 1 through 6 ................................................................................................................B-1
Appendix C Serial Communications Protocol
Contents
C-1 Overview ....................................................................................................................................C-1
C-2 Communication Settings.............................................................................................................C-1
C-2-1 RS-232 Port Settings..................................................................................................................C-1
C-2-2 Hyperterminal Setup ...................................................................................................................C-2
C-3 Serial Command Organization ....................................................................................................C-4
C-3-1 Command Structures ..................................................................................................................C-4
C-3-2 Command Addressing................................................................................................................. C-5
C-4 Command Operation ...................................................................................................................C-6
C-4-1 General Commands .................................................................................................................... C-6
C-4-1-1 Command Synchronize .............................................................................................................. C-6
C-4-1-2 Menu Command .........................................................................................................................C-6
C-4-1-3 View Programmed Channel Port Values .....................................................................................C-7
C-4-1-4 Identify Command .......................................................................................................................C-7
C-4-1-5 Message Serial Character Pacing Controls ................................................................................ C-8
C-4-1-6 Serial Message Error Control......................................................................................................C-9
C-4-2 Channel Input Port Commands .................................................................................................. C-10
C-4-2-1 Measured Channel Values Command ........................................................................................ C-10
C-4-2-2 Send Channel Input Port Programmed Values ........................................................................... C-11
C-4-2-3 Program Channel Input Port Values ........................................................................................... C-12
C-4-2-4 Clear Accumulated Values ......................................................................................................... C-12
C-4-3 Channel Output Port Control Commands ................................................................................... C-12
C-4-3-1 Channel Control Output Port Values .......................................................................................... C-12
C-4-3-2 Batch and Blend Control Commands ......................................................................................... C-13
C-4-4 Global Settings Services............................................................................................................ C-15
C-4-4-1 Global Setting V
alues ................................................................................................................C-15
C-4-4-2 Set Global Settings Values........................................................................................................ C-15
C-4-5 Communication Message Basics ..............................................................................................C-15
C-4-5-1 Message Structure .................................................................................................................... C-16
C-4-5-2 Message Format .......................................................................................................................C-16
C-5 Serial Value Programming ......................................................................................................... C-17
C-5-1 Read a Programmed Value ........................................................................................................C-17
C-5-2 Program a New Value ................................................................................................................C-18
C-5-3 Channel Input Port Values ......................................................................................................... C-19
C-5-4 Channel Output Port Values ...................................................................................................... C-21
C-5-5 Global Setting Values ................................................................................................................ C-22
iii
Contents
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Model 0254
Appendix D CE Certification
Essential Instructions .................................................................................................................D-1
Warranty, Local Sales/Service Contact Information...................................................................... Back Cover
Figures
Figure Page
Number Number
1-1 Model 0254 Signal Wiring ..........................................................................................................1-6
2-1 Panel Mount Cut-Out Dimensions ............................................................................................. 2-4
2-2 Table Top Stand Mount Installation ............................................................................................ 2-5
2-3 Power Supply Bracket and Bezel ............................................................................................... 2-8
2-4 Rack Installation.........................................................................................................................2-9
3-1 Home Screen.............................................................................................................................3-1
3-2 User Interface Screen Map ........................................................................................................ 3-6
Tables
Table Page
Number Number
3-1 Display Home Screen Fields......................................................................................................3-2
Model 0254
iv
Contents
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
THIS PAGE WAS
INTENTIONALLY
LEFT BLANK
1-1
Model 0254
Section 1 Introduction
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
1-1 Description
The Brooks Model 0254 is a versatile full-featured measurement and
control process instrument, with precision multiple channel capabilities.
The architecture supports a wide range of operating capabilities organized
to meet the requirements of nearly any high-accuracy measurement and
control application.
This system suits applications requiring advanced multiple channel
operation. It provides control services including rate, batch, and
proportional blending.
The system is all digital, based on microcomputer technology. There is no
power switch to be left off, no feature selection switches or jumpers, and
no analog trimming potentiometers.
It employs surface mount technology and incorporates floating-point digital
signal processing arithmetic. Built-in test capabilities enable simple
installation, and extensive self-tests ensures long-term operating reliability.
The rugged package is modular, providing an ideal solution for use in
demanding applications and tough environments. Options are available for
panel mount, rack mount, and desktop.
1-1-1 Architecture
The system is built on the basics of a multiple-port based architecture,
modularly organized, which are combined as channels, able to meet
present and future instrumentation requirements.
Channels provide for value measurements and quantity accumulation, rate
measurements, scalar measurements, supported by value scaling for
process voltage and current process signals. Signal acquisition uses digital
signal processing rejecting ambient noise and interference.
1-1-2 Communication
The communication facilities provide for data acquisition, command, and
control functions, supporting configuration programming and information
gathering.
1-1-3 Process Controls
Totalizer, batch, and proportional blending are readily achieved using the
front panel key controls or serial commands.
1-2
Model 0254
Section 1 Introduction
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
1-1-4 Operator Controls and Indicators
Operation may be completely controlled from its integral eight-key pad
used to view operation and programmed operating values. Key activations
are single touch with hold-and-repeat capability. The front panel key pad
provides splash proofing and environment protection.
The 0254 primary indicator is its large back lit liquid crystal graphic display
- visible at a distance even in low light conditions - to view values, support
programming operations, and indicate process state information. A user-
selectable audio indicator annunciates key activations.
1-1-5 Diagnostic Capabilities
Powerful automatic diagnostics support easy installation and ensure a
trouble-free operating life. Tests include memory facility scans, channel
input-output status, and communication status.
1-2 Specifications
Programmable Control and Measurement Functions
Four independent channels.
Rate, Batch, Blend, Measure: 0.000 to 999999
Blend Ratio Percent: 0.000 to 999.999
Totalizer: 0.000 to 19,999,999,999
Gas Correction Factor: 0.000 to 999.999
Measurement Engineering Units: Refer to Appendix A.
Time Base: day, hrs, min, sec, none
Decimal Point : X. to X.XXX
Valve Override: Normal, Open, Closed
Input and Output Signal Selection: Volts or mA, independent, mix or match
Input and Output Full Scale Setting: Independent, 0.000 to 999,999
Setpoint Source: Keypad or RS232
1-3
Model 0254
Section 1 Introduction
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Measurement and Setpoint Accuracy
0.090% of Full Scale
Electrical Specifications
Input Electrical Characteristics
Voltage Input: 0–5, 0-10, 1–5, 2–10 V
Volts Input Impedance: 10 K Ohms
Current Input: 0–20, 4–20 mA
Current Input Impedance: 100 Ohms
Output Electrical Characteristics
Voltage Output: 0–5, 0–10, 1–5, 2–10 V
Voltage Output Load: 2 K Ohms minimum
Current Output: 0–20, 4–20 mA
Current Output Load: 0–375 Ohms
Voltage Compensation
Voltage I/O signals are compensated to correct for voltage drop in the power
return circuit of the attached wiring
Valve Override (VOR) Output
Normal: Open circuit
Open: 8.0 V @ 4 mA
Closed: –4.0 V @ 4 mA
Channel Connectors (4)
15-pin female D
Provides signal and power to connected devices
Serial Port
EIA-TIA232D full duplex D9S Load 4.7 K max
Baud Rate = 9600
1-4
Model 0254
Section 1 Introduction
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Power for Model 0254 and Connected Equipment
Power Supply Requirements
Required: 12–24 Vdc, 2 Amps
Additional permitted: –15 Vdc, 1 Amp DC +/- 5%
Supply voltage flucuations up to +/- 10%
Max current draw per channel : 400 mA
Instrument power draw: 0.8 W
Power Connector: 9-pin female D, 5 A rated
Optional Power Supply Modules (available from Brooks)
• +15 Vdc/2 A, –15 Vdc/1 A, 100–240 Vac, 47–63 Hz
• +24 Vdc/2.5 A, 100–240 Vac, 47–63 Hz
Graphic Display
8-line x 40-character LCD display with backlight
Programmable Display Configuration
Two lines per channel
Line 1, Process Variable: Rate, Total, or Signal
Line 2, Setpoint: Rate, Batch, Blend, or Signal
Off: Unused individual lines can be turned off
Keypad/Display Window
8-Key Metal Dome Tactile with Selectable Audio Beep
Construction: Splash proof and chemically resistant
Hot Keys for Instant Access: Setpoint (rate, batch, blend), VOR,
Emergency Stop
Key Functionality: Full operation and programming capability (refer to
Figure 3-2 on p. 3-6)
Global Settings
Settings applied to all four channels
Information
Version/Check Sum: Factory-installed firmware information
Factory Setting: Allows reset of all channels to factory configuration
System Power
Key selectable power down, power up
1-5
Model 0254
Section 1 Introduction
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Control Service Settings
Audio Beep: Selectable audible keystroke, 2 kHz
Leading Zero Suppress: Selectable leading zero suppression on display
Power Setpoint Clear: Selectable memory retention for Setpoint/VOR
Network Address
Unique factory set identifier
Environment
Temperature/Humidity
Operating: 32 to 122°F (0 to 50°C); 0 to 95% non-condensing
Ship Storage: –40 to 185°F (–40 to 85°C); 0 to 95% non-condensing
Warm-up
15 min typ to rated accuracy
Data Reliability
Data Retention: Non-volatile RAM/ROM, 100 year retention
Self-Diagnostics: On power up, memory checksum, communications,
system status, display and keypad operation
Enclosure
Material: ABS Cycolac Resin FR23
Weight: 1.4 lbs (635 g)
Panel Cut-Out: 7.67 x 4.275 in., 0.125 in. radius maximum
(194.8 x 108.6 m, 3.2 m radius maximum); 0.25 in. (6.35 mm) maximum
panel thickness with optional panel mount kit
Certifications
CE Mark: EN61326-1
FCC: Part 15 Class A, Part 68
RoHS: EPD 2002/95/EC, 01Jul2006
UL: UL 61010 Electrical Safety for General Purpose
Indoor Use;
Altitude up to 2000 m;
RATED POLLUTION Degree 2
1-6
Model 0254
Section 1 Introduction
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Mounting Options
Panel Mount Kit: Brackets accept panel thickness up to 0.25 in. (6.35 mm)
Table Top Kit: Weighted base with fixed tilt for easy viewing
Rack Mount Kit: Hardware for mounting the Model 0254 and optional power
supply into 19-in. sub-rack.
Rack Mount Kit with 19-in. sub-rack: 19-in. sub-rack included with Rack
Mount Kit
Retrofit Applications, Model 0152/0154: Rack Mount Kit will adapt the
Model 0254 and power supply to the table top enclosure used for the
Model 0152/0154
Communications
Full communications capability for remote readout setpoint, control,
programming, and data acquisition
1-3 Signal Wiring
Figure 1-1 Model 0254 Signal Wiring
NOTES:
1) The following nodes are directly connected inside the 0254:
Pin 12 (COM PWR) of each DA15S channel
Pin 5 (GND) of the SERIAL PORT connector
Pins 1,2 and 6 (GND) of the +12-24 Vdc 9-Pin SUB D connector
All hex shield screwlocks for all SUB-D connectors
2) For each DA15S channel, Pins 6 (CMD RTN) and 8 (RATE RTN) are directly connected,
are connected via approximately 10KΩ to Pin 12 (COM PWR)
Pins 6 and 8 are not directly connected between channels
3) For all connectors, all unlisted pins are not internally connected
1-7
Model 0254
Section 1 Introduction
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
1-4 Optional Equipment
1-4-1 Mounting Kit Options
The Model 0254 can be mounted using the following kits:
Panel Mount Kit
Table Top Kit
Rack Mount Kit
Rack Mount Kit with 19-in. Rack
1-4-2 Power Supply Options
The Model 0254 supports the following power supply options:
• Factory supplied: +15 Vdc/2 A, –15 Vdc/1 A, 100–240 Vac, 47–63 Hz
• Factory supplied: +24 Vdc/2.5 A, 100–240 Vac, 47–63 Hz
• User supplied: 12–24 Vdc into 9-pin D connector, 5 A rated
1-4-3 Input/Output Pin-Out Adapters
The Model 0254 has four 15-pin female D-channel connectors. The pin
configuration is compatible with standard Brooks 0(4)-20 mA cables.
An optional pin-out adapter kit with four adapters is available for use with
Brooks 0(1)-5 Vdc cables.
1-8
Model 0254
Section 1 Introduction
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
THIS PAGE WAS
INTENTIONALLY
LEFT BLANK
2-1
Model 0254
Section 2 InstallationInstallation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
2-1 General
This section provides installation instructions for the Model 0254 Gas and
Liquid Mass Flow Secondary Electronics device.
2-2 Receipt of Equipment
When the instrument is received, the outside packing case should be
checked for damage incurred during shipment. If the packing case is
damaged, the local carrier should be notified at once regarding his liability.
A report should be submitted to your nearest Product Service Department.
Brooks Instrument
407 W. Vine Street
P.O. Box 903
Hatfield, PA 19440 USA
Toll Free (888) 554 FLOW (3569)
Tel (215) 362 3700
Fax (215) 362 3745
E-mail: BrooksAm@BrooksInstrument.com
www.BrooksInstrument.com
Brooks Instrument Brooks Instrument
Neonstraat 3 1-4-4 Kitasuna Koto-Ku
6718 WX Ede, Netherlands Tokyo, 136-0073 Japan
P.O. Box 428 Tel +81 (0) 3 5633 7100
6710 BK Ede, Netherlands Fax +81 (0) 3 5633 7101
Tel +31 (0) 318 549 300 Email: BrooksAs@BrooksInstrument.com
Fax +31 (0) 318 549 309
E-mail: BrooksEu@BrooksInstrument.com
Remove the envelope containing the packing list. Carefully remove the
instrument from the packing case. Make sure spare parts are not
discarded with the packing materials. Inspect for damaged or missing
parts.
2-3 Recommended Storage Practice
If intermediate or long-term storage of equipment is required, it is
recommended that the equipment be stored in accordance with the
following:
Within the original shipping container.
Stored in a sheltered area, preferably a warm, dry, heated warehouse.
–40 to 185°F (–40 to 85°C); 0 to 95% non-condensing.
Upon removal from storage, a visual inspection should be conducted to
verify the condition of equipment is "as received".
2-2
Model 0254
Section 2 Installation Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
2-4 Return Shipment
Prior to returning any instrument to the factory, contact your nearest Brooks
location for a Return Materials Authorization Number (RMA#). This can be
obtained from one of the following locations:
Brooks Instrument
407 W. Vine Street
P.O. Box 903
Hatfield, PA 19440 USA
Toll Free (888) 554 FLOW (3569)
Tel (215) 362 3700
Fax (215) 362 3745
E-mail: BrooksAm@BrooksInstrument.com
www.BrooksInstrument.com
Brooks Instrument Brooks Instrument
Neonstraat 3 1-4-4 Kitasuna Koto-Ku
6718 WX Ede, Netherlands Tokyo, 136-0073 Japan
P.O. Box 428 Tel +81 (0) 3 5633 7100
6710 BK Ede, Netherlands Fax +81 (0) 3 5633 7101
Tel +31 (0) 318 549 300 Email: BrooksAs@BrooksInstrument.com
Fax +31 (0) 318 549 309
E-mail: BrooksEu@BrooksInstrument.com
2-5 Transit Precautions
To safeguard against damage during transit, transport the instrument to the
installation site in the same container used for transportation from the
factory if circumstances permit.
2-6 Removal from Storage
Upon removal from storage, a visual inspection should be conducted to
verify the condition of the equipment is “as received.”
2-7 Ventilation and Mounting Requirements
Because of the low power consumption of the Model 0254, it does not
have ventilation requirements. However, the ambient temperature
surrounding the Model 0254 should not exeed 122°F (50°C). The optional
power supply modules are also ventilation-free and limited to an ambient
temperature of 122°F (50°C).
2-8 Cleaning Instructions
Do not use cleaning agents other than water because this might affect
color and marking of the equipment.
Use a clean, soft and damp cloth for cleaning.
2-3
Model 0254
Section 2 InstallationInstallation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
2-9 Cable Requirements
For compliance with the EMC directive 89/336/EEC, the equipment has to
be installed with shielded signal cables which are overall completely
screened with a shield of at least 80%. Sub-D connectors used must be
shielded with a metal shield. The cable screen should be connected to the
metal shell and shielded at both ends over 360°. The shield should be
terminated to earth ground.
The optional power supplies available from Brooks are always supplied
with a power cord that meets all agency certifications and has a protective
conductor for grounding purposes. Any replacement cords must have
similar construction and be certified by a recognized nartional test
laboratory.
2-10 Installation Instructions
Do not use cleaning agents other than water because this might affect
color and marking of the equipment.
2-10-1 Panel Mount Installation Instructions
1. Cut a hole according to the cut-out dimensions shown in Figure 2-1 on
p. 2-4.
2. Pass the Model 0254 enclosure through the cut-out.
3. Position one of the two brackets included in the Panel Mount kit on the
side of the enclosure, while aligning the two holes at the top and bottom
of the bracket flange with the holes at the corners of the back of the
enclosure. Secure the bracket to the enclosure with two of the screws
provided in the kit.
Do not over-tighten these attachment screws.
CAUTION
!
4. Secure the second bracket to the other side of the enclosure.
5. Tighten the panel mounting screws in the middle of both brackets to
secure the enclosure to the panel.
Do not over-tighten these attachment screws.
CAUTION
!
6. Connect the power supply cable with the D-connector to the power
connector on the back of the enclosure.
7. Connect the power supply AC cord to a power outlet.
8. Connect the cables to the RS-232 and Channel connectors, as
approprate for the application.
2-4
Model 0254
Section 2 Installation Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
Figure 2-1 Panel Mount Cut-Out Dimensions
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124

Brooks 0254 Instrucțiuni de utilizare

Categorie
Măsurare, testare
Tip
Instrucțiuni de utilizare