Microlife BP A6 BT Manual de utilizare

Categorie
Unități de tensiune arterială
Tip
Manual de utilizare
Microlife AG
Espenstrasse 139
9443 Widnau / Switzerland
www.microlife.com
IB BP A6 BT 4G E-V11 2520
Revision Date: 2020-10-
14
BP
A6 BT
Blood Pressure Monitor
EN
RU
1
9
BG
RO
19
29
CZ
SK
37
45
SL
SR
53
61
HU
HR
69
77
PL
85
Preparation
Sit on a back-supported chair and keep your legs uncrossed. / Сядьте, не
скрещивая ноги, на стул со спинкой. / Седнете на стол с облегалка и не
кръстосвайте краката си. / Așezați-va intr-un scaun cu spatar și nu încrucișați
picioarele / Sedněte si na židli s opěrkou a nepřekřižujte nohy. / Sadnite si na
stoličku s opierkou a neprekrižujte si nohy. / Usedite se na stol s podprtim
hrbtom in nogami na tleh (ne prekrižajte jih). / Sedite na stolicu koja podupire
leđa i nemojte prekrštati noge. / Üljön egy háttámlás székre és lábait ne rakja
keresztbe! / Sjednite na stolicu koja podupire leđa i neka Vam noge ne budu
prekrižene. / Usiądź na krześle wspieranym plecami i nie rozstawiaj nóg.
1.
3.
5.
2.
4.
6.
Fit the cuff closely, but not too tight. / Поместите манжету так, чтобы она
плотно, но не туго прилегала. / Стегнете добре маншета, но не твърде
много. / Fixați manseta ferm dar nu prea stransă / Nasaďte manžetu těsně, ale
ne příliš těsně. / Nasaďte manžetu tesne, ale nie príliš tesne. / Manšeto dobro
namestite, vendar ne pretesno. / Dobro zategnite manžetnu, ali ne previše
stegnuto. / Rögzítse a mandzsettát szorosan, de ne feszítse meg túl erősen! /
Postavite manžetu blizu, ali ne pretijesno. / Zaciśnij mankiet dokładnie, jednak
niezbyt silnie.
Avoid thick or close-fitting garments on the upper arm. / Освободите плечо от
плотной или плотно облегающей одежды. / Избягвайте дебели или плътно
прилепващи дрехи върху горната част на ръката. / Evitați imbracaminte
groasa sau stâmta pe braț / Vyhněte se těsnému nebo přiléhavému oděvu
na horní části paže. / Vyhnite sa tesnému alebo priliehavému odevu na hornej
časti ramena. / AIzogibajte se tesnim oblačilom. / Izbegavajte usku odeću na
nadlaktici. / Kerülje a vastag vagy szoros ruhákat a felkaron! / Izbjegavajte usku
odjeću na nadlaktici. / Unikaj grubej lub dopasowanej odzieży na ramieniu.
Place the artery-mark on the cuff over your artery. / Поместите манжету так,
чтобы значок артерии на манжете находился над артерией руки в сгибе
локтя. / Поставете маркера на маншета върху артерията. / Plasați semnul
artera al manșetei peste artera dumneavoastra / Umístěte manžetu se značkou
tepny nad Vaši tepnu. / Umiestnite manžetu so značkou tepny nad Vašu tepnu.
/ Namestite znak za arterijo, ki je na manšeti, na vašo arterijo. / Postavite oznaku
za arteriju na manžetni preko Vaše arterije. / Helyezze a mandzsetta artériajelzőjét
az artériája fölé! / Postavite oznaku arterije na manžeti preko Vaše arterije. /
Umieść mankiecie na tętnicy.
Position the cuff 1-2 cm above your elbow. / Расположите манжету на
1-2 см выше локтя. / Поставете маншета 1-2 см над лакътя. / Plasați
manșeta cu 1-2 cm deasupra articulatiei / Manžetu umístěte 1-2 cm nad loket.
/ Manžetu umiestnite 1-2 cm nad lakeť. / Namestite manšeto 1-2 cm nad
komolcem. / Postavite manžetnu 1-2 cm iznad Vašeg lakta. / A mandzsetta 1-2
centiméterrel legyen a könyökétől feljebb! / Postavite manžetu 1-2 cm iznad
Vašeg lakta. / Ustaw mankiet na 1-2 cm powyżej łokcia.
Keep your arm still and do not speak during the measurement. / Держите
руку неподвижно и не разговаривайте во время измерения. / Дръжте
ръката си неподвижна и не говорете по време на измерването. / Mějte
ruku uvolněnou a během měření nemluvte. / Majte ruku uvoľnenú a počas
merania nerozprávajte. / Držte ruku uvoľnenú a počas merania nerozprávajte. /
Med merjenjem ne premikajte roke in ne govorite. / Držite ruku mirno i nemojte
pričati tokom merenja. / Ne mozgassa a karját és ne beszéljen mérés közben! /
Držite ruku mirno i nemojte pričati tijekom mjerenja. / Trzymaj rękę nieruchomo i
nie mów podczas pomiaru.
A6 BT
2
1
3
4
65
13
14
15
8
9
19
7
25
27
20
1211
23 24 21
22
18
17
16
26
10
Microlife BP A6 BT Microlife BP A6 BTMicrolife BP A6 BT Guarantee CardBefore each measurement
Name of Purchaser / Ф.И.О. покупателя /
Име на купувача / Numele cumpărătorului / Imię
i nazwisko nabywcy / Jméno kupujícího / Meno
zákazníka / Ime in priimek kupca / Ime i prezime
kupca / Vásárló neve / Ime i prezime kupca
Serial Number / Серийный номер / Сериен
номер / Număr de serie / Numer seryjny /
Výrobní číslo / Výrobné číslo / Serijska števika /
Serijski broj / Sorozatszám / Serijski broj
Date of Purchase / Дата покупки / Дата на
закупуване / Data cumpărării / Data zakupu /
Datum nákupu / Dátum kúpy / Datum nakupa /
Datum kupovine / Vásárlás dátuma / Datum
kupovine
Specialist Dealer / Специализированный дилер /
Специалист дистрибутор / Distribuitor de spe-
cialitate / Przedstawiciel / Specializovaný dealer /
Špecializovaný predajca / Spezializirani trgovec /
Ovlašćeni diler / Forgalmazó / Ovlašteni prodavač
1.
2.
3.
Avoid eating, bathing, smoking or caffeine (approx. 30 min).
Избегайте приема пищи, купания, курения или приема кофеиносодержащих продуктов
(прибл. за 30 минут до измерения).
Избягвайте ядене, къпане, пушене или кофеин (около 30 минути).
Evitați sa mâncați, sa faceți baie, sa fumați, sau sa beți cafea(aproximativ 30 min).
Vyhněte se jídlu, koupání, kouření nebo kofeinu (přibližně 30 minut).
Vyhnite sa jedlu, kúpaniu, fajčeniu alebo kofeínu (približne 30 minút).
Izogibajte se hrani, kopanju in kajenju (približno 30 minut).
Izbegavajte hranu, kupanje, pušenje i kofein (približno 30 minuta).
Kerülje az evést, fürdést, dohányzást vagy koffeinfogyasztást (a megelőző kb. 30 percben)!
Izbjegavajte jesti, kupati se, pušiti i kofein (oko 30 min.)
Unikaj jedzenia, kąpieli, palenia lub kofeiny (około 30 minut).
Avoid activity and relax for 5-10 min.
В течение 5-10 мин. оставайтесь в расслабленное состояние и избегайте физической
нагрузки.
Избягвайте активности и се отпуснете в продължение на 5-10 минути.
Evitați orice activitate si relaxați-vă 5-10 min.
Vyhněte se fyzické aktivitě a odpočívejte 5-10 minut.
Vyhnite sa fyzickej aktivite a odpočívajte 5-10 minút.
Izogibajte se fizični aktivnosti in počivajte 5-10 minut.
Izbegavajte fizičku aktivnost i opustite se tokom 5-10 minuta.
Kerülje a fizikai tevékenységet, pihenjen 5-10 percig!
Izbjegavajte aktivnost i opustite se 5-10 min.
Unikaj aktywności i zrelaksuj się przez 5-10 minut.
Measure before medication intake.
Измеряйте до приема лекарства.
Измерете преди приема на лекарството ви.
Masurați inainte de a va lua medicația.
Měření proveďte před přijetím léku.
Meranie uskutočnite pred užitím liekov.
Merjenje izvedite pred zaužitjem zdravil.
Obavite merenje pre uzimanja leka.
Mérjen a gyógyszer bevétele előtt!
Izvedite mjerenje prije nego što uzmete lijek.
Zmierz przed przyjęciem leku.
1BP A6 BT
EN
Display
Read the instructions carefully before using this device.
Type BF applied part
Intended use:
This oscillometric blood pressure monitor is intended for measuring
non-invasive blood pressure in people aged 12 years or older.
It is clinically validated in patients with hypertension, hypotension,
diabetes, pregnancy, pre-eclampsia, atherosclerosis, end-stage
renal disease, obesity and the elderly.
The device can detect an irregular pulse suggestive of Atrial Fibril-
lation (AF). Please note that the device is not intended to diagnose
AF. A diagnosis of AF can only be confirmed by ECG. The patient
is advised to see a physician.
Dear Customer,
This device was developed in collaboration with physicians and
clinical tests carried out prove its measurement accuracy to be of
a very high standard.*
Microlife AFIBsens is the world’s leading digital blood pressure
measurement technology for the detection of atrial fibrillation (AF)
and arterial hypertension. These are the two top risk factors of
getting a stroke or heart disease. It is important to detect AF and
hypertension at an early stage, even though you may not experi-
ence any symptoms. AF screening in general and thus also with
the Microlife AFIB algorithm, is recommended for people of 65
years and older. The AFIB algorithm indicates that atrial fibrillation
may be present. For this reason, it is recommended that you visit
your doctor when the device gives an AFIB signal during your
blood pressure measurement. The AFIB algorithm of Microlife has
been clinically investigated by several prominent clinical investiga-
tors and showed that the device detects patients with AFIB at a
certainty of 97-100%.
1,2
If you have any questions, problems or want to order spare parts
please contact your local Microlife-Customer Service. Your dealer
or pharmacy will be able to give you the address of the Microlife
dealer in your country. Alternatively, visit the internet at
www.microlife.com where you will find a wealth of invaluable infor-
mation on our products.
Stay healthy – Microlife AG!
* This device uses the same measuring technology as the award
winning «BP 3BTO-A» model tested according to the British and
Irish Hypertension Society (BIHS) protocol.
1
Kearley K, Selwood M, Van den Bruel A, Thompson M, Mant D,
Hobbs FR et al.: Triage tests for identifying atrial fibrillation in
Microlife BP A6 BT
EN
1
START/STOP Button
2
Display
3
M-button (memory)
4
Time Button
5
Cuff Socket
6
AFIB/MAM Switch
7
Traffic Light Indicator
8
USB Port
9
Mains Adapter Socket
AT
Battery Compartment
AK
Cuff
AL
Cuff Connector
AM
- «Backward» Button
AN
+ «Forward» Button
AO
Lock Switch
AP
Systolic Value
AQ
Diastolic Value
AR
Pulse Rate
AS
Battery Display
BT
Cuff Check Indicator
BK
Arm Movement Indicator
BL
MAM Interval Time
BM
Atrial Fibrillation Indicator (AFIB)
BN
AFIB/MAM Mode
BO
Stored Value
BP
Date/Time
BQ
Pulse Indicator
Keep dry
2
primary care: a diagnostic accuracy study comparing single-lead
ECG and modified BP monitors. BMJ Open 2014; 4:e004565.
2
Wiesel J, Arbesfeld B, Schechter D: Comparison of the Microlife
blood pressure monitor with the Omron blood pressure monitor for
detecting atrial fibrillation. Am J Cardiol 2014; 114:1046-1048.
Table of Contents
1. Appearance of the Atrial Fibrillation Indicator for early
Detection (Active only in AFIB/MAM mode)
What is Atrial Fibrillation (AF)?
Who should be screened for Atrial Fibrillation?
Risk factors you can control
2. Using the Device for the First Time
Inserting the batteries
Setting the date and time
Selecting the correct cuff
Selecting standard or AFIB/MAM mode
3. Checklist for Taking a Reliable Measurement
4. Taking a Blood Pressure Measurement
How not to store a reading
How do I evaluate my blood pressure?
5. Data Memory
Viewing the stored values
Clearing all values
6. Battery Indicator and Battery change
Low battery
Flat battery – replacement
Which batteries and which procedure?
Using rechargeable batteries
7. Using a Mains Adapter
8. Bluetooth® Function
9. PC-Link Functions
10. Error Messages
11. Safety, Care, Accuracy Test and Disposal
Safety and protection
Device care
Cleaning the cuff
Accuracy test
Disposal
12. Guarantee
13. Technical Specifications
Guarantee Card (see Back Cover)
1. Appearance of the Atrial Fibrillation Indicator for
early Detection (Active only in AFIB/MAM mode)
This device is able to detect atrial fibrillation (AF). This symbol BM
indicates that atrial fibrillation was detected during the measure-
ment. Please refer to the next paragraph for information regarding
the consultation with your doctor.
Keep the arm still during measuring to avoid false readings.
This device may not or wrongly detect atrial fibrillation in
people with pacemakers or defibrillators.
In the presence of atrial fibrillation the diastolic blood pressure
value may not be accurate.
In the presence of atrial fibrillation using MAM-mode is
recommended for more reliable blood pressure measure-
ment.
What is Atrial Fibrillation (AF)?
Normally, your heart contracts and relaxes to a regular beat. Certain
cells in your heart produce electrical signals that cause the heart to
contract and pump blood. Atrial fibrillation occurs when rapid, disorga-
nized electrical signals are present in the heart’s two upper chambers,
called the atria; causing them to contract irregularly (this is called fibril-
lation). Atrial fibrillation is the most common form of heart arrhythmia.
Information for the doctor on frequent appearance of the
atrial fibrillation indicator
This device is an oscillometric blood pressure monitor that also
analyses pulse irregularity during measurement. The device is
clinically tested.
The AFIB symbol is displayed after the measurement, if atrial
fibrillation occurred during measuring. If the AFIB symbol
appears after having performed a full blood pressure measure-
ment episode (triplicate measurements), the patient is advised to
perform another measurement episode (triplicate measure-
ments). If the AFIB symbol appears again, we recommend the
patient to seek medical advice.
If the AFIB-symbol appears on the screen of the blood pressure
monitor, it indicates the possible presence of atrial fibrillation.
The atrial fibrillation diagnosis however, must be made by a
cardiologist based on ECG interpretation.
3BP A6 BT
EN
It often causes no symptoms, yet it significantly increases your risk of
stroke. You’ll need a doctor to help you control the problem.
Who should be screened for Atrial Fibrillation?
AF screening is recommended for people over 65 years of age, since
the chance of having a stroke increases with age. AF screening is
also recommended for people from the age of 50 years who have
high blood pressure (e.g. SYS higher than 159 or DIA higher than 99)
as well as those with diabetes, coronary heart failure or for those who
have previously had a stroke.
In young people or in pregnancy AF screening is not recommended
as it could generate false results and unnecessary anxiety. In addi-
tion, young individuals with AF have a low risk of getting stroke as
compared to elder people.
Risk factors you can control
Early diagnosis of AF followed by adequate treatment can signifi-
cantly reduce the risk of getting stroke. Knowing your blood pressure
and knowing whether you have AF is the first step in proactive stroke
prevention.
For more information visit our website: www.microlife.com/afib.
2. Using the Device for the First Time
Inserting the batteries
Switch the lock switch AO to «unlock» position. The battery
compartment AT is on the bottom of the device. Insert the batteries
(4 x 1.5 V, size AAA), thereby observing the indicated polarity.
Setting the date and time
1. After the new batteries are fitted, the year number flashes in the
display. You can set the year by pressing the M-button 3. To
confirm and then set the month, press the time button 4.
2. Press the «+» AN or the «-» AM button to set the month. Press
the time button 4 to confirm and then set the day.
3. Follow the instructions above to set the day, hour and minutes.
4. Once you have set the minutes and pressed the time button, the
date and time are set and the time is displayed.
5. If you want to change the date and time, press and hold the time
button down for approx. 3 seconds until the year number starts
to flash. Now you can enter the new values as described above.
Selecting the correct cuff
Microlife offers different cuff sizes. Select the cuff size to match the
circumference of your upper arms (measured by close fitting in the
centre of the upper arm).
Only use Microlife cuffs.
Contact your local Microlife Service if the enclosed cuff AK does
not fit.
Connect the cuff to the device by inserting the cuff connector AL
into the cuff socket 5 as far as it will go.
If you buy a spare Microlife cuff, please remove the cuff
connector AL from the cuff tube AL from the cuff supplied
with the original device and insert this cuff connector into
the tube of the spare cuff (valid for all cuff sizes).
Selecting standard or AFIB/MAM mode
Before each measurement, select standard (single measurement)
or AFIB/MAM mode (automatic triple measurement). In AFIB/MAM
mode, 3 measurements are automatically taken in succession and
the result is then automatically analysed and displayed. Because
the blood pressure constantly fluctuates, a result obtained in this
way is more reliable than when a single measurement is
performed.
To select AFIB/MAM mode, slide the AFIB/MAM switch 6
upwards to position «3» until the MAM-symbol BN appears on
the display. To change to standard mode (single measure-
ment), slide the AFIB/MAM switch downwards to position «1».
The bottom, right hand section of the display shows a 1, 2 or 3 to
indicate which of the 3 measurements is currently being taken.
There is a break of 15 seconds between the measurements.
A count down indicates the remaining time.
The individual results are not displayed. Your blood pressure will
only be displayed after all 3 measurements are taken.
Do not remove the cuff between measurements.
If one of the individual measurements was questionable, a fourth
one is automatically taken.
AF detection is only activated in AFIB/MAM mode.
3. Checklist for Taking a Reliable Measurement
Avoid activity, eating or smoking immediately before the
measurement.
Cuff size for circumference of upper arm
S 17 - 22 cm
M 22 - 32 cm
M - L 22 - 42 cm
L 32 - 42 cm
L - XL 32 - 52 cm
4
Sit down on a back-supported chair and relax for 5 minutes.
Keep the feet flat on the floor and do not cross your legs.
Always measure on the same arm (normally left). It is recom-
mended that doctors perform double arm measurements on a
patients first visit in order to determine which arm to measure in
the future. The arm with the higher blood pressure should be
measured.
Remove close-fitting garments from the upper arm. To avoid
constriction, shirt sleeves should not be rolled up - they do not
interfere with the cuff if they are laid flat.
Always ensure that the correct cuff size is used (marking on the
cuff).
Fit the cuff closely, but not too tight.
Make sure that the cuff is positioned 1-2 cm above the elbow.
The artery mark on the cuff (ca. 3 cm long bar) must lie over
the artery which runs down the inner side of the arm.
Support your arm so it is relaxed.
Ensure that the cuff is at the same height as your heart.
4. Taking a Blood Pressure Measurement
1. Select standard (single measurement) or
AFIB/MAM
mode
(automatic triple measurement): see details in chapter «2.».
2. Press the START/STOP button 1 to start the measurement.
3. The cuff will now pump up automatically. Relax, do not move
and do not tense your arm muscles until the measurement
result is displayed. Breathe normally and do not talk.
4. When the correct pressure is reached, the pumping stops and the
pressure falls gradually. If the required pressure was not reached,
the device will automatically pump some more air into the cuff.
5. During the measurement, the pulse indicator BQ flashes in the
display.
6. The result, comprising the systolic AP and the diastolic AQ blood
pressure and the pulse rate AR is displayed. Note also the
explanations on further display symbols in this booklet.
7. When the device has finished measuring, remove the cuff.
8. Switch off the device. (The monitor does switch off automatically
after approx. 1 min.).
AF detection is only activated in AFIB/MAM mode.
You can stop the measurement at any time by pressing the
START/STOP button (e.g. if you feel uneasy or an
unpleasant pressure sensation).
This monitor is specially tested for use in pregnancy and
pre-eclampsia. When you detect unusual high readings in
pregnancy, you should measure after a short while again
(eg. 1 hour). If the reading is still too high, consult your
doctor or gynecologist.
In pregnancy the AFIB symbol can be ignored.
How not to store a reading
As soon as the reading is displayed press and hold the START/
STOP button 1 until «M» BO is flashing. Confirm to delete the
reading by pressing the M button 3.
How do I evaluate my blood pressure?
The LED traffic light indicator on the left-hand side of the display
7
indicates within which range the measured blood pressure lies. The
value is either within the optimum (green), elevated (yellow) or high
(red) range. The classification corresponds to the following ranges
defined by international guidelines (
ESH, ESC, JSH
). Data in mmHg.
The higher value is the one that determines the evaluation.
Example: a blood pressure value of 140/80 mmHg or a value of
130/90 mmHg indicates «blood pressure too high».
5. Data Memory
This device automatically stores the last 99 measurement values.
Viewing the stored values
Switch the lock switch
AO
to «unlock» position. Press the M-button
3
briefly. The display first shows «
M
»
BO
and «
A
», which stands for the
average of all stored values.
Pressing the «+» AN or the «-» AM button repeatedly enables you
to move from one stored value to another. Press the M-button
again to exit the memory mode.
Pay attention that the maximum memory capacity of 99
memories is not exceeded.
When the 99 memory is full, the
oldest value is automatically overwritten with the 100th
Range Systolic Diastolic Recommendation
1. blood pressure
too high
135 85 Seek medical
advice
2. blood pressure
elevated
130 - 134 80 - 84 Self-check
3. blood pressure
normal
<130 <80 Self-check
5BP A6 BT
EN
value.
Values should be evaluated by a doctor before the
memory capacity is reached – otherwise data will be lost.
Clearing all values
If you are sure that you want to permanently remove all stored
values, hold down the M-button (the device must have been
switched off beforehand) until «CL» appears and then release the
button. To permanently clear the memory, press the M-button
while «CL» is flashing. Individual values cannot be cleared.
Cancel deletion:
press START/STOP button
1
while «
CL
»
is flashing.
6. Battery Indicator and Battery change
Low battery
When the batteries are approximately ¾ empty the battery symbol
AS
will flash as soon as the device is switched on (partly filled battery
displayed). Although the device will continue to measure reliably, you
should obtain replacement batteries.
Flat battery – replacement
When the batteries are flat, the battery symbol AS will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment AT at the back of the device.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
3. To set date and time, follow the procedure described in
«Section 2.».
The memory retains all values although date and time must
be reset – the year number therefore flashes automatically
after the batteries are replaced.
Which batteries and which procedure?
Use 4 new, long-life 1.5 V, size AAA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.
Using rechargeable batteries
You can also operate this device using rechargeable batteries.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
device as they may become damaged (total discharge as a
result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not
intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
7. Using a Mains Adapter
You can operate this device using the Microlife mains adapter
(DC 6V, 600 mA).
Only use the
Microlife
mains adapter available as an original
accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket 9 in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is consumed.
8. Bluetooth
®
Function
This device can be used in conjunction with a smartphone running
the «Microlife Connected Health+» App. The Bluetooth
®
connec-
tion is automatically active after the measurement has been
completed.
1. Activate Bluetooth
®
on your smartphone.
2. Activate the Bluetooth
®
function on the device. Press the «+»
button AN while the device is in stand-by mode. The icon «bt»
starts flashing on the display.
For more detailed information visit www.microlife.com/connect.
9. PC-Link Functions
This device can be used in conjunction with a personal computer
(PC) running the Microlife Blood Pressure Analyzer+ (BPA+) soft-
ware. The memory data can be transferred to the PC by
connecting the monitor via a cable.
If no download-voucher and cable is included download the BPA+
software from www.microlife.com/software and use a USB cable
with a Mini-B 5 pin connector.
During the connection, the device is completely controlled
by the computer.
6
10.Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «Err 3», is displayed.
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
11.Safety, Care, Accuracy Test and Disposal
Safety and protection
Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
This device comprises sensitive components and must be treated
with caution. Observe the storage and operating conditions
described in the «Technical Specifications» section.
Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
The cuffs are sensitive and must be handled with care.
Do not exchange or use any other kind of cuff or cuff connector
for measuring with this device.
Only pump up the cuff once fitted.
Do not use this device close to strong electromagnetic fields such
as mobile telephones or radio installations. Keep a minimum
distance of 3.3 m from such devices when using this device.
Do not use this device if you think it is damaged or notice
anything unusual.
Never open this device.
If the device is not going to be used for a prolonged period the
batteries should be removed.
Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
The measurement results given by this device is not a diag-
nosis. It is not replacing the need for the consultation of a physi-
cian, especially if not matching the patient’s symptoms. Do not
rely on the measurement result only, always consider other
potentially occurring symptoms and the patient’s feedback.
Calling a doctor or an ambulance is advised if needed.
Error Description Potential cause and remedy
«Err 1» Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat the
measurement.*
«Err 2»
BK
Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«Err 3»
BT
No pressure
in the cuff
An adequate pressure cannot be gener-
ated in the cuff. A leak may have occurred.
Check that the cuff is correctly connected
and is not too loose. Replace the batteries
if necessary. Repeat the measurement.
«Err 5» Abnormal
result
The measuring signals are inaccurate
and no result can therefore be displayed.
Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
«Err 6» AFIB/MAM
Mode
There were too many errors during the
measurement in AFIB/MAM mode,
making it impossible to obtain a final
result. Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
«HI» Pulse or cuff
pressure
too high
The pressure in the cuff is too high (over
299 mmHg) OR the pulse is too high
(over 200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«LO» Pulse too
low
The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*
«FL»
Problem with
Bluetooth
®
connection
If «FL» flashes on the display, the device
failed to establish a connection with your
smartphone. Make sure that Bluetooth
®
is activated and repeat the procedure.
7BP A6 BT
EN
Permanently high blood pressure values can damage your
health and must be treated by your doctor!
Always discuss your values with your doctor and tell him/her if
you have noticed anything unusual or feel unsure. Never rely
on single blood pressure readings.
Under no circumstances should you alter the dosages of
drugs or initiate a treatment without consulting your doctor.
Deviations between measurements taken by your doctor or in
the pharmacy and those taken at home are quite normal, as
these situations are completely different.
The pulse display is not suitable for checking the
frequency of heart pacemakers!
If you are pregnant, you should monitor your blood pressure
regularly as it can change drastically during this time.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.
Device care
Clean the device only with a soft, dry cloth.
Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and soap-
suds.
WARNING: Under no circumstances may you wash the
inner bladder!
Accuracy test
We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).
Disposal
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
12.Guarantee
This device is covered by a 5 year guarantee from the date of
purchase. During this guarantee period, at our discretion, Microlife
will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
Transport costs and risks of transport.
Damage caused by incorrect application or non-compliance
with the instructions for use.
Damage caused by leaking batteries.
Damage caused by accident or misuse.
Packaging/storage material and instructions for use.
Regular checks and maintenance (calibration).
Accessories and wearing parts: Batteries, power adapter
(optional).
The cuff is covered by a functional guarantee (bladder tightness)
for 2 years.
Should guarantee service be required, please contact the dealer
from where the product was purchased, or your local Microlife
service. You may contact your local Microlife service through our
website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee
will be granted if the complete product is returned with the original
invoice. Repair or replacement within guarantee does not prolong
or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.
13.Technical Specifications
Operating conditions:
10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
354
g (including batteries)
Dimensions:
160 x 80 x 32
mm
Measuring procedure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range:
20 - 280
mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure
display range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
4 x 1.5 V alkaline batteries; size AAA
Mains adapter DC 6V, 600 mA (optional)
8
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
The Bluetooth
®
word mark and logos are registered trademarks
owned by Bluetooth SIG, Inc. and any use of such marks by Micro-
life Corp. is under license. Other trademarks and trade names are
those of their respective owners.
Battery lifetime: approx. 400 measurements
(using new batteries)
IP Class: IP20
Reference to
standards:
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements
Accessories: 2 years
9BP A6 BT
RU
Дисплей
Перед использованием прибора внимательно прочтите
данное руководство.
Изделие типа BF
Предназначение:
Этот осциллометрический тонометр предназначен для неин-
вазивного измерения артериального давления у людей в
возрасте 12 лет и старше.
Прибор прошел клинические испытания для использования
пациентами с гипертонией, гипотонией, сахарным диабетом,
при беременности, преэклампсии, атеросклерозе, конечной
стадии почечной недостаточности, ожирении и у людей пожи-
лого возраста.
Прибор может обнаруживать неравномерный пульс, хара
к-
терный для мерцательной аритмии (AF). Обратите внимание,
что прибор не предназначен для диагностирования мерца-
тельной аритмии. Диагноз мерцательная аритмия может быть
подтвержден только с помощью ЭКГ. Пациенту рекоменду-
ется обратиться к терапевту.
Уважаемый покупатель,
Прибор был разработан в сотрудничестве с врачами, а клини-
ческие тесты подтвердили высокую точность его измерений.*
Microlife AF
IBsens -это ведущая в мире цифровая технология
измерения артериального давления для обнаружения
фибрилляции предсердий (AF) и артериальной гипертензии.
Это два главных фактора риска возникновения инсульта или
болезни сердца. Важно определить AF и гипертонию на
ранней стадии, даже если у вас нет никаких симптомов. AF-
скрининг совместно с алгоритмом Microlife AFIB рекоменду-
ется для людей старше 65 лет. Алг
оритм AFIB указывает на
то, что может присутствовать фибрилляция предсердий. Если
устройство выдает сигнал AFIB во время измерения артери-
ального давления, следует обратиться к врачу. Алгоритм
AFIBMicrolife был клинически проверен несколькими извест-
нымипрофессионалами и показал, что устройство выявляет-
пациентов с AFIB с точностью 97-100%.
1,2
При возникновении вопросов, проблем или для заказа запа-
сных частей, пожалуйста, обращайтесь в местный сервисный
центр Microlife. Ваш дилер или аптека могут предоставить Вам
адрес дилера Microlife в Вашей стране. В качестве альтерна-
тивы, посетите в Интернете страницу www.microlife.com, где
Вы сможете найти ряд полезных сведений по нашему
изделию.
Будьте здоровы – Microlife AG!
Microlife BP A6 BT
RU
1
Кнопка СТАРТ/СТОП
2
Дисплей
3
Кнопка M (Память)
4
Кнопка Time (Время)
5
Гнездо для манжеты
6
Переключатель AFIB/MAM
7
Индикатор уровня давления
8
Порт USB
9
Гнездо для блока питания
AT
Отсек для батарей
AK
Манжета
AL
Соединитель манжеты
AM
- Кнопка «Назад»
AN
+ Кнопка «Вперед»
AO
Переключатель блокировки
AP
Систолическое давление
AQ
Диастолическое давление
AR
Частота пульса
AS
Индикатор разряда батарей
BT
Индикатор правильности надевания манжеты
BK
Индикатор движения ру
ки
BL
Интервал времени MAM
BM
Индикатор мерцательной аритмии (AFIB)
BN
Режим AFIB/MAM
BO
Сохраненное значение
BP
Дата/Время
BQ
Индикатор пульса
Хранить в сухом месте
10
* В приборе использована та же технология измерений, что
и в отмеченной наградами модели «BP 3BTO-A», которая
успешно прошла клинические испытания в соответствии с
протоколом Британского и Ирландского Гипертонического
Общества (BIHS).
1
Kearley K, Selwood M, Van den Bruel A, Thompson M, Mant D,
Hobbs FR et al.: Triage tests for identifying atrial fibrillation in
primary care: a diagnostic accuracy study comparing single-lead
ECG and modified BP monitors. BMJ Open 2014; 4:e004565.
2
Wiesel J, Arbesfeld B, Schechter D: Comparison of the Microlife
blood pressure monitor with the Omron blood pressure monitor for
detecting atrial fibrillation. Am J Cardiol 2014; 114:1046-1048.
Table of Contents
1. Appearance of the Atrial Fibrillation Indicator for early
Detection (Active only in AFIB/MAM mode)
What is Atrial Fibrillation (AF)?
Who should be screened for Atrial Fibrillation?
Risk factors you can control
2. Using the Device for the First Time
Inserting the batteries
Setting the date and time
Selecting the correct cuff
Selecting standard or AFIB/MAM mode
3. Checklist for Taking a Reliable Measurement
4. Taking a Blood Pressure Measurement
How not to store a reading
How do I evaluate my blood pressure?
5. Data Memory
Viewing the stored values
Clearing all values
6. Battery Indicator and Battery change
Low battery
Flat battery – replacement
Which batteries and which procedure?
Using rechargeable batteries
7. Using a Mains Adapter
8. Bluetooth® Function
9. PC-Link Functions
10. Error Messages
11. Safety, Care, Accuracy Test and Disposal
Safety and protection
•Device care
Cleaning the cuff
Accuracy test
Disposal
12. Guarantee
13. Technical Specifications
Guarantee Card (see Back Cover)
1. Появление индикатора мерцательной аритмии
для ранней диагностики (активируется только в
режиме AFIB/MAM)
Этот прибор может выявлять мерцательную аритмию (AF).
Этот символ BM обозначает, что мерцательная аритмия обна-
ружена во время измерения. Пожалуйста, обратитесь к следу-
ющему абзацу, чтобы получить информацию касательно
консультации с Вашим врачом.
Держите руку неподвижно во время измерения, чтобы
избежать ошибочных результатов.
Информация для врача в случае частого появления
индикатора мерцательной аритмии
Этот прибор является осциллометрическим измерителем
артериального давления, который анализирует также нере-
гулярность пульса во время измерения. Прибор прошел
клинические испытания.
Символ мерцательной аритмии появляется на дисплее в
случае, если мерцательная аритмия присутствовала во
время измерения. Если символ AFIB появляется после окон-
чания полного измерения артериального давления (трех-
кратного последовательного измерения), пациенту реко-
мендуется пов
торить измерение (трехкратное последова-
тельное измерение). Если снова появится символ AFIB -
пациенту рекомендуется обратиться за помощью к врачу.
Если на экране монитора измерителя артериального
давления появляется символ AFIB, это указывает на
возможное присутствие мерцательной аритмии. Диагноз
мерцательной аритмии, однако, должен быть сделан карди-
ологом на основе расшифровки ЭКГ.
11BP A6 BT
RU
Этот прибор может не обнаруживать мерцательную
аритмию у людей с кардиостимуляторами или кардио-
дефибрилляторами.
При наличии фибрилляции предсердий значение
диастолического артериального давления может быть
неточным.
При фибрилляции предсердий для надежных пока-
заний, давление рекомендуется измерять в режиме
MAM.
Что такое мерцательная аритмия (AF)?
В норме сердце сокращается и расслабляется в регулярном
ритме. Определенные клетки в сердце генерируют электриче-
ские сигналы, которые вызывают сокращения сердца и пере-
качивание крови. Мерцательная аритмия возникает, когда
быстрые, беспорядочные электрические сигналы в двух
верх
них камерах сердца, называемых предсердиями, вызы-
вают их нерегулярные сокращения (это называется фибрил-
ляциями). Мерцательная аритмия является наиболее распро-
страненной формой сердечных аритмий. Часто при этом не
возникает никаких симптомов, но значительно увеличивается
риск возникновения инсульта. Вам необходимо обратиться к
доктору, чтобы контролировать эту проблему.
Кто должен быть обследован на наличие мерца
тельной
аритмии?
Скрининг AF рекомендуется для людей старше 65 лет, так как
с возрастом возрастает вероятность возникновения инсульта.
Скрининг AF также рекомендуется для людей в возрасте от 50
лет, имеющих высокое артериальное давление (например,
SYS выше 159 или DIA выше 99), а также с диабетом, ишеми-
ческой болезнью сердца, или для тех, кто ранее перенес
инсульт.
AF-скрининг не ре
комендуется проводить у молодых людей
или во время беременности, так как это может привести к
ошибочным результатам и ненужным тревогам. Кроме того,
молодые люди с диагнозом AF имеют низкую вероятность
возникновения инсульта по сравнению с людьми пожилого
возраста.
Факторы риска, которыми Вы можете управлять
Ранняя диагностика мерцательной аритмии с последующим
лечением может зн
ачительно снизить риск возникновения
инсульта. Знание вашего артериального давления и знание,
есть ли у вас мерцательная аритмия - это первый шаг в
профилактике инсульта.
For more information visit our website: www.microlife.com/afib.
2. Использование прибора в первый раз
Inserting the batteries
Передвиньте переключатель блокировки AO в положение
«Разблокировано». Отсек для батарей AT расположен на
нижней части прибора. Вставьте батареи (4 х 1,5В, размер
AAA), соблюдая полярность.
Установка даты и времени
1. После установки новых батарей на дисплее замигает
числовое значение года. Год устанавливается нажатием
кнопки M 3. Для того, чтобы подтвердить введенное
значение и затем устан
овить месяц, нажмите кнопку Time
(Время) 4.
2. Нажмите кнопку «+» AN или «-» AM для установки месяца.
Нажмите кнопку Time (Время) 4 чтобы подтвердить и
затем установите день.
3. Следуя вышеприведенным инструкциям, установите день,
час и минуты.
4. После установки минут и нажатия кнопки Time (Время) на
экране появятся дата и время.
5. Для изменения даты и времени нажмите и удер
живайте
кнопку Time (Время) приблизительно в течение 3 секунд,
пока не начнет мигать год. После этого можно ввести новые
значения, как это описано выше.
Подбор подходящей манжеты
Microlife предлагает манжеты разных размеров. Выберите
манжету, размер которой соответствует обхвату Вашего
плеча (измеренному при плотном прилегании посредине
плеча).
Пользуйтесь только манжетами Microlife!
Размер манжеты для обхвата плеча
S 17 - 22 cm (см)
M 22 - 32 cm (см)
M - L 22 - 42 cm (см)
L 32 - 42 cm (см)
L - XL 32 - 52 cm (см)
12
Обратитесь в местный сервисный центр Microlife, если
приложенная манжета AK не подходит.
Подсоедините манжету к прибору, вставив соединитель
манжеты AL в гнездо манжеты 5 до упора.
Если вы приобрели запасную манжету Microlife: отсое-
дините разъем трубки для подачи воздуха AL от
манжеты AL, поставленной с устройством, и подсоеди-
ните этот разъем к запасной манжете (разъем подходит
для всех размеров манжеты).
Выберите режим «Обычный» или «AFIB/MAM»
Перед каждым измерением, необходимо выбрать режим
работы прибора: обычный режим (одинарное измерение) или
режим AFIB/MAM (тр
ойное измерение). В режиме AFIB/MAM
прибор автоматически выполняет 3 измерения, которые
следуют один за другим, данные всех выполненных изме-
рений автоматически анализируется и результат выводится
на дисплей. Так как артериальное давление постоянно изме-
няется, результат, полученный таким образом, более
надежен, чем одно измерение.
Чтобы выбрать режим AFIB/MAM, сдвиньте переключатель
AFIB/MAM, 6 вверх до положения «3»,
пока на дисплее не
появится символ MAM BN. Чтобы перейти в стандартный
режим (одно измерение), сдвиньте переключатель AFIB/
MAM вниз в положение «1».
В нижнем правом участке дисплея отображается цифра 1,
2 или 3, указывающая на то, какое из трех измерений
выполняется в настоящий момент.
Между измерениями установлен перерыв в 15 секунд.
Отсчет отображает оставшееся время.
Отдел
ьные результаты не отображаются. Артериальное
давление будет показано только после выполнения трех
измерений.
Не снимайте манжету между измерениями.
Если одно из трех отдельных измерений вызывает
сомнения, то автоматически будет произведено четвертое.
Определение фибрилляции предсердий (аритмия)
возможно только в режиме AFIB/MAM.
3. Рекомендации для получения точных резуль-
татов измерения
Избегайте физической активности, не ешьте и не курите
непосредственно перед измерением.
Присядьте на стул со спинкой на пять минут и расслабь-
тесь. Поставьте ноги на пол ровно и не скрещивайте их.
Всегда проводите измерения на одной и той же руке
(обычно на левой). Рекомендуется, чтобы во время
первого визита пациента, врач провел измерения на двух
руках, чтобы определить на какой руке нужно производить
измерения в дальнейшем. Потом измерения проводятся
только на той руке, на которой давление оказалось выше.
Снимите облегающую одежду с плеча. Не следует закаты-
вать рукав рубашки, это может привести к сдавливанию,
рукава из тонкой ткани не мешают измерению, если приле-
гают свободно.
Всегда проверяйте, что используется правильный размер
манжеты (маркировка на манжете).
Плотно наложите манжету, но не слишком туго.
Убедитесь, что манжета расположена на 1-2 cm (см)
выше локтя.
Метка артерии на манжете (линия длиной около 3 cm
(см)) должна находиться над артерией с внутренней
стороны руки.
Расположите руку так, чтобы она оставалась рассла-
бленной.
Уб
едитесь, что манжета находится на той же высоте, что
и сердце.
4. Выполнение измерений артериального
давления
1. Выберите «обычный режим» (одинарное измерение) или
«Режим AFIB/MAM (тройное измерение): подробная
информация см. главу «2.».
2. Нажмите кнопку СТАРТ/СТОП 1 для начала измерения.
3. Теперь будет произведена автоматическая накачка
манжеты. Расслабьтесь, не двигайтесь и не напрягайте
руку до тех пор, пока не отобразится результат измерения.
Дышите нормально и не разговаривайте.
4. Если измерение успеш
но завершено, подкачка прекраща-
ется и происходит постепенный сброс давления. Если
требуемое давление не достигнуто, прибор автоматически
произведет дополнительное нагнетание воздуха в
манжету.
5. Во время измерения, индикатор пульса BQ мигает на
дисплее.
13BP A6 BT
RU
6. Затем отображается результат, состоящий из систоличе-
ского AP и диастолического AQ артериального давления, а
также пульса AR. См. также пояснения по другим показа-
ниям дисплея в этой инструкции.
7. По окончанию измерения снимите и уберите манжету.
8. Отключите прибор. (Прибор автоматически отключится
приблизительно через 1 минуту).
Определение фибрилляции предсердий (аритмия)
возможно только в режиме AFIB/MAM.
Вы можете остановить измерение в любой момент
нажатием кнопки СТАРТ/СТОП (например, если Вы
испытываете неудобство или неприятное ощущение от
нагнетаемого давления).
Этот прибор специально тестировался для применения
при беременности и преэклампсии. Если во время
беременности Вы обнаружили необычно высокий
результат, то Вы должны осуществить повторное изме-
рение (например через 1 час). Если результат по-преж-
нему высокий, то проконсультируйтесь со своим
лечащим врачом или гинекологом.
Во время беременности значок AFIB можно игнориро-
вать.
Как от
менить сохранение результата
Когда результат появится на дисплее, нажмите и держите
нажатой кнопку СТАРТ/СТОП. 1 пока кнопка «M» BO не
начнет мигать. Подтвердите удаление, нажав кнопку M 4.
Как определить артериальное давление
Световые диоды в левой части дисплея 7 указывают на
диапазон, в который попадает измереное артериальное
давление. Измереное давление находится либо в опти-
мальном (зелены
й), повышенном (желтый), либо высоком
(красный) диапазоне. Классификация данных по измеренному
давлению на диапозоны происходит по международным
директивам (ESH, ESC, JSH). Данные выражены в мм.рт.ст.
Оценка давления определяется по наивысшему значению.
Например: давление 140/80 mm Hg (мм рт.ст.) и давление 130/
90 mm Hg (мм рт.ст.) оба оцениваются как «артериальное
да
вление очень высокое».
5. Память
Это устройство автоматически сохраняет последние 99 изме-
рений.
Просмотр сохраненных величин
Передвиньте переключатель блокировки AO в положение
«Разблокировано». Кратко нажмите кнопку M 3. Сначала на
дисплее покажется знак «M» BO и «A», который обозначает
среднее всех сохраненных значений.
Нажимайте кнопки «+» AN или «-» AM чтобы просмотреть
сохраненные в памяти значения. Снова нажмите кнопку М
что
бы выйти из режима пямяти.
Обратите внимание, что максимальный объем памяти в
99 измерений не может быть превышен. Когда память
заполнена, 100th измерение записывается вместо
самого раннего. Значения должны быть отслежены
врачом до достижения максимального объема памяти
иначе данные будут потеряны.
Удаление всех значений
Если Вы уверены в том, что хотите удалить все хранимые
значения без возможности восстановления, уд
ерживайте
кнопку M в нажатом положении (предварительно прибор необ-
ходимо выключить) до появления «CL» и затем отпустите
кнопку. Для очистки памяти нажмите кнопку M в момент, когда
мигает «CL». Отдельные значения не могут быть удалены.
Отмена удаления: нажмите кнопку СТАРТ/СТОП 1 во
время мигания «CL».
Диапазон
Систо-
личе-
ское
Диасто-
личе-
ское Рекомендация
1. артериальное
давление
слишком
высокое
135 85 Обратитесь за
медицинской
помощью
2. повышенное
артериальное
давление
130 - 134 80 - 84 Самостоя-
тельный контроль
3. артериальное
давление в
норме
<130 <80 Самостоя-
тельный контроль
Диапазон
Систо-
личе-
ское
Диасто-
личе-
ское Рекомендация
14
6. Индикатор разряда батарей и их замена
Батареи почти разряжены
Если батареи использованы приблизительно на ¾, то при
включении прибора символ элементов питания AS будет
мигать (отображается частично заряженная батарея).
Несмотря на то, что прибор продолжит надежно проводить
измерения, необходимо подготовить новые элементы
питания на замену.
Замена разряженных батарей
Если батареи разряжены, то при включении прибора символ
элементов питания AS бу
дет мигать (отображается разря-
женная батарея). Дальнейшие измерения не могут произво-
диться до замены батарей.
1. Откройте отсек батарей AT на нижней части прибора.
2. Замените батареи, убедившись, что соблюдена поляр-
ность в соответствии с символами в отсеке.
3. Для того, чтобы установить дату и время, следуйте проце-
дуре, описанной в «Разделе 2.».
В памяти сохраняются все значения, но дата и время
будут сброшеныпоэтому после замены батарей год
автоматически замигает.
Элементы питания и процедура замены
Пожалуйста, используйте 4 новые щелочные батареи
на 1,5V (В) с длительным сроком службы размера AAA.
Не используйте батареи с истекшим сроком годности.
Дocтaньте батареи, если прибор не будет использо-
ваться в течение длительного периода времени.
Использование аккумуляторов
С прибором можно работать, используя аккумуляторные
батареи.
Пожалуйста, используйте только один тип аккумуля-
торных батарей «NiMH».
Батареи необходимо вынуть и перезарядить, если
появляется символ элементов питания (разряженная
батарея). Они не должны оставаться внутри прибора,
поскольку могут выйти из строя (даже в выключенном
приборе батареи продолжают разряжаться).
Всегда вынимайте аккумуляторы, если не собираетесь
пользоваться прибором в течение недели или более!
Аккумуляторы НЕ могут заряжаться в приборе!
Повторно зарядите аккумуляторы во внешнем
зарядном устройстве и ознакомьтесь с информацией
по зарядке, уходу и сроку службы!
7. Использование блока питания
Вы можете работать с прибором при помощи блока питания
Microlife (Постоянный ток 6В, 600 мА).
Используйте только блоки питания Microlife, относя-
щиеся к оригинальным принадлежностям и рассчи-
танные на соответствующее напряжение.
Убедитесь в том, что ни блок питания, ни кабель не
повреждены.
1. Вставьте кабель блока питания в гнездо блока питания 9
в приборе.
2. Вставьте вилку блока питания в розетку.
При подключении блока питания ток элементов питания не
потребляется.
8. Функция Bluetooth
®
Это устройство можно использовать в сочетании сосмар-
тфоном, на котором установлено «Microlife Connected
Health+» App. Bluetooth
®
автоматически активируется после
завершения измерения.
1. Активируйте функцию Bluetooth
®
на вашем смартфоне.
2. Активируйте функцию Bluetooth® на вашем приборе.
Нажмите кнопку «+» AN, когда прибор находится в режиме
ожидания. На дисплее начнет мигать символ «bt».
Для получения более подробной информации посетите сайт
www.microlife.com/connect.
9. Функции связи с компьютером
Прибор может подключаться к персональному компьютеру
(ПК) при использовании программного обеспечения Microlife
Blood Pressure Analyzer+ (BPA+). Данные из памяти переда-
ются на ПК путем соединения монитора с кабелем ПК.
Если в комплекте нет ваучера для загрузки и кабеля, то
скачайте программное обеспечение BPA + с веб-сайта
www.microlife.com/software и используйте USB-кабель с 5-
контактным разъемом Mini-B
Во время соединения с ПК прибор управляется компью-
тером.
15BP A6 BT
RU
10.Сообщения об ошибках
Если во время измерения происходит ошибка, то процедура
измерения прерывается и выдается сообщение об ошибке,
например «Err 3».
* Пожалуйста, немедленно проконсультируйтесь с врачом,
если эта или какая-либо другая проблема возникает
повторно.
Если Вам кажется, что результаты отличаются от
обычных, то, пожалуйста, внимательно прочтите
информацию в «Разделе 1.».
11.Техника безопасности, уход, проверка точности
и утилизация
Техника безопасности и защита
Следуйте инструкциям по использованию. В этом доку-
менте содержатся важные сведения о работе и безопа-
сности этого устройства. Перед использованием устрой-
ства, пожалуйста, внимательно прочитайте этот документ и
сохраните его для дальнейшего использования.
Прибор может использоваться только в целях, описанных в
данной инструкции. Изготовитель не несет ответствен-
нос
ти за повреждения, вызванные неправильным исполь-
зованием.
В состав прибора входят чувствительные компоненты,
требующие осторожного обращения. Ознакомьтесь с усло-
виями хранения и эксплуатации, описанными в разделе
«Технические характеристики»!
Оберегайте прибор от:
- воды и влаги
- экстремальных температур
- ударов и падений
- загрязнения и пыли
- прямых солнечных лучей
- жары и холода
Манже
ты представляют собой чувствительные элементы,
требующие бережного обращения.
Ошибка Описание Возможная причина и устранение
«Err 1» Сигнал
слишком
слабый
Импульсные сигналы на манжете
слишком слабые. Повторно наложите
манжету и повторите измерение.*
«Err 2»
BK
Ошибочны
е сигналы
Во время измерения манжета зафик-
сировала ошибочные сигналы,
вызванные, например, движением
или сокращением мышц. Повторите
измерение, держа руку неподвижно.
«Err
3»
BT
Отсутст-
вует
давление в
манжете
Манжета не может быть накачана до
необходимого уровня давления.
Возможно, имеет место утечка.
Проверьте, что манжета подсоеди-
нена правильно и не слишком
свободна. При необходимости заме-
ните батареи. Повторите измерение.
«Err 5» Ошибочны
й результат
(артефакт)
Сигналы измерения неточны, из-за
чего отображение результатов невоз-
можно. Пр
очтите рекомендации для
получения надежных результатов
измерений и затем повторите изме-
рение.*
«Err 6» Режим AFI
B/MAM
Во время измерения произошло
слишком много ошибок в режиме
AFIB/MAM, поэтому получение окон-
чательного результата невозможно.
Прочтите рекомендации для полу-
чения надежных результатов изме-
рений и затем повторите измерение.*
«HI» Пульс или
давление
манжеты
сл
ишком
высоки
Давление в манжете слишком
высокое (свыше 299 mm Hg (мм
рт.ст.)) ИЛИ пульс слишком высокий
(свыше 200 ударов в минуту). Отдох-
ните в течение 5 минут и повторите
измерение.*
«LO» Пульс
слишком
низкий
Пульс слишком низкий (менее
40 ударов в минуту). Повторите изме-
рение.*
«FL» Проблема
с Blue-
tooth
®
соединени
ем
Мигающий символ «FL» на экране
означает, что соединение прибора со
смартфоном не было установлено.
Убедитесь, что функция Bluetooth®
активирована, и повторите процедуру.
Ошибка Описание Возможная причина и устранение
16
Не меняйте другие части манжеты или коннектор манжеты
для измерений с этим устройством.
Производите накачку только наложенной манжеты.
Не используйте устройство вблизи источников сильных
электромагнитных полей, например рядом с мобильными
телефонами или радиостанциями. Во время использо-
вания устройства минимальное расстояние от источников
таких полей должно составлять 3,3 m (м).
Не используйте прибор, ес
ли вам кажется, что он повре-
жден, или если вы заметили что-либо необычное.
Никогда не вскрывайте прибор.
Если прибор не будет использоваться в течение длитель-
ного периода времени, то из него следует вынуть бата-
рейки.
Прочтите дальнейшие указания по безопасности в
отдельных разделах этой инструкции.
Результаты и
змерения, которые предоставляет этот
прибор, не являются диагнозом. Они не заменяют необхо-
димость консультации врача, особенно если они не соот-
ветствуют симптомам пациента. Не полагайтесь только на
результат измерения, всегда рассматривайте другие
потенциальные симптомы и жалобы пациента. Обратитесь
к врачу или вызовите скорую в случае необходимости.
Постоянно повышенное артериальное давление может
нанес
ти ущерб Вашему здоровью, и в этом случае Вам
необходимо обратиться к врачу!
Всегда сообщайте врачу о Вашем давлении и сообщайте
ему/ей, если Вы заметили что-нибудь необычное или
чувствуете неуверенность. Никогда не полагайтесь на
результат однократного измерения артериального
давления.
Ни при каких обстоятельствах не меняйте дозировку
лекарств и не занимайт
есь самолечением без консуль-
тации вашего лечащего врача.
Расхождения между результатами измерений, получен-
ными врачом или в аптеке, и результатами, полученными в
домашних условиях, также являются вполне нормальными,
поскольку ситуации, в которых проводятся измерения,
совершенно различны.
Показания пульса не пригодны для использования в
качестве контроля частоты кардиостимулятора!
Во вре
мя беременности следует тщательно следить за
артериальным давлением, поскольку на протяжении этого
периода оно может существенно меняться!
Позаботьтесь о том, чтобы дети не могли использовать
прибор без присмотра, поскольку некоторые его мелкие
части могут быть проглочены. При поставке прибора с
кабелями и шлангами возможен риск удушения.
Уход за прибором
Используйте для чистки прибора только сухую, мягкую ткань.
Очистка манжеты
Осторожно удалите пятна с манжеты с помощью влажной
тряпки и мыльной воды.
ПРЕДУПРЕЖДЕНИЕ: Ни при каких обстоятельствах не
допускается стирка внутренней эластичной камеры!
Проверка точности
Мы рекомендуем проверять точность прибора каждые 2 года
либо после механического воздействия (например, падения).
Для проведения теста обратитесь в местный сервисный центр
Microlife (см. введение).
Утилизация
Батареи и электронные приборы следует утилизиро-
вать в соответствии с принятыми нормами и не выбра-
сывать вместе с бытовыми отходами.
12.Гарантия
На прибор распространяется гарантия в течение 5 лет с даты
приобретения. В течение этого гарантийного периода, по
нашему усмотрению, Microlife бесплатно отремонтирует или
заменит неисправный продукт.
Вскрытие или изменение устройства аннулирует гарантию.
Следующие пункты исключены из гарантии:
Транспортные повреждения и риски, связанные с тран-
спортом.
Повреждения, вызванные неправильным применением
или несоблюдением инструкции по п
рименению.
Повреждения, вызванные утечкой батарей.
Повреждения, вызванные несчастным случаем или непра-
вильным использованием.
Упаковка / хранение материалов и инструкции по приме-
нению.
17BP A6 BT
RU
Регулярные проверки и обслуживание (калибровка).
Аксессуары и изнашивающиеся части: Батареи, адаптер
питания (при необходимости).
На манжету распространяется гарантия (герметичность
воздушного клапана) на 2 года.
Если требуется гарантийное обслуживание, обратитесь к
дилеру, у которого был приобретен продукт, или в местную
службу поддержки Microlife. Выможете связаться с местным
сервисом Microlife через наш сайт:
www.microlife.com/support
Компенсация ограничена стоимос
тью продукта. Гарантия
будет предоставлена, если весь товар будет возвращен с
оригинальным счетом. Ремонт или замена в рамках гарантии
не продлевает и не восстанавливает сначала гарантийный
срок. Юридические претензии и права потребителей не огра-
ничены этой гарантией.
13.Технические характеристики
Данный прибор соответствует требованиям директивы ЕС о
медицинском оборудовании 93/42/EEC.
Право на внесение технических изменений сохраняется за
производителем.
Торговая марка и логотип Bluetooth
®
являются зарегистриро-
ванными торговыми марками, принадлежащими Bluetooth
SIG, Inc. Любое использование этих торговых марок компа-
нией Microlife Corp. является лицензированным. Другие
торговые марки и торговые названия принадлежат их соответ-
ствующим владельцам.
Условия
применения:
от +10 °C до +40 °C
максимальная относительная влаж-
ность 15 - 95 %
Условия хранения: от -20 °C до +55 °C
максимальная относительная влаж-
ность 15 - 95 %
Масса: 354 g (г) (включая батарейки)
Размеры: 138 x 94,5 x 62,5 mm (мм)
Процедура из
ме-
рения:
осциллометрическая, в соответствии с
методом Короткова: фаза I систоличе-
ская, фаза V диастолическая
Диапазон изме-
рений:
20 - 280 mm Hg (мм рт.ст.) – артери-
альное давление
40 - 200 ударов в минутупульс
Индикация
давления в
манжете: 0 - 299 mm Hg (мм рт.ст.)
Минимальный шаг
индикации: 1 mm Hg (мм рт.ст.)
Статическая
точность:
давление в пределах ± 3 mm Hg
(мм рт.ст.)
Точность изме-
рения пульса: ±5 %
считанного значения
Источник питания:
4 x 1,5V (В) щелочные батарейки
размера AAA
Блок питания постоянного тока 6В, 600
мА (опционально)
Срок службы
батареи:
примерно 400 измерений (при исполь-
зовании новых щелочных батарей)
Класс защиты: IP20
Соответствие стан-
дартам:
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Ожидаемый срок
службы:
Прибор: 5 лет или 10000 измерений
Комплектующие: 2 года
18
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94

Microlife BP A6 BT Manual de utilizare

Categorie
Unități de tensiune arterială
Tip
Manual de utilizare