Renishaw FCR25 Manualul utilizatorului

Tip
Manualul utilizatorului

Acest manual este potrivit și pentru

SP25M user's guide
Document part number: H-1000-5104-06-D
SP25M user's guide
www.renishaw.com
Issued 10 2019
1
General information
© 2005 ‐ 2019 Renishaw plc. All rights reserved.
This document may not be copied or reproduced in whole or in part, or transferred to any other media or language, by any means, without
the prior written permission of Renishaw.
The publication of material within this document does not imply freedom from the patent rights of Renishaw plc.
Disclaimer
RENISHAW HAS MADE CONSIDERABLE EFFORTS TO ENSURE THE CONTENT OF THIS DOCUMENT IS CORRECT AT THE DATE OF
PUBLICATION BUT MAKES NO WARRANTIES OR REPRESENTATIONS REGARDING THE CONTENT. RENISHAW EXCLUDES LIABILITY,
HOWSOEVER ARISING, FOR ANY INACCURACIES IN THIS DOCUMENT.
Trademarks
RENISHAW, the probe symbol used in the RENISHAW logo and REVO are registered trademarks of Renishaw plc in the United Kingdom and
other countries. apply innovation and names and designations of other Renishaw products and technologies are trademarks of Renishaw
plc or its subsidiaries.
All brand names and product names used in this document are trade names, service marks, trademarks, or registered trademarks of their
respective owners.
All trademarks and trade names are acknowledged.
WEEE
The use of this symbol on Renishaw products and/or accompanying documentation indicates that the product should not be mixed with the
general household waste upon disposal. It is the responsibility of the end user to dispose of this product at a designated collection point for
waste electrical and electronic equipment (WEEE) to enable reuse or recycling. Correct disposal of this product will help save valuable
resources and prevent potential negative effects on the environment. For more information, please contact your local waste disposal service
or Renishaw distributor.
Warranty
Renishaw plc warrants its equipment for a limited period (as set out in our Standard Terms and Conditions of Sale) provided that it is installed
exactly as defined in associated Renishaw documentation.
Prior consent must be obtained from Renishaw if non-Renishaw equipment (e.g. interfaces and/or cabling) is to be used or substituted. Failure
to comply with this will invalidate the Renishaw warranty.
Claims under warranty must be made from authorised service centres only, which may be advised by the supplier or distributor.
SP25M user's guide
www.renishaw.com
Issued 10 2019
2
Care of equipment
Renishaw probes and associated systems are precision tools used for obtaining precise measurements and must therefore be treated with
care.
Changes to Renishaw products
Renishaw reserves the right to improve, change or modify its hardware or software without incurring any obligations to make changes to
Renishaw equipment previously sold.
Packaging
To aid end user recycling and disposal the materials used in the different components of the packaging are stated here:
Packaging component Material 94/62/EC code 94/62/EC number
Outer box Non-corrugated fibreboard PAP 21
Outer box Corrugated fibreboard PAP 20
Storage box Polypropylene PP 5
Packing foam Low density polyethylene LDPE 4
Patents
Features of the SP25M system and associated products, equipment and techniques are the subjects of one or more of the following patents
and patent applications:
EP1086352 JP4062515 US6772527
EP1368615 JP4726303 US6909983
EP1505362 JP5210536 US7146741
SP25M user's guide
www.renishaw.com
Issued 10 2019
3
Product compliance
EU declaration of conformity
Contact Renishaw plc or visit www.renishaw.com/EU for the full EU declaration.
EMC conformity
This equipment must be installed and used in accordance with this installation guide. This product is intended for industrial use only and
should not be used in a residential area or connected to a low voltage power supply network which supplies buildings used for residential
purposes.
FCC (USA only)
Information to user (47 CFR 15.105)
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These
limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to
cause harmful interference, in which case you will be required to correct the interference at your own expense.
Information to user (47 CFR 15.21)
The user is cautioned that any changes or modifications not expressly approved by Renishaw plc or authorised representative could void the
user's authority to operate the equipment.
Equipment label (47 CFR 15.19)
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
1. This device may not cause harmful interference.
2. This device must accept any interference received, including interference that may cause undesired operation.
SP25M user's guide
www.renishaw.com
Issued 10 2019
4
REACH regulation
Information required by Article 33﴾1﴿ of Regulation ﴾EC﴿ No. 1907/2006 ﴾“REACH”﴿ relating to products containing substances of very high
concern (SVHCs) is available at:
www.renishaw.com/REACH
China RoHS
Contact Renishaw plc or visit www.renishaw.com/ChinaRoHS for the full China RoHS tabulation.
SP25M user's guide
www.renishaw.com
Issued 10 2019
5
International safety instructions
BG ‐ ПРЕДУПРЕЖДЕНИЕ
Моля, обърнете на приложение 1 и прочетете инструкциите за безопасност на вашия собствен език, преди за разопаковате и
монтирате този продукт.
CZ ‐ VÝSTRAHA
Před rozbalením a instalací tohoto výrobku si přečtěte bezpečnostní pokyny ve vlastním jazyce uvedené v příloze 1.
DA - ADVARSEL
Læs sikkerhedsinstrukserne i Appendix 1 FØR udpakning og installation af dette produkt.
DE - WARNHINWEIS
Bevor Sie dieses Produkt auspacken und installieren, konsultieren Sie bitte Anhang 1 und lesen Sie die Sicherheitshinweise in Ihrer Sprache.
EL ‐ ΠΡΟΕΙΔΟΠΟΙΗΣΗ
Γυρίστε στο Κεφάλαιο 1 και διαβάστε τις οδηγίες ασφαλείας στη δική σας γλώσσα προτού ανοίξετε αυτό το προϊόν για να το
εγκαταστήσετε.
EN - WARNING
Before unpacking and installing this product, please consult Appendix 1 and read the safety instructions in your language.
ES - ADVERTENCIA
Consulte el apéndice 1 y lea las instrucciones de seguridad en su idioma antes de desempaquetar e instalar este producto.
ET - HOIATUS
Palun vaadake 1. lisa ning lugege enne selle toote lahtipakkimist ja paigaldamist ohutusjuhend läbi.
FI - VAROITUKSIA
Lue liitteessä 1 olevat omalla kielelläsi kirjoitetut turvaohjeet ennen tämän tuotteen pakkauksen avaamista ja asentamista.
FR - AVERTISSEMENT
Consulter l'annexe 1 et les instructions de sécurité dans votre propre langue avant de déballer et d'installer ce produit.
SP25M user's guide
www.renishaw.com
Issued 10 2019
6
GA - RABHADH
Téigh chuig aguisín 1 agus déan na treoracha sábháilteachta a léamh i do theanga féin le do thoil sula ndéantar an táirge seo a dhíphacáil
agus a shuiteáil.
HR - NAPOMENA
Prije nego što proizvod izvadite iz ambalaže i ugradite ga, otvorite Prilog 1 i pročitajte sigurnosne upute na svom jeziku.
HU – FIGYELMEZTETÉS
A termék kicsomagolása és telepítése előtt olvassa el az 1. számú függelékben található, az Ön anyanyelvén hozzáférhető biztonsági
utasításokat.
IT - AVVISO
Prima di aprire ed installare questo prodotto, leggere le istruzioni di sicurezza nella vostra lingua riportate nell'Appendice 1.
JA ‐ 警告
この製品を箱から取り出し設置する前に、付録 1 に記載された安全性に関する注意書きをお読みください。
LT – ĮSPĖJIMAS
Prieš išpakuodami ir įdiegdami produktą, turite grįžti prie 1 priedo ir perskaityti nurodymus dėl saugos savo kalba.
LV – BRĪDINĀJUMS
Pirms šī izstrādājuma izsaiņošanas un uzstādīšanas izskatiet 1. pielikumā sniegtās drošības instrukcijas savā valodā.
MT - TWISSIJA
Jekk jogħġbok mur f'appendiċi 1 u aqra l‐istruzzjonijiet tas‐sigurtà fil‐lingwa tiegħek qabel ma toħroġ dan il‐prodott mill‐ippakkjar u
tinstallah.
NL - WAARSCHUWING
Ga naar appendix 1 en lees de veiligheidsinstructies in uw eigen taal, voordat u dit product uitpakt en installeert.
PL ‐ OSTRZEŻENIE
Przed rozpakowaniem i zainstalowaniem tego produktu prosimy o zapoznanie się z Dodatkiem 1 i przeczytanie zaleceń dotyczących
bezpieczeństwa w danym języku.
PT ‐ ADVERTÊNCIA
Você deve retornar ao Anexo 1 e ler as instruções de segurança em seu idioma antes de desembalar e instalar este produto.
SP25M user's guide
www.renishaw.com
Issued 10 2019
7
RO - AVERTISMENT
Înainte de a desface ambalajul şi a instala acest produs, vă rugăm să căutaţi Anexa 1 şi să citiţi cu atenţie instrucţiunile de siguranță, în limba
română.
SK ‐ VÝSTRAHA
Pred rozbalením a inštaláciou tohto produktu si pozrite prílohu 1 a prečítajte si bezpečnostné pokyny vo vašom jazyku.
SL - OPOZORILO
Preden izdelek vzamete iz embalaže in ga vgradite, odprite Prilogo 1 in preberite varnostna navodila v svojem jeziku.
SV - VARNING
Gå till bilaga 1 och läs säkerhetsinstruktionerna på ditt eget språk innan du packar upp och installerar denna produkt.
TW ‐ 警告
在拆開和安裝本品之前,請翻頁至附錄 1 讀母語的安全指示。
中文警告
在拆包和安装本品之前,翻到附1阅读中文版安全明。
SP25M user's guide
www.renishaw.com
Issued 10 2019
8
General safety recommendations
CAUTION: Before unpacking and installing the SP25M probe system, the user should carefully read the safety instructions below
and ensure that they are followed at all times by all operators using the probe system. Use of controls or adjustments, or
performance of procedures other than those specified herein may result in hazardous infra red radiation exposure.
Operators must be trained in the use and application of the SP25M probe system and accompanying products, in the context of the
machine it is fitted to, before being allowed to operate that machine.
NOTE: References are made below to features indicated [†] [‡] [♦] on the illustrations shown below. Please ensure that you clearly
understand all safety instructions. Familiarisation with the SP25M system components, as shown in the following sections is
recommended:
System components overview
Schematic diagram of probe system components
Schematic diagram of FCR25 flexible change rack
Schematic diagram of FCR25 TC flexible change rack
The SP25M probe system has mechanical overtravel protection provided in the probe +Z axis, by a fixed bumpstop. The machine
control system must therefore be able to stop the motion of the machine, in this axis of the probe, before the bumpstop is reached.
If this is not the case, the user must wear eye protection during operation in case of stylus breakage.
Care should be taken to ensure that the optical windows ﴾indicated [♦]﴿, located on both body and module, do not become damaged as they
are made of glass and could cause injury.
CAUTION: Permanent magnets are used in some components of the SP25M system and associated products. It is important to
keep them away from items which may be affected by magnetic fields, e.g. data storage systems, pacemakers and watches etc.
SP25M user's guide
www.renishaw.com
Issued 10 2019
9
LED safety
The SP25M body contains embedded high power LED sources ﴾indicated [†]﴿ which emit invisible infra‐red radiation. These sources are
exposed when an SM25-# or TM25-20 module is not attached.
Removing the module breaks two sets of interlock switch contacts ﴾indicated [‡]﴿ to automatically switch off the LED power and assure user
safety.
At suitable intervals, the interlock contacts should be inspected and checked to ensure that they are clean and free from airborne
contamination such as dust, debris or swarf. In unlikely circumstances, such contamination could cause a short circuit of the pins and thus
increase the risk of sending power to the LEDs, without a module being attached. Never connect conducting objects to, or between, the
contacts. Follow the cleaning instructions in the Maintenance section.
Before inspecting, always remove the SP25M body from the probe head.
In the event of serious damage to, or a rupture of, any part of the SP25M body or scanning module outer casing, IMMEDIATELY disconnect
power source, remove and do not attempt to re-use the parts, and contact your supplier for advice.
Safety illustrations
These diagrams show features, indicated [†] [‡] [♦] which are referred to within these safety recommendations.
SP25M probe body
End view showing kinematic joint to module
SP25M kinematic joint to module
SM25 scanning modules
End view showing kinematic joint to body
SP25M kinematic joint to body
SP25M user's guide
www.renishaw.com
Issued 10 2019
10
Introduction
NOTE: Before unpacking and installing the SP25M probe system, please read the safety information and ensure that they are
followed at all times during system installation and operation.
This document is intended to provide help and guidance during initial installation, integration and subsequent use of the of the SP25M probe
system. It is assumed that the system will be installed on a coordinate measuring machine using a controller system that has had full SP25M
system integration in the form of a dedicated Renishaw controller or an AC3 (ISA bus) analogue interface PC card.
SP25M is a compact and versatile probe system that provides both scanning capability for form measurement or reverse engineering and
touch-trigger probing for geometric size and position measurement.
The SP25M system provides highly accurate scanning measurement with stylus lengths from 20 mm to 400 mm as well as full TP20 touch-
trigger probe module compatibility to suit a range of applications.
The SP25M system uses the Renishaw autojoint to enable repeatable mounting to the PH10M PLUS, PH10MQ PLUS and PH6M probe heads.
Multiwired extension bar mounting of the SP25M system is also possible to extend the reach and enable more access to part features.
The range of SM25 scanning modules are designed to cover specific stylus length ranges whilst maintaining excellent accuracy performance.
The system can be readily expanded to further increase the scanning range, allow TP20 touch-trigger measurement or utilise an automatic
changer system.
SP25M user's guide
www.renishaw.com
Issued 10 2019
11
Design principles of the SP25M system
The SP25M design features an optical transducer sensor system which is located within the SP25M probe body. The design also features a
spring pivoting motion system which is located within the SM25 scanning modules.
Two infrared light emitting diodes (IREDs) mounted in the SP25M probe body project invisible infrared beams on to two mirrors mounted on
the pivoting motion structure within the SM25 scanning module. The mirrors reflect the beam back to the SP25M probe body where their
change in position is detected by two position sensitive devices (PSDs) when the stylus is deflected. The PSDs provide signal outputs in the
three probe axes; P, Q and R.
NOTE: The SP25M probe system does not have a fixed rate, gain or resolution and the P, Q and R probe axis outputs are non-linear
and non-orthogonal. These outputs are converted to X, Y and Z signals by the calibration routine. Renishaw offer support and
advice on scanning calibration algorithms suited to SP25M.
The SM25 scanning modules are designed to provide an optimised level of accuracy and contact force over a specified stylus range. This
minimises the reduction in performance seen in other types of scanning probes as the stylus length increases.
SP25M user's guide
www.renishaw.com
Issued 10 2019
12
System components overview
System overview
The modular design of the SP25M system enables the flexibility for optimum configuration by the user. Please refer to the schematic
diagrams showing the SP25M system components in the SP25M installation section of this document.
SP25M probe body
The SP25M probe body houses the optical transducer system and attaches to PH10M PLUS, PH10MQ PLUS or PH6M using a Renishaw
autojoint connection.
SM25 scanning modules
There are five scanning modules in the SM25 range which enable accurate scanning measurements at different stylus lengths. SM25-1,
SM25-2, SM25-3 and SM25-4 are recommended for use with linear stylus arrangements and SM25-5 is recommended for use with non-
linear and star stylus arrangements.
SH25 stylus holders
The range of SH25 stylus holders enable accurate scanning measurements with effective stylus lengths ranging from 20 mm to 400 mm.
SH25-1, SH25-2, SH25-3 and SH25-4 are recommended for use with linear stylus arrangements. SH25-2A, SH25-3A, SH25-4A and SH25-5
are recommended for use with non-linear and star stylus arrangements.
TM25-20 adaptor module
For rapid touch-trigger measurement, the TM25-20 adaptor module may be used which provides full compatibility with the whole range of
TP20 touch-trigger probe modules.
SP25M user's guide
www.renishaw.com
Issued 10 2019
13
FCR25 flexible change rack
Automated and repeatable changing of the SM25 scanning modules, SH25 stylus holders and TM25-20 is possible with FCR25. This triple
port system is designed for mounting on the Renishaw MRS rail system.
FCR25-L3 and FCR25-L6 leg mounted flexible change racks
These variants of the FCR25 flexible change rack and designed to be mounted to the bed of the CMM with the integrated leg assembly.
FCR25 TC change rack
The FCR25 TC provides automated and repeatable changing of SM25 scanning modules whilst keeping the ports at the same operating
temperature as the SP25M probe body. This ensures that there is no difference in temperature between the SP25M probe body and the
SM25 scanning modules eliminating thermal variation and providing optimum metrology.
AC3 analogue interface PCB card
An ISA bus card to enable integration of the SP25M system with a CMM manufacturer's own controller.
SP25M user's guide
www.renishaw.com
Issued 10 2019
14
Probe system components
* Other variants exist
Scanning
The SM25 range of scanning modules are only compatible with the respective range of numbered stylus holders. Each scanning module is
optimised to provide reliable and accurate measurement at a range of working lengths.
Touch-trigger probing
The TM25-20 adaptor module provides touch-trigger probing capability by utilising the full range of TP20 probe modules. Please visit
www.renishaw.com/cmmguides for more information on the Renishaw TP20 touch-trigger probe system.
SP25M user's guide
www.renishaw.com
Issued 10 2019
15
FCR25 flexible change rack system
FCR25 is a flexible change rack system that can be either mounted to an MRS rail or supplied with a standalone leg which can be fixed to the
bed of the CMM. FCR25 provides automatic and repeatable changing of SP25M system scanning modules and stylus holders.
There are a number of port inserts that are required to interchange SH25 stylus holders and also TP20 modules. The PA25-SH port insert is
used to allow the changing of the whole range of SH25 stylus holders, the PA25-20 should be used to allow TP20 modules to be changed
using FCR25.
FCR25 can be fixed to an MRS rail in multiples, allowing more numbers of ports for housing and changing probe system components. There
are three port and six port versions of the standalone FCR25 also available.
SP25M user's guide
www.renishaw.com
Issued 10 2019
16
FCR25 TC temperature controlled flexible
change rack
FCR25 TC is similar in function to the standard FCR25, the main difference being that temperature compensation is provided to optimise
scanning module metrology performance.
No port inserts are compatible with FCR25 TC which means that only the SM25 scanning module components of the SP25M system can be
changed using this rack.
FCR25 TC can be mounted directly to an MRS rail, there is also a three port standalone version that can be fixed directly to the bed on the
CMM.
SP25M user's guide
www.renishaw.com
Issued 10 2019
17
SP25M specification
Characteristic Requirements
Probe attributes Scanning with 3-axis measurement (X, Y, Z)
Touch-trigger probing using TP20 modules
Measurement range ±0.5 mm deflection in all directions in all orientations
Overtravel range
X, Y ±2 mm
+Z 1.7 mm
-Z 1.2 mm
Resolution Capable of <0.1 μm
Spring rate Nominally 0.6 N/mm - when using a module's shortest specified
stylus
Nominally 0.2 N/mm - when using a module's longest specified
stylus
Probe dimensions Ø25 mm × length dependant on module used
Weight
SP25M body 65 g (2.29 oz)
SM25-1 scanning module 35 g (1.23 oz) (including SH25-1 but excluding stylus)
SM25-2 scanning module 40 g (1.41 oz) (including SH25-2 but excluding stylus)
SM25-3 scanning module 49 g (1.73 oz) (including SH25-3 but excluding stylus)
SM25-4 scanning module 71 g (2.50 oz) (including SH25-4 but excluding stylus)
SM25-5 scanning module 45 g (1.59 oz) (including SH25-5 but excluding stylus)
TM25-20 adaptor module 40 g (1.41 oz) (including TP20 STD module, but excluding stylus)
Effective stylus length range
(Always observe the specified stylus range for the scanning module
being used. Use Renishaw's M3 range of styli.)
SM25-1 + SH25-1 EWL 20 mm - 50 mm (0.79 in - 1.97 in) using 20 mm - 50 mm stylus
range
SM25-2 + SH25-2 / SH25-2A EWL 50 mm - 105 mm (1.97 in - 4.14 in) using 20 mm - 75 mm
stylus
SM25-3 + SH25-3 / SH25-3A EWL 120 mm - 200 mm (4.73 in - 7.88 in) using 20 mm - 100 mm
stylus
SM25-4 + SH25-4 / SH25-4A EWL 220 mm - 400 mm (8.67 in - 15.75 in) using 20 mm - 200 mm
stylus
SM25-5 + SH25-5 EWL 20 mm - 100 mm (0.79 in - 3.94 in) using 20 mm - 100 mm
stylus
SP25M user's guide
www.renishaw.com
Issued 10 2019
18
Mounting Multiwired autojoint - compatible with PH10M PLUS / PH10MQ
PLUS and PH6M probe heads, extension bars and ACR1 / ACR3
sensor changers
Crash protection
±X, ±Y, ‐Z Via break out of module or stylus holder
+Z Via integral bump-stop design
Signal outputs Non-linear and non-orthogonal analogue outputs - rate gain and
resolution are not fixed
Signal outputs Non-linear and non-orthogonal analogue outputs - rate gain and
resolution are not fixed
Power supply +12 V ﴾±5%﴿, ‐12 V ﴾+10% / +8%﴿, +5% ﴾+10% / ‐13%﴿ dc at probe
Probe calibration Requires that non-linear, third order polynomial calibration method
is used
Change rack options
FCR25 Three port unit which mounts on MRS
FCR25-L3 Three port standalone rack version
FCR25-L6 Six port standalone rack version
FCR25 TC Three port unit which mounts on MRS and warms scanning module
FCR25 TC-L3 Three port standalone rack version
Interface options
UCC S3
UCC2-2 systems
OEM controller systems AC3 interface card
Interface for TP20 also required if applicable
SP25M user's guide
www.renishaw.com
Issued 10 2019
19
Dimensional information
SM25-1
SH25-1
SM25-2
SH25-2
SM25-2
SH25-2A
SM25-3
SH25-3
SM25-3
SH25-3A
SM25-4
SH25-4
SM25-4
SH25-4A
SM25-5
SH25-5
Range (mm) 20 - 50 20 - 75 20 - 75 20 - 100 20 - 100 20 - 200 20 - 200 20 - 100
EWL (mm) 20 - 50 50 - 105 50 - 105 120 - 200 120 - 200 220 - 400 220 - 400 20 - 100
SP25M user's guide
www.renishaw.com
Issued 10 2019
20
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133

Renishaw FCR25 Manualul utilizatorului

Tip
Manualul utilizatorului
Acest manual este potrivit și pentru

în alte limbi